MANUAL CHANGES September, 01
Manual for Model Number 81110A,81104A
Manual printed on April 2000
Edition1.1, E0400
Manual Part Number 81110-91021

Make all ERRATA corrections.

Check the following table for your instrument serial prefix/serial number/EDC and make the listed
changes to your manual

New Item

Serial Prefix or Manual Serial Prefix or Manual
Serial Number Change Serial Number Change

ERRATA

81104A 81110A
DE387 00466 DE387 00805
Serialnumber independent 1 Serialnumber independent 1




5 September, 2001 Page 1 of 4
MODEL 81110A,81104A
____________________________________________________________________


Reference Quide: 81110A/81105A (P/N 81110-91021), page 106

Chapter: Agilent 81110A/81105A Specifications

Width Accuracy
81111A: +/- 0.5% +/- 250ps(*) after selfcal, typical +/-3% +/-250ps(*) without selfcal
81105A: +/- 5% +/- 250ps(*)

(*) Width accuracy specification is valid up to 5.5Vpp amplitude. Above 5.5Vpp, the width
may increase typically up to +300ps .


Page 103, Power consumption: 170VA max.

Page 104, Specifications change to read:

Recalibration period
3 years recommended

Page 7, Output Modules for Agilent 81110A Mainframes add:

The 81111A output modules are fitted with single ended outputs. Therefore the connectors
for the complementary outputs have no functionality. Complementary outputs are only
available when 81112A output modules are installed.

(The 81110A always is fitted with connectors for the normal and the complementary
outputs, regardless of the configuration. It has been designed this way to allow the
easy retrofit of the output channels.)


Page 106, Glitch-free timing changes change to read:

This applies to continous mode with timing values < 100 ms (frequency: > 10 Hz),
and consecutive values between one-half and twice the previous value.


Page 106, Width add:

The pulse width is specified at fastest transitions.


Page 107, Delay add:

Delay and Double Pulse Delay are specified at fastest leading edges.




__________________________________________________________________________________
05.09.01 Page 2 of 4
MODEL 81110A,81104A
____________________________________________________________________

INDEX OF MANUAL CHANGE

MANUAL FRAME
CHANGE
ERRATA
1 See attached
Declaration of
Conformity




__________________________________________________________________________________
05.09.01 Page 3 of 4
MODEL 81110A,81104A
____________________________________________________________________

MANUAL CHANGE 1

On page 102, Specifications change to read:
See attached Declaration of conformity




__________________________________________________________________________________
05.09.01 Page 4 of 4
DECLARATION OF CONFORMITY
According to ISO/IEC Guide 22 and CEN/CENELEC EN 45014



Manufacturer's Name: Agilent Technologies Deutschland GmbH
Manufacturer's Address: Boeblingen Verifications Solutions (BVS)
Herrenberger Str. 130
D-71034 Boeblingen
Declares, that the product
Product Name: Family of Pulse-/Data Generators
System Number: 81100
Product Modules: 81101A 50 MHz Pulse/Pattern Generator
81104A 80 MHz Pulse/Pattern Generator
81110A 330/165 MHz Pulse/Pattern Generator
81111A 165 MHz , 10 V Output Module
81112A 330 MHz , 3.5 V Output Module
81130A 400/660 MHz Pulse/Pattern Generator
81131A 400 MHz , 3.5 V Output Module
81132A 660 MHz , 2.5 V Output Module
E8305A (a) VXI Plug-in 250 MHz Pulse Generator
(a)
E8306A VXI Plug-in 100 MHz Clock Generator
(a)
E8311A VXI Plug-in 165MHz Pulse/Pattern Generator
E8312A (a) VXI Plug-in 330MHz Pulse/Pattern Generator
Conforms with the following European Directives:
The product herewith complies with the requirements of the Low Voltage Directive 73/23/EEC and the EMC Directive 89/336/EEC (including
93/68/EEC) and carries the CE Marking accordingly.

Conforms with the following product standards:
EMC (Technical The product modules marked by (a) herewith comply with the requirements of the EMC Directive 89/336/EEC (including
Construction File) 93/68/EEC) and carry the CE Marking accordingly (European Union).
Against: EMC test specification EN 55011:1991 (Group 1, Class A)
As detailed in Electromagnetic Compatibility (EMC) Certificate of Compliance No. B801356L
Assessed by: CETECOM ICT Services GmbH, D-66117 Saarbruecken

Standard Limit
EMC IEC 61326-1:1997+A1:1998 / EN 61326-1:1997+A1:1998
CISPR 11:1997 / EN 55011:1998 Group 1 Class A (1)
IEC 61000-4-2:1995+A1:1998 / EN 61000-4-2:1995 4kV CD, 8kV AD
IEC 61000-4-3:1995 / EN 61000-4-3:1995 3 V/m, 80-1000 MHz
IEC 61000-4-4:1995 / EN 61000-4-4:1995 0.5kV signal lines, 1kV power lines
IEC 61000-4-5:1995 / EN 61000-4-5:1995 0.5 kV line-line, 1 kV line-ground
IEC 61000-4-6:1996 / EN 61000-4-6:1996 3V, 0.15-80 MHz
IEC 61000-4-11:1994 / EN 61000-4-11:1994 1 cycle/100%

Canada: ICES-001:1998
Australia/New Zealand: AS/NZS 2064.1

Safety IEC 61010-1:1990+A1:1992+A2:1995 / EN 61010-1:1993+A2:1995
Canada: CSA C22.2 No. 1010.1:1992

Supplemental Information:
(1)
The products were tested in a typical configuration with Agilent Technologies test systems.




2001-May-02 Hans-Martin Fischer
Date Name
Product Regulations Engineer
Title

For further information, please contact your local Agilent Technologies sales office, agent or distributor.