Installation and
Verification Manual

HP 8566B Spectrum Analyzer
Includes Option 400, Option 462,
and Option 857




HP Part No. 08566-90169
Printed in USA September 1993
Notice.
The information contained in this document is subject to change
without notice.
Hewlett-Packard makes no warranty of any kind with regard to this
material, including but not limited to, the implied warranties of
merchantability and fitness for a particular purpose. Hewlett-Packard
shall not be liable for errors contained herein or for incidental
or consequential damages in connection with the furnishing,
performance, or use of this material.




@ Copyright Hewlett-Packard Company 1993
All Rights Reserved. Reproduction, adaptation, or translation without
prior written permission is prohibited, except as allowed under the
copyright laws.
1400 Fountaingrove Parkway, Santa Rosa CA, 95403-1799, USA
Certification Hewlett-Packard Company certifies that this product met its
published specifications at the time of shipment from the factory.
Hewlett-Packard further certifies that its calibration measurements
are traceable to the United States National Institute of Standards and
Technology, to the extent allowed by the Institute's calibration facility,
and to the calibration facilities of other International Standards
Organization members.

Regulatory Chapter [cross reference to location of declaration of conformity &
Information noise information] contains regulatory information.




Warranty This Hewlett-Packard instrument product is warranted against defects
in material and workmanship for a period of one year from date of
shipment. During the warranty period, Hewlett-Packard Company
will, at its option, either repair or replace products which prove to be
defective.
For warranty service or repair, this product must be returned to a
service facility designated by Hewlett-Packard. Buyer shall prepay
shipping charges to Hewlett-Packard and Hewlett-Packard shall pay
shipping charges to return the product to Buyer. However, Buyer shall
pay all shipping charges, duties, and taxes for products returned to
Hewlett-Packard from another country.
Hewlett-Packard warrants that its software and firmware designated
by Hewlett-Packard for use with an instrument will execute
its programming instructions when properly installed on that
instrument. Hewlett-Packard does not warrant that the operation
of the instrument, or software, or firmware will be uninterrupted or
error-free.
LIMITATION OF WARRANTY
The foregoing warranty shall not apply to defects resulting from
improper or inadequate maintenance by Buyer, Buyer-supplied
software or interfacing, unauthorized modification or misuse,
operation outside of the environmental specifications for the
product, or improper site preparation or maintenance.
NO OTHER WARRANTY IS EXPRESSED OR IMPLIED.
HEWLETT-PACKARD SPECIFICALLY DISCLAIMS THE IMPLIED
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE.
EXCLUSIVE REMEDIES
THE REMEDIES PROVIDED HEREIN ARE BUYER'S SOLE AND
EXCLUSIVE REMEDIES. HEWLETT-PACKARD SHALL NOT BE
LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR
CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT,
TORT, OR ANY OTHER LEGAL THEORY.



...
III
Assistance Product maintenance agreewmats and other customer assistance
agreements are available for Hewlett-FWkard products.
Rw any assistance, contact your nearest Him&&t-Rzckard Sales and
Service Ome.




Safety Notes The following safety notes are used throughout this manual.
Familiarize yourself with each of the notes and its meaning before
operating this instrument.

Caution Caution denotes a hazard. It calls attention to a procedure that, if
not correctly performed or adhered to, could result in damage to or
destruction of the instrument. Do not proceed beyond a caution sign
until the indicated conditions are fully understood and met.

Warning Warning denotes a hazard. It calls attention to a procedure
which, if not correctly performed or adhered to, could result in
injury or loss of life. Do not proceed beyond a warning note until
the indicated conditions are fully understood and met.


Instruction The instruction manual symbol. The product is marked with this
ManuaI symbol when it is necessary for the user to refer to the instructions in
the manual.




iv
General Safety
Considerations
Warning Before this instruct is switched on, make sure it has been
properly grounded through the protective conductor of the ac
power cable to a socket outlet provided with protective earth
contact.
Any interruption of the protective (grounding) conductor, inside
or outside the instrument, or disconnection of the protective
earth terminal can result in personal injury.

Warning There are many points in the instrument which can, if contacted,
cause personal injury. Be extremely careful.
Any adjustments or service procedures that require operation
of the instrument with protective covers removed should be
performed only by trained service personnel.

Caution Before this instruct is switched on, make sure its primary power
circuitry has been adapted to the voltage of the ac power source.
Failure to set the ac power input to the correct voltage could cause
damage to the instrument when the ac power cable is plugged in.




V
How to Use This
Manual
This manual uses the Front-Panel Ke 3 This represents a key physically located on the
following instrument.
conventions: Screen Text This indicates text displayed on the instrument's
screen.




HP 8566B Included with the HP Model 8566B spectrum analyzer are manuals:
Documentation The Installation and Verification, the Operating and Programming
Manual, and the Performance Tests and Adjustments Manual.
Description

HP 8566B Installation HP part number 08566-90169
and Verification Contents: General information, installation, specifications,
Manual characteristics, and operation verification.

HP 8566B Operating HP part number 08566-90040
and Programming Contents: Manual and remote operation, including complete syntax
Manual and command description. Accopanying this manual is the seperate,
pocket-sized Quick Reference Guide, HP part number 59558970.

HP 8566B HP part number 08566-90168
Performance Tests and Contents: Electrical performance tests and adjustment procedures.
Adjustments Manual

HP 8566B RF Section HP part number 08566-90210 Contents: RF section service
Troubleshooting and information.
Repair Manual

HP 8566B IF-Display HP part number 08566-90085 Contents: IF-Display section service
Section information.
Troubleshooting and
Repair Manual




vi
Contents
1. General Information
What You'll Find in This Chapter . . . . . . . . . . . l-l
Introducing the HP 8566B . . . . . . . . . . . . . . . l-l
Safety . . . . . . . . . . . . . . . . . . . . . . . l-2
Accessories Supplied . . . . . . . . . . . . . . . . l-2
Accessories Available . . . . . . . . . . . . . . . . l-5
Options . . . . . . . . . . . . . . . . . . . . . . l-6
Instruments Covered by This Manual . . . . . . . . l-7
Serial Numbers . . . . . . . . . . . . . . . . . . . l-7
Calibration Cycle . . . . . . . . . . . . . . . . . . l-8
HP-IB Address Selection . . . . . . . . . . . . . . l-8
Bench Operation . . . . . . . . . . . . . . . . . . l-8
Electrostatic Discharge Information . . . . . . . . . . l-9
Static-Safe Accessories . . . . . . . . . . . . . . . l-10
Routine Maintenance . . . . . . . . . . . . . . . . . l-11
Cleaning the Display . . . . . . . . . . . . . . . . l-11
Cleaning the RF Section's Fan Filter . . . . . . . . . 1-12
Replacing the RF Section's Battery . . . . . . . . . 1-12
Ordering Information . . . . . . . . . . . . . . . . . 1-16
Direct Mail Order System . . . . . . . . . . . . . . 1-16
Direct Phone-Order System . . . . . . . . . . . . . 1-16
Returning the Instrument for Service . . . . . . . . . 1-17
Service lags . . . . . . . . . . . . . . . . . . . . 1-17
Original Packaging . . . . . . . . . . . . . . . . . 1-17
Other Packaging . . . . . . . . . . . . . . . . . . 1-18
Sales and Service Offices . . . . . . . . . . . . . . . l-20

2. Instailation
What You'll Find in This Chapter . . . . . . . . . . . 2-l
Safety . . . . . . . . . . . . . . . . . . . . . . . 2-l
Preparation for Use . . . . . . . . . . . . . . . . . . 2-2
Initial Inspection . . . . . . . . . . . . . . . . . . 2-2
Operating Environment . . . . . . . . . . . . . . . 2-3
Power Requirements . . . . . . . . . . . . . . . . 2-3
`Ib Install Standard Instruments . . . . . . . . . . . . 2-4
lb Install Option 908 and 913 Instruments . . . . . . . 2-7
lb Install Option 010 Instruments . . . . . . . . . . . 2-11
lb Set the HP-IB Address . . . . . . . . . . . . . . . 2-18
From the Front Panel . . . . . . . . . . . . . . . . 2-18
From the HP-IB Bus . . . . . . . . . . . . . . . . 2-18
From the HP-IB Address Switch . . . . . . . . . . . 2-18




Contents-l
3. Specifications
Introduction . . . . . . . . . . . . . . . . . . . . . 3-l
Frequency . . . . . . . . . . . . . . . . . . . . . . 3-l
Measurement Range . . . . . . . . . . . . . . . . 3-l
Displayed Values . . . . . . . . . . . . . . . . . . 3-l
Center Frequency . . . . . . . . . . . . . . . . 3-l
Frequency Span . . . . . . . . . . . . . . . . . 3-2
Resolution . . . . . . . . . . . . . . . . . . . . . 3-3
Resolution Bandwidth . . . . . . 3-3
Resolution Bandwidth (Option 462 6 dB*Bandwidthsj 3-3
Resolution Bandwidth (Option 462 Impulse
Bandwidth) . . . . . . . . . . . . . . . . . . 3-3
Bandwidth Shape . . . . . . . . . . . . . . . . . 3-4
Spectral Purity . . . . . . . . . . . . . . . . . . . 3-4
Amplitude . . . . . . . . . . . . . . . . . . . . . . 3-5
Measurement Range . . . . . . . . . . . . . . . . 3-5
Displayed Values . . . . . . . . . . . . . . . . . . 3-5
Scale . . . . . . . . . . . . . . . . . . . . . . 3-5
Accuracy . . . . . . . . . . . . . . . . . . . . . 3-6
Reference Lines Accuracy . . . . . . . . . . . . . . 3-6
Dynamic Range . . . . . . . . . . . . . . . . . . . 3-6
Spurious Responses . . . . . . . . . . . . . . . . 3-6
Image, Multiple, and Out-of-Band Responses . . . . 3-7
Residual Responses . . . . . . . . . . . . . . . . 3-8
Gain Compression . . . . . . . . . . . . . . . . 3-8
Displayed Average Noise Level (Sensitivity) . . . . 3-8
Amplitude Uncertainty . . . . . . . . . . . . . . . . 3-10
Table Footnotes . . . . . . . . . . . . . . . . . . . 3-12
Marker . . . . . . . . . . . . . . . . . . . . . . . . 3-13
Sweep . . . . . . . . . . . . . . . . . . . . . . . . 3-14
Inputs . . . . . . . . . . . . . . . . . . . . . . . . 3-14
RF INPUT . . . . . . . . . . . . . . . . . . . . . 3-14
IF INPUT . . . . . . . . . . . . . . . . . . . . . 3-14
outputs . . . . . . . . . . . . . . . . . . . . . . . 3-15
CAL OUTPUT 3-15
1STLOOUTPUT : : : : : : : : : : : : : : : : : : 3-15
SWEEP + TUNE OUTPUT . . . . . . . . . . . . . 3-15
Options . . . . . . . . . . . . . . . . . . . . . . . 3-16
400 Hz Power Line Frequency Operation Option 400 . 3-16
General . . . . . . . . . . . . . . . . . . . . . . . 3-17
HP-IB Interface Functions . . . . . . . . . . . . . . 3-17
Environmental . . . . . . . . . . . . . . . . . . . 3-17
Temperature . . . . . . . . . . . . . . . . . . . 3-17
Altitude . . . . . . . . . . . . . . . . . . . . . 3-17
Power Requirements . . . . . . . . . . . . . . . . 3-17
Humidity . . . . . . . . . . . . . . . . . . . . . . 3-17
Operation . . . . . . . . . . . . . . . . . . . . 3-17
Storage . . . . . . . . . . . . . . . . . . . . . 3-17
EM1 . . . . . . . . . . . . . . . . . . . . . . . . 3-17
X-Rays . . . . . . . . . . . . . . . . . . . . . . . 3-17
Serial Prefix 3004A and Above . . . . . . . . . . 3-17
Serial Prefix 3001A and Below . . . . . . . . . . 3-17
Warm-Up Time . . . . . . . . . . . . . . . . . . . 3-19
Operation . . . . . . . . . . . . . . . . . . . . 3-19
Frequency Reference . . . . . . . . . . . . . . . 3-19

Contents-2
Weight . . . . . . . . . . . . . . . . . . . . . . . 3-19
Dimensions . . . . . . . . . . . . . . . . . . . . . 3-20
4. Performance Characteristics
Frequency . . . . . . . . . . . . . . . . . . . . . . 4-l
Resolution . . . . . . . . . . . . . . . . . . . . . 4-l
Spectral Purity . . . . . . . . . . . . . . . . . . . 4-l
Noise Sidebands . . . . . . . . . . . . . . . . . 4-l
Power-Line Related Sidebands . . . . . . . . . . . 4-l
Stability . . . . . . . . . . . . . . . . . . . . . . 4-3
Residual FM . . . . . . . . . . . . . . . . . . . 4-3
Drift . . . . . . . . . . . . . . . . . . . . . . . 4-3
Amplitude . . . . . . . . . . . . . . . . . . . . . . 4-4
Third Order Intermodulation Distortion . . . . . . . 4-4
Third Order Intercept (TOI) . . . . . . . . . . . . . 4-4
Synthesis-Related Spurious Sidebands . . . . . . . . 4-5
Input Attenuator Uncertainties . . . . . . . . . . . 4-5
RF Gain Uncertainty . . . . . . . . . . . . . . . . 4-5
Inputs . . . . . . . . . . . . . . . . . . . . . . . . 4-6
IF INPUT . . . . . . . . . . . . . . . . . . . . . 4-6
RF INPUT . . . . . . . . . . . . . . . . . . . . . 4-6
VIDEO INP . . . . . . . . . . . . . . . . . . . . . 4-7
IF INP . . . . . . . . . . . . . . . . . . . . . . . 4-7
EXT TRIGGER . . . . . . . . . . . . . . . . . . . 4-7
FREQREFERENCEEXT . . . . . . . . . . . . . . 4-8
outputs . . . . . . . . . . . . . . . . . . . . . . . 4-8
Display Outputs . . . . . . . . . . . . . . . . . . 4-8
Recorder Outputs . . . . . . . . . . . . . . . . . . 4-8
SWEEP . . . . . . . . . . . . . . . . . . . . . . 4-8
VIDEO . . . . . . . . . . . . . . . . . . . . . . . 4-9
PENLIFT . . . . . . . . . . . . . . . . . . . . . . 4-9
21.4 MHz IF OUTPUT . . . . . . . . . . . . . . . . 4-9
FREQ REFERENCE INT . . . . . . . . . . . . . . . 4-9
10MHzOUT . . . . . . . . . . . . . . . . . . . . 4-9
VIDEO OUT . . . . . . . . . . . . . . . . . . . . 4-10
IF OUT 4-10
IF OUTPUT' : : : : : : : : : : : : : : : : : : : : 4-10
General . . . . . . . . . . . . . . . . . . . . . . . 4-10
Acoustic Noise Emission/ Geraeuschemession . . . . . 4-10
Function Descriptions . . . . . . . . . . . . . . . . . 4-11
Sweep . . . . . . . . . . . . . . . . . . . . . . . 4-11
Sweep Time . . . . . . . . . . . . . . . . . . . 4-11
Zero Frequency Span . . . . . . . . . . . . . . 4-11
Marker . . . . . . . . . . . . . . . . . . . . . 4-11
Video Bandwidth . . . . . . . . . . . . . . . . . 4-11
Cathode Ray Tube . . . . . . . . . . . . . . . . 4-12
Instrument State Storage . . . . . . . . . . . . . 4-12
Remote Operation . . . . . . . . . . . . . . . . 4-12




Contents-3
5. Operation Verification
What You'll Find in This Chapter . . . . . . . . . . . 5-1
Test System Configuration . . . . . . . . . . . . . . . 5-4
Equipment Connections . . . . . . . . . . . . . . . . 5-5
Program Loading . . . . . . . . . . . . . . . . . . . 5-6
Program Operation . . . . . . . . . . . . . . . . . . 5-7
HP-IB Addresses . . . . . . . . . . . . . . . . . . . 5-8
Error Messages . . . . . . . . . . . . . . . . . . . . 5-9
Test Descriptions . . . . . . . . . . . . . . . . . . . 5-9
1. Input Attenuator Switching Check . . . . . . . . 5-10
2. IF Gain Uncertainty . . . . . . . . . . . . . . . 5-11
3. Scale Fidelity (Log) . . . . . . . . . . . . . . . 5-12
4. Scale Fidelity (Linear) . . . . . . . . . . . . . . 5-13
5. Log Scale Switching Uncertainty . . . . . . . . . 5-14
6. Resolution Bandwidths . . . . . . . . . . . . . . 5-15
6. Impulse and Resolution Bandwidths (Option 462
Impulse Bandwidth) . . . . . . . . . . . . . . . 5-16
7. Line Related Sidebands . . . . . . . . . . . . . 5-17
8. Average Noise Level . . . . . . . . . . . . . . . 5-18
9. Residual Responses . . . . . . . . . . . . . . . 5-19
10. Sweep + Tune Out Accuracy . . . . . . . . . . 5-20
11. Second Harmonic Distortion . . . . . . . . . . . 5-21
12. Frequency Span Accuracy . . . . . . . . . . . . 5-22
13. Gain Compression . . . . . . . . . . . . . . . 5-23
14. Frequency Response . . . . . . . . . . . . . . 5-24
15. Third Order Intermodulation Distortion . . . . . 5-25
16. Calibrator Output Amplitude Accuracy . . . . . . 5-26
17. First LO Output Power . . . . . . . . . . . . . 5-27




Contents-4
Figures
l-l. HP 8566B with Accessories Supplied . . . . . . . . . l-3
l-2. AC Power Cables Available . . . . . . . . . . . . . l-4
l-3. Typical Serial Number Label . . . . . . . . . . . . l-8
l-4. Static-Safe Workstation . . . . . . . . . . . . . . . l-9
l-5. Display Bezel Screws . . . . . . . . . . . . . . . . l-11
l-6. Removing the Bottom Cover . . . . . . . . . . . . 1-13
l-7. Location of Al5 Controller Assembly . . . . . . . . 1-13
l-8. Location of Battery on Al5 Controller Assembly . . . 1-14
l-9. Factory Packaging Materials for each Section . . . . . 1-18
2-l. Rear-Panel Cable Connections . . . . . . . . . . . . 2-4
2-2. Voltage Selector Board and Fuse . . . . . . . . . . . 2-5
2-3. Removing the Handles and Trim . . . . . . . . . . . 2-7
2-4. Removing the Information-Card Tray . . . . . . . . 2-8
2-5. Attach the Rack Mount Flanges . . . . . . . . . . . 2-8
2-6. Voltage Selector Board and Fuse . . . . . . . . . . . 2-9
2-7. Rear-Panel Cable Connections . . . . . . . . . . . . 2-10
2-8. Removing the Handles and Trim . . . . . . . . . . . 2-11
2-9. Removing the Information-Card Tray . . . . . . . . 2-12
2-10. Attaching the Rack Handles and Flanges . . . . . . . 2-12
2-l 1. Rear-Panel Cable Connections . . . . . . . . . . . . 2-13
2-12. Voltage Selector Board and Fuse . . . . . . . . . . . 2-14
2-13. Attaching the Inner-Member Brackets . . . . . . . . 2-15
2-14. Slide Adapter for Non-HP System Rack Cabinets . . . 2-16
2-15. Removing the Bottom Cover . . . . . . . . . . . . 2-19
2-16. Location of Al5 Controller Assembly . . . . . . . . 2-19 .
2-17. Address Switch (Shown In Factory Preset Position) . . 2-20
2-18. Address Switch Set to 4 . . . . . . . . . . . . . . . 2-21
3-l. Specified Average Displayed Noise Level, 100 Hz to 2.5
GHz Non-preselected Tuning Range . . . . . . . . 3-9
3-2. Specified Average Displayed Noise Level, 2.0 GHz to 22
GHz Preselected Tuning Range . . . . . . . . . . 3-9
3-3. Instrument Dimensions with Handles . . . . . . . . 3-20
3-4. Instrument Dimensions without Handles . . . . . . . 3-20
4-l. Typical Spectrum Analyzer Resolution . . . . . . . . 4-l
4-2. Single Sideband Noise Normalized to 1 Hz BW versus
Offset from Carrier . . . . . . . . . . . . . . . 4-2
4-3. Typical SSB Noise at 5.0 GHz Center Frequency
Normalized to 1 Hz BW versus Offset from Carrier
and Analyzer Resolution. May be Limited by
Average Noise Level. . . . . . . . . . . . . . . 4-3
4-4. Typical Optimum Dynamic Range . . . . . . . . . . 4-4
5-l. Dual Bus (MTS) System Connection . . . . . . . . . 5-5
5-2. Single Bus System Connection . . . . . . . . . . . . 5-6
5-3. RF Input and Calibration Controls . . . . . . . . . . 5-8




Contents-5
lhbles
1 - 1. Static-Safe Accessories . . . . . . . . . . . . . . . l-10
l-2. Factory Packaging Materials . . . . . . . . . . . . . 1-18
l-3. Hewlett-Packard Sales and Service Offices . . . . . . 1-21
2-l. Decimal and Binary Address Codes . . . . . . . . . 2-22
5-l. Tests Performed . . . . . . . . . . . . . . . . . . 5-2
5-2. Tests Not Performed . . . . . . . . . . . . . . . . 5-3
5-3. Equipment Summary . . . . . . . . . . . . . . . . 5-4
5-4. HP-IB Addresses . . . . . . . . . . . . . . . . . . 5-8




Contents-6
1
General Information
What You'll Find in This chapter introduces you to the HP 8566B spectrum analyzer
This Chapter and its options and accessories. Refer to Chapter 2 for information
on inspecting and installing the HP 8566B. Refer to Chapter 3 and
Chapter 4 for a complete listing of instrument specifications and
characteristics. Refer to Chapter 5 for Operation Verification.
Introducing the HP 8566B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . l-2
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . l-2
Accessories Supplied . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . l-2
Accessories Available . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..l- 5
Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . l-6
Instruments Covered by This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . l-7
Serial Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . l-8
Calibration Cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..l- 8
HP-IB Address Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . l-9
Bench Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..l- 9
Electrostatic Discharge Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . l-10
Static-Safe Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . l-12
Routine Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . l-13
Cleaning the Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-13
Cleaning the RF Section's Fan Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . l-14
Replacing the RF Section's Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-14
Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . l-17
Returning the Instrument for Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . l-18
Service Tags . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..l-18
Original Packaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 18
Other Packaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . l-19
Sales and Service Offices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . l-21




Introducing the The HP 8566B spectrum analyzer is capable of measuring signals
HP 8566B from -135 dBm to +30 dBm over a frequency range of 100 Hz to
22 GHz. The HP 8566B provides preselection from 2 to 22 GHz. The
frequency range of the instrument can be extended, unpreselected,
to 110 GHz using HP 11970 Series mixers, and to 325 GHz using other
commercially available mixers.
The HP 8566B is a complete, self-contained instrument that requires
only an external ac power source for operation. A set of two ac
power cables, suitable for use in the country to which the instrument
is originally shipped, are included with the instrument.




General Information l-l
Safety Before installing or operating this instrument, you should familiarize
yourself with the safety marking on the instrument and the safety
instructions in the manuals. The instrument has been manufactured
and tested in accordance with international safety standards.
However, to ensure safe operation of the instrument and personal
safety of the user, the cautions and warnings in the manuals must be
followed. Refer to the summary of the safety information located
near the front of this manual.

Accessories Supplied Figure l-l illustrates the instrument with the supplied accessories. In
accordance with international safety standards, both sections of this
instrument are equipped with three-wire ac power cables. Various
power cables are available to connect the HP 8566B to the types of ac
power outlets unique to specific geographic areas. See Figure l-2. The
cables appropriate for the area to which the instrument is originally
shipped are included with the instrument.



FRONT VIEW



I F- D i s p l a y
Se c t i o n




RF Sect ion




REAR VIEW h


IF- Display
Sect ion




RF Section




CABLES




1-2 General Information
Item Description HP Fart Number
1 1ST LO OUTPUT SMA termination 1810-0118
2 Front-panel IF semirigid jumper 85660-20101
3 BNC jumper cable (quantity 3) 85660-60117
4 Bus interconnect cable (W31) 85662-60220
5 Coax interconnect cable (W30) 85662-60093
6 Line power cables (2 each) see Figure l-2



Figure l-l. HP 8566B with Accessories Supplied




General Information 1.3
CABLE
I
PLUG I CABLE CABLE
I FOR USE
PLUG TYPE * * HP PART DESCRIPTION LENGTH COLOR IN COUNTRY
NUMBER CM ( I N C H E S )

250V
8120-1351 Straight* BS1363A 229 (90) Mint Gray Great Britain,
8120-1703 90" 229 (90) Mint Gray Cyprus. N i g e r ia,
Singapore,
2 imbabwe




I
FE%
Argentina,




0
250V 8120-1369 Straight* NZSS198/ASCll 201 (79) Gray
Australia,
8 120-0696 90