File name Cost-Effectice.pdf Cost Effective Techniques used to Validate the Performance of the microK Resistance Thermometry Instrument with sub mK Uncertainty Paul Bramley Metrosol Limited, Towcester, UK Neil Robinson Isothermal Technology Limited, Southport, UK This article describes techniques used to validate the performance of the microK thermometry instrument, which was designed for use in secondary temperature calibration and high accuracy temperature measurement applications (to <0.4mK uncertainty). The microK is the first precision thermometer that can work with both resistance thermometers and thermocouples and provides sub mK uncertainties with SPRTs and mK uncertainties with thermocouples. With a resistance ratio accuracy of 0.4ppm and voltage accuracy of 0.25 |