File name A6013L Capacitor Meter Tester.pdfA6013L 2. SPECIFICATIONS
DIGITAL C METER OPERATION MANUAL 2-1.GENERAL SPECIFICATIONS
1. FEATURES Display :LCD (Liquid Crystal Display) Max
Easy and correct readout. Indication 1999.
High measuring accuracy. LCD Backlight and data hold.
Measurements are possible even under a strong Measurement:C (Capacitance)
magnetic field. Range :single 9 position, whole range value (from
LSI-circuit provides high reliability and durability. 0.1pF to 20000uF)
Input overload protection is provided. Zero Adjustment :Manual (range:±20pF)
LCD display for low power consumption and clear Calibrate Adjustment :Have two internal adjustment. One
readout even in bright ambient light conditions. is panel Zero adjustment.
In-line pushbuttons allow one hand operation. Over-input :Display shows "1".
Light-weight and compact construction for easy Sampling Time :3/second
operation. Operating Temp :0 to 40.
Low battery condition is indicated on the LCD Operating Humidity :80% MAX.R.H.
display. Power Supply:Single, standard 9 volt battery. NEDA
1604IEC6F22
Battery Life :Alkaline type approx.: 200 hours.
2. Panel Layout Zinc-Carbon type approx.: 100 hours
LCD Display: 3½ Typical consumption current :3~4mA (200pF-200uF)
digits, character Standard Accessories: Test alligator clips (red &
15mm high black)...1 pair.
Back Light Button Instruction manual...1 pc.
Switch: Press this
button to switch on 2-2. ELECTRICAL SPECIFICATION
back light if the dark Accuracy is ±(percentage of reading + number of digit)
circumstance light at 23±5,<80%RH.
Range Accuracy Resolution Test Max
makes the reading Frequency indication
|