File name www.thinksrs.com-UGA comparison.pdf Detailed Comparison of SRS UGA Systems UGA Hydrogen UGA UGA LT Model UGA100 /3 UGA100 ULT100 Numbers UGA200 /3 UGA200 ULT200 UGA300 /3 UGA300 ULT300 Applications Up to 100% H2 Less than 25% H2 Less than 3% H2 Corrosive gases Corrosive gases No corrosive gases Hydrocarbons and air Hydrocarbons and air Hydrocarbons and air Turbo Pump High compression Medium compression Medium compression ratio for hydrogen ratio for hydrogen ratio for hydrogen Features Dual diaphragm pump system Single diaphragm pump Low roughing pressure (<0.3 mbar) Higher roughing pressure (2 mbar) Ion gauge monitors sample chamber Ion gauge optional Roughing line pressure monitor Roughing line pressure monitor only Sampling from several bar to 10-4 mbar Atmospheric sampling only Heated sample (100 |