Text preview for : testlog_5700a.txt part of Fluke testlog 5700a Fluke 5700a logs initial_28_03_2016 testlog_5700a.txt
Back to : testlog_5700a.txt | Home
Fluke 5700A - detected, S/N XXXXXXX, Version: JA*
This is Series I unit
-i- Reading initial calibration data from 5700A
Unit last calibrated : 1308 days ago
Unit running time : 111438 hr
CAL CONST 6.5V reference voltage : +6.891366504559953E+00
CAL CONST 13V reference voltage : +1.379482200889601E+01
CAL CONST 10KOHM standard resistance : +1.000044087249608E+04
CAL CONST, Zero calibration temperature : +2.245000076293945E+01
CAL CONST, All calibration temp : +2.245000076293945E+01
Calibrator UNDER REMOTE CONTROL
Calibrator STABLE (settled within spec)
Calibrator is cooking report over SERIAL
User string PUD : #2055700A
Fatal errors history: "
3/17/93 8:18:20 Fault 230: 5725 RESET (power-up or watchdog timer)
3/30/93 7:07:49 Fault 4014: Sequencer Timed Out Waiting for 5725
10/05/94 18:53:17 Fault 218: 5725 +400V Supply Too Large
10/14/94 20:32:02 Fault 218: 5725 +400V Supply Too Large
11/07/94 12:27:47 Fault 218: 5725 +400V Supply Too Large
11/07/94 12:28:06 Fault 218: 5725 +400V Supply Too Large
1/06/95 7:56:16 Fault 218: 5725 +400V Supply Too Large
1/06/95 7:56:39 Fault 218: 5725 +400V Supply Too Large
1/06/95
12:34:42 Fault 218: 5725 +400V Supply Too Large
1/25/95 7:59:37 Fault 230: 5725 RESET (power-up or watchdog timer)
1/25/95 8:00:18 Fault 230: 5725 RESET (power-up or watchdog timer)
2/03/95 7:02:33 Fault 230: 5725 RESET (power-up or watchdog timer)
4/07/95 12:53:33 Fault 225: 5725 Fan Not Working
4/07/95 12:53:34 Fault 225: 5725 Fan Not Working
4/07/95 12:53:35 Fault 225: 5725 Fan Not Working
4/07/95 12:53:36 Fault 225: 5725 Fan Not Working
"
HP 3458A detected
TEMP? = 37.5 C
Fluke 5700A readback output = 10.000000000 V
10V test result: 1.000001492E+01 VDC [deviation 1.4920 ppm]
Program completed!