Keysight and Yokowo Co., Ltd. PENPROBE.CA RF Probe-test Solution up to 110GHz
PENPROBE. CA's Features Detachable Probe-unit
Recently, higher frequency RF test requirements are
PENPROBE.CA series embodies several favorable features All the Probe-units in the PENPROBE.CA are detachable
increasing rapidly based on the development of high
to support our customers. PENPROBE.CA consists of from the probe body. Each Probe-unit in Form A (30 degrees
speed semiconductor devices. 67GHz VNA with 500GHz
Probe-unit and probe body. touchdown angle) can be dismounted from the probe body
or 1THz extenders, already commercially available. To
and replaced by different contactor configuration Probe-unit
cope with these measurement equipment PENPROBE.CA's Probe-unit is detachable from its probe
among GSG, GS, SG with various pitch. For different
advancements, YOKOWO decided to provide new RF body, thus offering a low cost solution by using the probe
frequencies Probe-unit can also be replaced among 40GHz,
probe-test solutions for our customers. Let us introduce body for different frequency testing. For this purpose,
50GHz, 67GHz and 110GHz. Please note that Probe-unit
our product "PENPROBE.CA" below. Probe-unit is also available as a separate item.
of Form A (30 degrees touchdown angle) can't be mounted
Another feature is the installation easiness of PENPROBE.CA on the Form B (45 degrees touchdown angle) probe body ,
PENPROBE.CA thanks to the good coplanarity of plated Ni-contactors. This Form C (dual, 30 degrees) nor Form D (dual, 45degrees). This
PENPROBE.CA is designed for RF probe-test with coaxial advantage will become valuable especially in the GSGSG relationship is valid for other Form's Probe-unit as well.
cable system. Its probe-contactors are made of photo- type contactors.
lithographic plated Ni on the laminated polyimide film, The detachable probe-unit system offers a low cost test
Also a Bias-Tee can be used with the 110GHz use PEN-
thus realizing good contactor co-planarity, durability and solution to our customers. Once you buy Form A type PEN-
PROBE.CA, thus minimizing the insertion loss compared to
RF performance such as insertion loss/ return loss. PEN- PROBE.CA of a given part number, you can use its probe
the ordinary wide-band Bias-Tee. We'll expand this Bias-Tee
PROBE.CA has various frequency range models and probe- body by purchasing a different type Probe-unit for cost cut.
inclusion for the lower frequency PENPROBE.CA in future.
contactor configurations aiming for versatile RF probe- For this purpose, Probe-unit is also available as a separate
testing. item.
Contactor Coplanarity