Text preview for : HP 4263B Users Guide.pdf part of Agilent HP 4263B Users Guide Agilent HP 4263B Users Guide.pdf



Back to : HP 4263B Users Guide.pdf | Home

Agilent 4263B LCR Meter
User's Guide




Agilent Part No. 04263-90022
Printed in JAPAN December 2000


Fourth Edition
Documentation Map
User's Guide ( This book
Is a handy reference to help you to get started using your 4263B, basic measurements and
commonly used features are explained.
Operation Manual (Furnished with the 4263B.)

Privides information on initial inspection, how to operate the 4263B, in-depth reference
information, general information, and meintenance information.
In User's Guide.

Chapter 1, Preparation for Use
For initial turn ON of the 4263B
Chapter 2, Operating the 4263B
Basic Measurement operation
Getting acquired with the 4263B : for beginners
Handy reference for measurement task : for all users
Chapter 3, Measurement Example
Measurement examples for typical 4263B applications
Capacitor measurement
Inductor measurement
Transformer measurement


In the User's Guide, information on the following subjects is not discussed:

Initial Inspection Maintenance
GPIB Remote Control Speci cation
Handler Interface Error Messages

For detailed information on these subjects, refer to the Operation Manual.
Contents
1. Preparation for Use
In This Chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Power Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
To Set Power LINE Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
To Set Power LINE Frequency . . . . . . . . . . . . . . . . . . . . . . . . 1-3
2. Operating the 4263B
In This Chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Resetting 4263B to its Default Settings . . . . . . . . . . . . . . . . . . . . 2-2
Connecting Test Fixture . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Setting the Cable Length . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Selectting the Measurement Parameter . . . . . . . . . . . . . . . . . . . . 2-3
Setting the Test Signal Frequency . . . . . . . . . . . . . . . . . . . . . . . 2-4
Setting the Test Signal Level . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Setting the DC Bias Source Voltage . . . . . . . . . . . . . . . . . . . . . . 2-5
Selecting the Measurement Time Mode . . . . . . . . . . . . . . . . . . . . 2-5
Setting the Averaging Rate . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Selecting the Measurement Range . . . . . . . . . . . . . . . . . . . . . . . 2-6
Auto Range mode |Automatically Selecting the Optimum Measurement Range 2-6
Hold Range mode|Selecting the Measurement Range of Your Choice . . . . . 2-6
Selecting the Trigger Mode . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
Setting the Trigger Delay Time . . . . . . . . . . . . . . . . . . . . . . . . 2-7
Performing the OPEN Correction
|Canceling the stray admittance in parallel with the DUT . . . . . . . . . 2-8
Performing the SHORT Correction
|Canceling the residual impedance in series with the DUT . . . . . . . . . 2-9
Using the Comparator Function . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Setting the Limit Values . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Starting the Sort . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Using the Contact Check Function
|Monitoring the connection of test electrodes and DUT . . . . . . . . . . 2-11
Using the Deviation Measurement Function . . . . . . . . . . . . . . . . . . 2-11
Setting the Deviation Reference Values . . . . . . . . . . . . . . . . . . . 2-11
Selecting the Deviation Mode . . . . . . . . . . . . . . . . . . . . . . . . 2-12
Selecting the Display Mode . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
Using the Level Monitor Function . . . . . . . . . . . . . . . . . . . . . . . 2-14
Selecting the Beeper Mode . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
Setting the Printer|Printing the measurement data . . . . . . . . . . . . . . 2-15
Connecting the DUT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
Applying the DC Bias . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17
Making a Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17
Veri cation of Current Settings . . . . . . . . . . . . . . . . . . . . . . . . 2-17
If You Have a Problem . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-18
Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-19
Default Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-19
Measurement Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . 2-19

Contents-1
Accessories Available . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-20
16064B LED Display/Trigger Box . . . . . . . . . . . . . . . . . . . . . 2-20
Test Fixtures and Test Leads . . . . . . . . . . . . . . . . . . . . . . . 2-20
Accessories List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22
Measurement Range Setting . . . . . . . . . . . . . . . . . . . . . . . . 2-23
Other Topics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-23
3. Measurement Examples
In This Chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Electrolytic Capacitor Measurement|For High Capacitance . . . . . . . . . . . 3-2
DUT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Measurement Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Measurement Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
For More Information . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Inductor Measurement|Versatile measurement parameters . . . . . . . . . . . 3-6
DUT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Test Fixture . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Measurement Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Measurement Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
For More Information . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Transformer Measurement (Option 001 Only) . . . . . . . . . . . . . . . . . 3-11
DUT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Measurement Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Measurement Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
For More Information . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14




Contents-2
Figures
2-1. Connecting a Test Fixture . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
2-2. Printer Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
2-3. Connecting the DUT . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16




Tables
1-1. Line Voltage Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2




Contents-3
1
Preparation for Use
In This Chapter
First you must set the 4263B to match the available power LINE voltage, before turning the
4263B ON.
If the 4263B's power LINE voltage and frequency are properly set and ready to use, you can
skip this chapter.



Power Requirements
The 4263B's power source requirements are as follows:
LINE Voltage : 100 / 120 / 220 / 240 V ac (610%)
LINE Frequency : 47 to 66 Hz
Power Consumption : 45 VA maximum




To Set Power LINE Voltage
1. Con rm that the power cable is disconnected.
2. Slide the LINE Voltage selector on the rear panel to match the ac LINE voltage which will
be used (see Table 1-1).




Preparation for Use 1-1
4263B

Table 1-1. Line Voltage Selection

Voltage Selector Line Voltage Required Fuse

(a) 115 V 100 / 120 V T 0.5 A 250 V
(Agilent part number 2110-0202)
(b) 230 V 220 / 240 V T 0.25 A 250 V
(Agilent part number 2110-0201)




1-2 Preparation for Use
4263B


To Set Power LINE Frequency
Note In this manual, the BLUE shift key is expressed as , the top of the key is
not labeled \blue".


1. Connect the power cable to the power cord receptacle on the rear panel.
2. Push the LINE switch in and the 4263B will emit a beep when it turns ON, and the self
tests will be performed. (If any message is displayed, see \Error Messages" at the back of
Operation manual.) The 4263B will be ready for operation after a message like the following
is displayed.




3. Press . The following con guration menu is displayed.




4. Press until Line blinks, then press .




A blinking item means that it is currently selected.
5. If the setting does not match the ac line frequency, press to toggle the setting
between 50 HZ and 60 HZ.
6. Press to set the line frequency.



Preparation for Use 1-3
4263B

7. Exit the con gration menu by selecting Exit.


Note The power line frequency setting is stored and is not changed after reset or
power-o. Once you set it, you do not need to set the line frequency again as
long as the same power line frequency is being used.




1-4 Preparation for Use
2
Operating the 4263B
In This Chapter
Basic operation of the 4263B is explained.




Operating the 4263B 2-1
4263B


Resetting 4263B to its Default Settings
1. Press to select the reset menu.




2. Select Yes using or , and press .

The 4263B will be reset to its default settings. For more information about the default settings,
see \Default Settings" later in this chapter.

Connecting Test Fixture
Connect the test xture to the UNKNOWN terminals as follows:




Figure 2-1. Connecting a Test Fixture


See information on available test xtures, \Accessories Available" later in this chapter.




2-2 Operating the 4263B
4263B


Setting the Cable Length
The cable length correction function cancels the phase shift error caused by the cable length.
When using the Agilent test leads, perform the cable length correction as follows:
1. Press . Cable lengths 0 m, 1 m, 2 m, and 4 m will be displayed.




The blinking cable length is the current setting.
2. Select the desired cable length using or . To determine which length you should
select, see \Accessories Available" later in this chapter.
3. Press .




Selectting the Measurement Parameter
1. Press . The primary measurement parameters are displayed.




2. Select the desired primary parameter using or , and press .

3. Then the secondary parameters are displayed in the same manner as above.
The secondary parameters which can be selected dier depending on the primary parameter.
Refer to \Measurement Parameters" later in this chapter.




4. Select the desired secondary parameter using or , and press .




Operating the 4263B 2-3
4263B


Setting the Test Signal Frequency
Press . The test signal frequency selection menu is displayed.

Select the desired frequency using or , and press .

You can also select the test signal frequency by pressing until the desired frequency is
displayed.




Note that the 10 kHz and 20 kHz (Option 002 only) test frequency are not available when the
cable length setting is 4 m, and the 100 kHz test frequency is not available when the cable
length setting is 2 m or 4 m.


Setting the Test Signal Level
1. Press . The test signal level selection menu is displayed.

2. Enter the desired value using the numeric keys and the engineering key . For
example, to set the level to 245 mV, press ( or press
).

You can also set the value using or .




3. Press to set the test signal level.




2-4 Operating the 4263B
4263B


Setting the DC Bias Source Voltage
1. Press . The DC bias setting menu is displayed.




2. Select the desired DC bias voltage value using or , and press .

Now the DC bias source is selected. For how to apply the DC bias voltage, see \Applying the
DC Bias", later in this chapter.



Selecting the Measurement Time Mode
1. Press to select the measurement time mode (Short, Med or Long). The Meas Time
annunciator ( 9 ) displays the measurement time setting.



Setting the Averaging Rate
1. Press .




2. Enter the averaging rate using the numeric keys. You can enter integer values from 1 to
256. Also, you can increase or decrease the value using or .

3. Press to set the value and to exit.




Operating the 4263B 2-5
4263B


Selecting the Measurement Range
Auto Range mode
|Automatically Selecting the Optimum Measurement Range
Press . The 4263B's range mode is changed from \Hold" to \Auto", or from \Auto" to
\Hold". When the Hold Range annunciator( 9 ) is OFF, the 4263B is set to the auto range mode.




Hold Range mode|Selecting the Measurement Range of Your Choice
1. Press .




2. Press or until the desired range is displayed. Or, input the impedance value to
be measured using the numeric keys and the engineering key . The 4263B will select
the optimum measurement range setting.
3. Press . to set the measurement range.

The available ranges are 0.1
, 1
, 10
, 100
, 1 k
, 10 k
, 100 k
, and 1 M
. To determine
which measurement range you should select, see \Measurement Range Setting" later in this
chapter.




2-6 Operating the 4263B
4263B


Selecting the Trigger Mode
Press until the Trigger annunciator ( 9 ) points to the desired trigger mode (Int, Man or
Ext).




To trigger a measurement in each mode, see \Making a Measurement" later in this chapter.



Setting the Trigger Delay Time
1. Press .




2. Enter the desired trigger delay time using the numeric keys. (For example, to set 0.5 sec,
press .) You can set the trigger delay time from 0.000 sec to 9.999 sec.
You can also set the trigger delay time using or ;.

3. Press .




Operating the 4263B 2-7
4263B


Performing the OPEN Correction
|Canceling the stray admittance in parallel with the DUT
1. Con rm that the test xture is connected to the UNKNOWN terminals without a DUT
connected.
2. Press . The OPEN correction menu is displayed.




Press or until Openmeas blinks, and press . The OPEN correction is
performed. During that time, the following message is displayed.




After a while, the 4263B completes OPEN correction with the message Open Correction
Complete, and returns to the measurement state.
If \Out Of Limit", a WARNING message, is displayed, the OPEN admittance is so high that it
would be unsuitable for OPEN correction data. This is only a WARNING, the OPEN correction
data will still be used. However, you must verify the test xture connection to the UNKNOWN
terminals and the procedure used to perform the OPEN correction.
Verify that the test xture is correctly connected to the UNKNOWN terminal.
Verify that nothing is connected to the test xture's test electrode.
Perform OPEN correction again after verifying the above items.




2-8 Operating the 4263B
4263B


Performing the SHORT Correction
|Canceling the residual impedance in series with the DUT
1. Con gure the test electrodes in a SHORT con guration by connecting the High and Low
electrodes to each other or by connecting a shorting bar to the test xture.
2. Press . The SHORT correction menu is displayed.




3. Press or until Shortmeas blinks, and press . SHORT correction is
performed. During that time, the following message is displayed.




After a while, the 4263B completes SHORT correction with the message Short Correction
Complete, and returns to the measurement state.
If \Out Of Limit", a WARNING message, is displayed, the SHORT impedance is so high that
it would be unsuitable for SHORT correction data. This is only a WARNING, the SHORT
correction data will still be used. However, you must verify the test xture connection to the
UNKNOWN terminals and the procedure used to perform the SHORT correction.
Verify that the test xture is correctly connected to the UNKNOWN terminal.
Verify that the test xture's test electrodes are correctly shorted.
Perform SHORT correction again after verifying the above items.




Operating the 4263B 2-9
4263B


Using the Comparator Function
Setting the Limit Values
1. Press to set the lower limit of the primary parameter, and to set the higher
limit. Press to set the lower limit of the secondary parameter, and
to set the higher limit.
For example, the following menu is displayed when is pressed.




2. Enter the value using the numeric keys, then press to set the value. You can set the
value from 0999.99210 to 999.99210 .
14 14




Starting the Sort




3. Press . The Comptr On annunciator( 9 ) turns ON, and the comparator function is
turned ON.
The sorting results are HIGH, IN, and LOW.
Where,
HIGH greater than the higher limit
IN between the higher and lower limits
LOW less than the lower limit
The 4263B shows the comparison results using the display, beeper, printer, and
16064B LED Display/Trigger Box.
For result output to the display, see \Selecting the Display Mode" later in this chapter.
For result output to the beeper, see \Selecting the Beeper Mode" later in this chapter.
For result output to the printer, see \Setting the Printer|Printing the measurement data"
later in this chapter.
For result output to the 16064B, see \Accessories Available" later in this chapter.



2-10 Operating the 4263B
4263B


Using the Contact Check Function
|Monitoring the connection of test electrodes and DUT




Press , and the Cont Chk annunciator( 9 ) turns ON.

When the contact check result is Fail, the 4263B displays N.C. (No-Contact).

Using the Deviation Measurement Function
Setting the Deviation Reference Values
1. Press .




Press or until 1RefEnt blinks, and press .

2. You can now input the primary parameter's reference value. Enter the reference value with
the numeric keys, and press to set the value.




3. Then the 4263B displays the menu for setting the secondary parameter's reference value.
Enter the reference value with the numeric keys, and press to set the value.




Operating the 4263B 2-11
4263B


Selecting the Deviation Mode
4. The primary parameter's mode can be set by selecting Pri and pressing .




1ABS Measured value0Reference
1% (Measured value0Reference)/Reference 2100 %
Off Turning the deviation mode OFF.
Select the deviation measurement mode with or , and press .

5. The secondary parameter's mode can be set by selecting Sec and pressing in the same
manner as the primary. Select the desired mode with or , and press .




2-12 Operating the 4263B
4263B


Selecting the Display Mode
Press




Press or to select Data, Cmprtr, Digit or Off, and press .
The Measurement Display mode (Data) shows the measurement data:




The Comparison Display mode (Cmprtr) shows the comparison results:




When Digit is selected, the following menu is displayed, allowing you to set the number of
digits displayed for the measured value.




When Off is selected, the 4263B will not display the measurement results (Display OFF
mode).




Operating the 4263B 2-13
4263B


Using the Level Monitor Function
1. Press , and the following menu is displayed.




2. Select Off, Imon, Vmon or Both using .

When Imon is selected, the 4263B will monitor the actual signal current owing through
the DUT.
When Vmon is selected, the 4263B will monitor the actuall signal voltage across the DUT.
When Both is selected, the 4263B will monitor both current and voltage.
When Off is selected, the level monitor function is turned OFF.
3. Press .

You can see the level monitor values on the LCD display by pressing several times;.




2-14 Operating the 4263B
4263B


Selecting the Beeper Mode
To change the beeper mode for the comparator result reporting:
1. Press .




2. Select Beep using or , and press .




3. Select the beep mode using or , and press .
4. Select Exit using or , and press .




Setting the Printer|Printing the measurement data
1. Use an GPIB compatible printer, set to the Listen Always mode.
2. Connect the printer to the 4263B's GPIB port on the rear panel.
3. Turn the printer ON.
4. Set the 4263B to talk only mode (Set the 4263B's GPIB address to 31).
a. Press . Or press or to change the value.




b. Press . The Talk Only annunciator( 9 ) turns ON, and the printer begins printing the
measurement data.




Operating the 4263B 2-15
4263B




Figure 2-2. Printer Output


When you want to disable printing, change the GPIB address to an address other than 31 (for
example, 17, which is the default setting).



Connecting the DUT
Connect the DUT to the test electrodes.




Figure 2-3. Connecting the DUT




2-16 Operating the 4263B
4263B


Applying the DC Bias




Press to apply the DC bias. The DC Bias ON/OFF indicator is ON.

(Press again to turn OFF the DC bias. The DC Bias ON/OFF indicator is OFF.)




Making a Measurement
In the internal trigger mode| The 4263B makes continuous free-running measurements.
In the manual trigger mode| Press when you want to trigger a measurement.
In the external trigger mode| Connect the external trigger source to the EXT TRIGGER
terminal on the 4263B's rear panel, and apply a TTL level trigger signal to trigger a
measurement. (For details, see Operation Manual .)
Note that it must be set to the external trigger mode to trigger from an external handler or
the 16064B LED Display/Trigger Box.



Veri cation of Current Settings
Press .

The display on the right side of the LCD changes, and each time is pressed, the next
current setting is displayed.
1. Test signal frequency and Test signal level
2. DC bias setting and Averaging rate
3. Trigger delay time and Cable length
4. Comparator limits for primary parameter
5. Comparator limits for secondary parameter
6. Level monitor value




Operating the 4263B 2-17
4263B


If You Have a Problem
If any of the problems listed below occur, follow the instructions described.
If you nd yourself lost when operating the 4263B, you can get back on track by:
To return to the measurement mode, press several times.

To return to the default settings, press . (If the reset not accepted, con rm that
the Key Lock annunciator is turned ON. See next.)
If the 4263B does not accept key input:
Check whether or not the Key Lock annunciator( 9 ) is ON. If so,
Press . The Key Lock annunciator( 9 ) turns OFF and the front-panel keys are
unlocked.
Check that the 16064B LED display/trigger box is connected to the 4263B and it is set to
lock out the keys. If so, unlock the keys from the 16064B.
If the 4263B does not display measurement results:
The display mode is set to the Display OFF mode.
1. If the 4263B is in the key lockout mode, cancel the key lockout mode. (See previous
description.)
2. Press to change the display mode to a mode other than Display OFF.

If ------ or OVLD is displayed:
The measurement result is out of the measurable range. Check the DUT and make sure the
measurement range is properly set.




2-18 Operating the 4263B
4263B


Reference
Default Settings
Frequency : 1 kHz Trigger delay : 0 ms
Test voltage level : 1 Vrms Comparator : OFF
DC Bias : OFF Contact check : OFF
DC Bias source : 0V Display mode : Measurement mode
Measurement parameter : Cp-D Beep mode : FAIL mode
Deviation measurement : OFF Cable length :0m
Measurement range : Auto Display digits :5
Measurement time : MEDium Level monitor : OFF
Averaging rate : 1 OPEN/SHORT correction data is cleared
Trigger mode : Internal


Measurement Parameters
Primary Secondary Parameters Which
Parameter May Be Selected For Th is
Primary Parameter

Z Z : impedance ( absolute value )
Y
Y : admittance ( absolute value )
: phase angle
R X R : resistance
G B Ls : equivalent series inductance
Lp : equivalent parallel inductance
Cp D Q G Rp Cs : equivalent series capacitance
Cs D Q Rs Cp : equivalent parallel capacitance
Q : quality factor
Lp D Q G Rp Rdc
D : dissipation Factor
Ls D Q Rs Rdc G : conductance
L2 N 1/N M R2 X : reactance
B : susceptance
Rp : equivalent parallel resistance
Rs : equivalent series resistance
Rdc :dc resistance
N : turns ratio of transformer1
M :mutual inductance1
L2 : inductance1
R2 : dc resistance1




1 This para meter is measured using the transformer measurement con guration (two-terminal measurement
con guration).




Operating the 4263B 2-19
4263B


Note Primary parameter L2 and secondary parameters Rdc, N, M, and R2 can only
be used with Option 001 (N / M / DCR measurement function addition). These
parameters are not displayed on the menu if the 4263B is not equipped with
Option 001.
To measure the primary parameter L2, the transformer measurement
con guration is required. So use the 16060A Transformer Test Fixture.


Accessories Available
16064B LED Display/Trigger Box

The 16064B LED Display/Trigger Box triggers a measurement when its trigger key is pressed,
and displays the contact check and comparison results using LEDs. It allows you to manually
operate the comparator function of the 4263B.




Test Fixtures and Test Leads

For measurement versatility, various types of test xtures and test leads are available for the
4263B. In this section, main test xtures and test leads are described using gures. When using
these test xtures and test leads, set the 4263B to the corresponding cable length of the test
xture or test leads being used.




2-20 Operating the 4263B
4263B




Operating the 4263B 2-21
4263B

Accessories List

16034E Test Fixture (For SMD or Chip type DUT)
16047A Test Fixture (For Axial or Radial DUT)
16047B Test Fixture (For Axial or Radial DUT)
16047C HF Test Fixture (For Axial or Radial DUT)
16047D Test Fixture (For Axial or Radial DUT)
16048A Test Leads (1 m, BNC)
16048B Test Leads (1 m, SMC)
16048D Test Leads (2 m, BNC)
16048E Test Leads (4 m, BNC)
16060A Transformer Test Fixture
16065A External Bias Test Fixture
16065C External Bias Adapter
16085B Terminal Adapter: Converts 4 terminal pair connector to APC7 connector.
16089A Kelvin Clip Leads (1 m, two large clips)
16089B Kelvin Clip Leads (1 m, two medium clips)
16089C Kelvin Clip Leads (1 m, two IC clips)
16089D Alligator Clip Leads (1 m, four medium clips)
16089E Kelvin Clip Leads (1 m, two large clips)
16092A1 RF Spring Clip: Axial Radial and SMD
16093A1 RF Two-Terminal Binding Post
16093B1 RF Three-Terminal Binding Post
16094A1 2,
RF Probe Tip/Adapter
16095A3 LF Probe Adapter
16191A1 Side Electrode SMD Test Fixture
16192A1 Parallel Electrode SMD Test Fixture
16193A1 Small Side Electrode SMD Test Fixture
16194A1 Wide Temperature SMD Test Fixture
16314A 50
/4-Term Converter 100Hz-10MHz
16334A Test Fixture (For SMD or Chip type DUT)
16451B Dielectric Test Fixture
16452A Magnetic Test Fixture
16064B LED Display/Trigger Box (with GO/NO-GO display and trigger button)
1 16085B adapter required.
2 Cables adapted to APC7 on each end required.
3 Don't connect ground-lead to 4263B.

Note There is some possibility that available accessories are changed. Refer to latest
accessories catalogue about the latest information.




2-22 Operating the 4263B
4263B


Measurement Range Setting
Range Setting Optimum Measurement Range

0.1
1 jZj 100 m

1
100 m
< jZj 1

10
1
< jZj 10

100
10
< jZj < 1 k

1 k
1 k
jZj < 10 k

10 k
10 k
jZj < 100 k

100 k
2 100 k
jZj < 1 M

1 M
2 1 M
jZj < 10 M


1 This range is available when the test level settings is higher than
315 mV.
2 This range is not available for the 100 kHz test frequency
setting.




Other Topics
For details on these functions, see the Operation Manual.
Initial Inspection | Chapter 1 of the Operation Manual
Load correction | Chapter 1, Chapter 3 and Chapter 7 of the Operation Manual
Key Lock Function | Chapter 2 and Chapter 3 of the Operation Manual
GPIB | Chapter 4 and Chapter 5 of the Operation Manual
Handler Interface | Chapter 3, Chapter 6, and Appendix B of the Operation Manual
Save / Recall | Chapter 2 and Chapter 3 of the Operation Manual
Backup Function | Chapter 3 of the Operation Manual
Speci cation | Chapter 8 of the Operation Manual
Maintenance | Chapter 9 of the Operation Manual
Error Messages | \Error Messages" in back of the Operation Manual




Operating the 4263B 2-23
3
Measurement Examples
In This Chapter
This chapter provides the typical measurement examples




Measurement Examples 3-1
4263B

Electrolytic Capacitor Measurement|For High Capacitance
The 4263B's measurement accuracy and wide measurement range are the right tools to make
precise measurements of electrolytic capacitor parameters.
Electrolytic capacitors are generally high capacitance, so their impedance is low. The 4263B has
the 100 m
measurement range, and keeps its high measurement accuracy when measuring low
impedance. For example, the 4263B measures an aluminum electrolytic capacitor, 22,000 F, at
the test frequency of 120 Hz, with about 0.5 % accuracy. You can try this measurement using
the following procedure.

DUT
Aluminum electrolytic capacitor ( 22,000 F 6 20 % )
Requirements
Test Fixture : 16089B Kelvin Clip Leads

Measurement Setup
Measurement parameter : Cs-D1
Test frequency : 120 Hz
Test signal level : 1 Vrms2
1 For high capacitance measurement, equivalent series
parameter Cs-D is commonly used.
2 Even though LVL:1 Vrms is set, less than 0.5Vrms
which is maximum appliable volatge for testing
electrolytic capacitors in accordance to JIS standard,
will be applied to the Aluminum electrolytic capacitor
under test.


Measurement Procedure


1. Reset the 4263B.
a. Press .




b. Press until Yes blinks, and press .




3-2 Measurement Examples
4263B

2. Connect test xture to the UNKNOWN terminals as follows.




3. Set the cable length to 1 m.
a. Press .




b. Select 1 m using or , and press .

4. Set measurement parameter to Cs-D.
a. Press , and the following menu is displayed.




b. Select Cs using or , and press .




c. Select D using or , and press .




Measurement Examples 3-3
4263B

5. Set the test frequency to 120 Hz.
Press .

Press or to set the frequency to 120Hz, and press .

6. Perform the OPEN correction.
a. Separate the test lead clips ( Nothing must be connected to the test lead clips).
b. Press and the following menu is displayed.




c. Select OpenMeas using or , and press .




After a while, the OPEN correction is completed. (If Out Of
Limit is displayed, see \Performing the OPEN Correction
|Canceling the stray admittance in parallel with the DUT" in Chapter 2.)
7. Perform the SHORT correction.
a. Short the electrodes of the test xture as shown in the following gure.




3-4 Measurement Examples
4263B


b. Press , and the following menu is displayed.




c. Select ShortMeas using or , and press .




After a while, the SHORT correction is completed.(If Out Of
Limit is displayed, see \Performing the SHORT Correction
|Canceling the residual impedance in series with the DUT" in Chapter 2.)
8. Connect the DUT to the test xture, and the measurement result will be displayed.




For More Information
To apply the DC bias | See \Setting the DC Bias Source Voltage" in Chapter 2.
To print out the measurement result | See \Setting the Printer|Printing the measurement
data" in Chapter 2




Measurement Examples 3-5
4263B


Inductor Measurement|Versatile measurement parameters
The 4263B oers many kinds of measurement parameters for LCR measurement. In addition to
these parameters, Option 001 adds ability to make turns ratio (N), mutual inductance (M), DC
resistance (DCR) measurements.
This example shows a basic measurement for an inductor, and its DCR. You can measure both
inductance and DCR without resetting the measurement con guration.

DUT
Coil ( 220 H 6 5 % @ 100kHz )

Requirements
Test Fixture : 16047A
Test Fixture
16047A

Measurement Setup
Measurement parameter : Lp-Q and Lp-Rdc
Test frequency : 100 kHz
Test signal level : 100 mVrms

Measurement Procedure


1. Reset the 4263B.
a. Press .




b. Press until Yes blinks, and press .

2. Connect the test xture to the UNKNOWN terminals.




3-6 Measurement Examples
4263B




3. Select measurement parameter Lp-Q.
a. Press and the following menu is displayed.




b. Select Lp using or , and press .




c. Select Q using or , and press .

4. Set the test signal frequency to 100 kHz.
Press .

Set the frequency to 100 kHz using or , and press .




Measurement Examples 3-7
4263B

5. Set the test signal level to 100 mV.
Press .

Set the level to 100 mV using the numeric keys or , and press

6. Perform the OPEN correction.
a. Remove any device inserted in the test electrodes to create an OPEN condition (Nothing
should be connected to the test electrodes).
b. Press , and the following menu is displayed.




c. Select OpenMeas using or , and press .




After a while, the OPEN correction is completed. (If Out Of
Limit is displayed, see \Performing the OPEN Correction
|Canceling the stray admittance in parallel with the DUT" in Chapter 2.)
7. Perform the SHORT correction.
a. Insert the shorting plate to the test xture as shown in the following gure:




3-8 Measurement Examples
4263B


b. Press , and the following menu is displayed.




c. Select ShortMeas using or , and press .




After a while, the SHORT correction is completed. (If Out Of
Limit is displayed, see \Performing the SHORT Correction
|Canceling the residual impedance in series with the DUT" in Chapter 2.)
8. Connect the DUT to the test xture and the measurement result will be displayed.




Note Step 9 is for an 4263B with Option 001 only.



9. Change the measurement parameter to Lp-Rdc.
a. Press and the following menu is displayed.




b. Press to select Lp.




Measurement Examples 3-9
4263B


c. Select Rdc using or , and press . The measurement result will be
displayed.




For More Information
To select other measurement parameters | See \Selectting the Measurement Parameter" in
Chapter 2.
To print out the measurement result | See \Setting the Printer|Printing the measurement
data" in Chapter 2




3-10 Measurement Examples
4263B


Transformer Measurement (Option 001 Only)
With the 4263B's ability to measure turns ratio (N), mutual inductance (M), and DC resistance
(DCR), transformer-parameter calculations are no longer time-consuming tasks. Moreover
the 16060A Transformer Test Fixture makes it easy to setup transformer measurement
con gurations.
The following example shows how easy it is to measure turns ratio (N), mutual inductance (M),
and dc resistance (DCR) measurement of transformer.

DUT
Transformer ( 1 : 8 )

Requirements
Test Fixture : 16060A Transformer Test Fixture

Measurement Setup
Measurement : L2-N and L2-R2
parameter
Test frequency : 100 kHz
Test signal Level : 100 mVrms

Measurement Procedure


1. Reset the 4263B.
a. Press .




b. Press until Yes blinks, and press .




Measurement Examples 3-11
4263B

2. Connect the test xture to the UNKNOWN terminals.




3. Set the measurement parameter to L2-N.
a. Press and the following is displayed.




b. Select L2 using or , and press .




c. Select N using or , and press .

4. Set the test frequency to 100 kHz.
Press .

Set the frequency to 100 kHz using , and press .

5. Set the test signal level to 100 mVrms.
Press .

Set the level to 100 mV using the numeric keys or , and press .




3-12 Measurement Examples
4263B

6. Perform the OPEN correction.
a. Short the red clips together and short the black clips together, then separate the shorted
red and black sets of clips from each other. ( See the following gure.)




b. Press , and the following menu is displayed.




c. Select OpenMeas using or , and press .




After a while, the OPEN correction is completed. (If OUT OF
LIMIT is displayed, see \Performing the OPEN Correction
|Canceling the stray admittance in parallel with the DUT" in Chapter 2.)

Note Do not use the SHORT correction function of the 4263B when the L2-N,
L2-1/N, L2-M, or L2-R2 measurement parameters are selected.




Measurement Examples 3-13
4263B


7. Connect the DUT to the test xture and the measurement result will be displayed.




Set the switch to the opposite position if the 4263B displays OVLD as the measured value
of N. The 4263B cannot measure a value of N less than 0.9, and OVLD means that the
measurement result is out of range.
The leading sign of N indicates the polarity of transformer as follows:




For More Information
To select other parameters | You can measure L2-M (mutual inductance) and L2-R2 (dc
resistance) without changing the measurement con guration. To change the measurement
parameter, see \Selectting the Measurement Parameter" in Chapter 2.
To print out the measurement result | See \Setting the Printer|Printing the measurement
data" in Chapter 2




3-14 Measurement Examples