Text preview for : HP E1418A USER.pdf part of Agilent HP E1418A USER Agilent HP E1418A USER.pdf



Back to : HP E1418A USER.pdf | Home

Agilent 75000 Series C




Agilent E1418A
8/16-Channel D/A Converter Module
User's Manual and SCPI Programming Guide


Where to Find it - Online and Printed Information:
System installation (hardware/software)............. VXIbus Configuration Guide*
Agilent VIC (VXI installation software)*

Module configuration and wiring........................ This Manual
SCPI programming.............................................. This Manual
SCPI example programs...................................... This Manual
SCPI command reference ................................... This Manual
Register-Based Programming ............................. This Manual

VXIplug&play programming ............................. VXIplug&play Online Help
VXIplug&play example programs...................... VXIplug&play Online Help
VXIplug&play function reference ...................... VXIplug&play Online Help
Soft Front Panel information............................... VXIplug&play Online Help

VISA language information ................................ Agilent VISA User's Guide

Agilent VEE programming information ............. Agilent VEE User's Manual

*Supplied with Agilent Command Modules, Embedded Controllers, and VXLink.




*E1418-90002*
Manual Part Number: E1418-90002
Printed in Malaysia E0706
Contents
Agilent E1418A User's Manual

Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Safety Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
WARNINGS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
User Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

Chapter 1. Module Setup and Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Using This Chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Module Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Ordering Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Field Kits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Terminal Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Functional Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Front Panel Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Setting the Logical Address Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Module Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Installation in a Mainframe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Terminal Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Screw-Type Terminal Module (Standard) . . . . . . . . . . . . . . . . . . . . . . 22
Wiring the Terminal Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Attaching the Terminal Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Removing the Terminal Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Terminal Module Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Option A3E . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Option A3H . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Option 135 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Terminal Module Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Configuring the Terminal Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
P/J Jumper . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
V/I Jumper . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Terminal Module Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Voltage and Current Output Connections . . . . . . . . . . . . . . . . . . . . . . 33
Combining Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Connecting an External Trigger Source . . . . . . . . . . . . . . . . . . . . . . . 35
Using the CAL Output Terminals . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Initial Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Device Driver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Module Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38




Agilent E1418A User's Manual Contents 1
Chapter 2. Programming Examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Using This Chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Program Examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Querying Module Identification and Configuration . . . . . . . . . . . . . . . . . . . . 40
Configuration Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
APPLy Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Error Checking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
I/O Error Checking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
SCPI Error Checking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
SOURce Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Setting the Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Controlling the Output Relay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Triggering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Using *OPC? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Combining Output Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57

Chapter 3. Agilent E1418A SCPI Command Reference . . . . . . . . . . . . . . . . . . . 59
Using This Chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Command Types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Common Command Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
SCPI Command Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Linking Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
SCPI Command Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
ABORt . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
APPLy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
APPLyn:CURRent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
APPLyn:VOLTage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
CALibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
CALibration:CONDition? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
CALibration:CONFigure:RESistance? . . . . . . . . . . . . . . . . . . . . . . . . 68
CALibration:RESet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
CALibration:STORe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
CALibration:TEST? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
CALibration:VALue:RESistance . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
CALibrationn:CONFigure:CURRent? . . . . . . . . . . . . . . . . . . . . . . . . 71
CALibrationn:CONFigure:VOLTage? . . . . . . . . . . . . . . . . . . . . . . . . 72
CALibrationn:VALue:CURRent? . . . . . . . . . . . . . . . . . . . . . . . . . . 73
CALibrationn:VALue:VOLTage? . . . . . . . . . . . . . . . . . . . . . . . . . . 74
DIAGnostic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
DIAGnostic:CALibration:OUTPut[:MODE] . . . . . . . . . . . . . . . . . . . . 76
DIAGnostic:CALibration:OUTPut[:MODE]? . . . . . . . . . . . . . . . . . . . . 77
DIAGnostic:CONFigure? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
DIAGnostic:OUTPut:ALL[:STATe] . . . . . . . . . . . . . . . . . . . . . . . . . 78
DIAGnostic:SOURcen:FUNCtion:MODE . . . . . . . . . . . . . . . . . . . . . . 78
DIAGnostic:TRIGgered? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
INITiate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
INITiate[:IMMediate] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80




2 Agilent E1418A User's Manual Contents
Chapter 3. Agilent E1418A SCPI Command Reference (continued)
OUTPut . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
OUTPut:TTLTrgn[:STATe] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
OUTPut:TTLTrgn[:STATe]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
OUTPutn[:STATe] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
OUTPutn[:STATe]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
SOURce . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
SOURcen:CURRent[:LEVel][:IMMediate][:AMPLitude] . . . . . . . . . . . . . 85
SOURcen:CURRent[:LEVel][:IMMediate][:AMPLitude]? . . . . . . . . . . . . . 86
SOURcen:CURRent[:LEVel]:TRIGgered[:AMPLitude] . . . . . . . . . . . . . . 87
SOURcen:CURRent[:LEVel]:TRIGgered[:AMPLitude]? . . . . . . . . . . . . . . 88
SOURcen:FUNCtion:MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
SOURcen:FUNCtion:MODE? . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
SOURcen:VOLTage[:LEVel][:IMMediate][:AMPLitude] . . . . . . . . . . . . . 91
SOURcen:VOLTage[:LEVel][:IMMediate][:AMPLitude]? . . . . . . . . . . . . . 92
SOURcen:VOLTage[:LEVel]:TRIGgered[:AMPLitude] . . . . . . . . . . . . . . 93
SOURcen:VOLTage[:LEVel]:TRIGgered[:AMPLitude]? . . . . . . . . . . . . . . 94
STATus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
STATus:OPERation:CONDition? . . . . . . . . . . . . . . . . . . . . . . . . . . 96
STATus:OPERation:ENABle . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
STATus:OPERation:ENABle? . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
STATus:OPERation[:EVENt]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
STATus:PRESet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
STATus:QUEStionable:CONDition? . . . . . . . . . . . . . . . . . . . . . . . . 98
STATus:QUEStionable:ENABle . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
STATus:QUEStionable:ENABle? . . . . . . . . . . . . . . . . . . . . . . . . . . 99
STATus:QUEStionable[:EVENt]? . . . . . . . . . . . . . . . . . . . . . . . . . . 99
SYSTem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
SYSTem:ERRor? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
SYSTem:VERSion? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
TEST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
TEST:NUMBer? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
TEST:TST[:RESults]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
*TST? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
TRIGger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
TRIGger[:IMMediate] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
TRIGger:SOURce . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
TRIGger:SOURce? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
IEEE-488.2 Common Command Quick Reference . . . . . . . . . . . . . . . . . . . . 106
Agilent E1418A Command Quick Reference . . . . . . . . . . . . . . . . . . . . . . . 107

Appendix A. Agilent E1418A Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . 111
DC Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
DC Current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
General Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112
Power Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112
Cooling Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112


Agilent E1418A User's Manual Contents 3
Appendix B. Agilent E1418A Register-Based Programming . . . . . . . . . . . . . . . . . 113
Addressing the Registers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
A16 Registers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
A24 Registers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
The Base Address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
Register Offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118
Reset and Registers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
Register Maps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
A16/A24 REGISTERS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
A24 REGISTERS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
Register Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
Manufacturer ID Register . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
Logical Address Register . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
Device Type Register . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
VXI Status/Control Register . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
VXI Offset Register . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128
Calibration Control Register . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129
Card Configuration Register . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130
Software Trigger Register . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131
Trigger Control Register . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131
Interrupt Status Register . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133
Isolation Status Register . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133
Channel Program Jumper Register . . . . . . . . . . . . . . . . . . . . . . . . . . 134
Channel Trigger Register . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135
Channel Mode Register . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 136
Channel Relay Control Register . . . . . . . . . . . . . . . . . . . . . . . . . . . 136
Card Control Register . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137
Main_DAC Immediate Registers . . . . . . . . . . . . . . . . . . . . . . . . . . . 138
Main_DAC Triggered Registers . . . . . . . . . . . . . . . . . . . . . . . . . . . 138
Offset_DAC Registers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
Gain_DAC Registers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
Undefined Registers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
Channel Voltage Offset Calibration Registers . . . . . . . . . . . . . . . . . . . . 140
Channel Voltage Gain Calibration Registers . . . . . . . . . . . . . . . . . . . . . 141
Channel Current Offset Calibration Registers . . . . . . . . . . . . . . . . . . . . 142
Channel Current Gain Calibration Registers . . . . . . . . . . . . . . . . . . . . . 143
Calibration Resistor Value Registers . . . . . . . . . . . . . . . . . . . . . . . . . 144
Voltage Calibration Status Register . . . . . . . . . . . . . . . . . . . . . . . . . 144
Current Calibration Status Register . . . . . . . . . . . . . . . . . . . . . . . . . . 145
Calibration Isolation Status Register . . . . . . . . . . . . . . . . . . . . . . . . . 146
Calibration Card Configuration Register . . . . . . . . . . . . . . . . . . . . . . . 147
Calibration Checksum Register . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147
Module Serial Number Registers . . . . . . . . . . . . . . . . . . . . . . . . . . . 148
Register Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 148




4 Agilent E1418A User's Manual Contents
Appendix C. Agilent E1418A Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . 151
Error Types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151
Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 152

Appendix D. Voltage/Current Output Adjustment . . . . . . . . . . . . . . . . . . . . . . 155
Using This Appendix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155
Calibration Constants and Non-Volatile Memory . . . . . . . . . . . . . . . . . . . . 156
Equipment Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 156
Making Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 156
Adjustment Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 157
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 157
Voltage Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 158
Current Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 160
Storing the Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163
Return the Module to Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163
Example Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163

Appendix E. Configuration and Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . 167
Using This Appendix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 167
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 167
Checking Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 167
Changing Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 168
Adding 8 Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 169
Installing Isolated/Non-Isolated Plug-on Modules . . . . . . . . . . . . . . . . . . 172
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174
Isolating an Assembly (Self-Test) . . . . . . . . . . . . . . . . . . . . . . . . . . 174
Assembly Exchange . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175
To Exchange an Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175

Index. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177




Agilent E1418A User's Manual Contents 5
Notes




6 Agilent E1418A User's Manual Contents
Certification
Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. Agilent
Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and
Technology (formerly National Bureau of Standards), to the extent allowed by that organization's calibration facility, and to the calibration
facilities of other International Standards Organization members.

Warranty
This Agilent Technologies product is warranted against defects in materials and workmanship for a period of one (1) year from date of
shipment. Duration and conditions of warranty for this product may be superseded when the product is integrated into (becomes a part
of) other Agilent products. During the warranty period, Agilent Technologies will, at its option, either repair or replace products which
prove to be defective.
For warranty service or repair, this product must be returned to a service facility designated by Agilent Technologies. Buyer shall prepay
shipping charges to Agilent and Agilent shall pay shipping charges to return the product to Buyer. However, Buyer shall pay all shipping
charges, duties, and taxes for products returned to Agilent from another country.
Agilent warrants that its software and firmware designated by Agilent for use with a product will execute its programming instructions
when properly installed on that product. Agilent does not warrant that the operation of the product, or software, or firmware will be
uninterrupted or error free.
Limitation Of Warranty
The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by Buyer, Buyer-supplied products
or interfacing, unauthorized modification or misuse, operation outside of the environmental specifications for the product, or improper site
preparation or maintenance.
The design and implementation of any circuit on this product is the sole responsibility of the Buyer. Agilent does not warrant the Buyer's
circuitry or malfunctions of Agilent products that result from the Buyer's circuitry. In addition, Agilent does not warrant any damage that
occurs as a result of the Buyer's circuit or any defects that result from Buyer-supplied products.
NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. Agilent SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES
OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Exclusive Remedies
THE REMEDIES PROVIDED HEREIN ARE BUYER'S SOLE AND EXCLUSIVE REMEDIES. Agilent SHALL NOT BE LIABLE
FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER BASED ON CON-
TRACT, TORT, OR ANY OTHER LEGAL THEORY.

Notice
The information contained in this document is subject to change without notice. Agilent Technologies MAKES NO WARRANTY OF
ANY KIND WITH REGARD TO THIS MATERIAL, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. Agilent shall not be liable for errors contained herein or for
incidental or consequential damages in connection with the furnishing, performance or use of this material. This document contains
proprietary information which is protected by copyright. All rights are reserved. No part of this document may be photocopied, reproduced,
or translated to another language without the prior written consent of Agilent Technologies, Inc. Agilent assumes no responsibility for the
use or reliability of its software on equipment that is not furnished by Agilent.

U.S. Government Restricted Rights
The Software and Documentation have been developed entirely at private expense. They are delivered and licensed as "commercial
computer software" as defined in DFARS 252.227- 7013 (Oct 1988), DFARS 252.211-7015 (May 1991) or DFARS 252.227-7014 (Jun
1995), as a "commercial item" as defined in FAR 2.101(a), or as "Restricted computer software" as defined in FAR 52.227-19 (Jun 1987)(or
any equivalent agency regulation or contract clause), whichever is applicable. You have only those rights provided for such Software and
Documentation by the applicable FAR or DFARS clause or the Agilent standard software agreement for the product involved.

Agilent E1418A 8/16 Channel D/A Converter Module User's Manual
Edition 2 Rev 2




Copyright