E5061B-3L5 LF-RF
Network Analyzer with Option 005
Impedance Analysis Function
Data Sheet
E5061B-3L5 with Option 005; NA plus ZA in one box
To ensure the performance and E5061B-3L5 block diagram
reliability of electronic equipment,
it is crucial to evaluate impedance
characteristics of various electronic Zin = 1 M/50
components used in the circuits. The ATT = 20 dB/0 dB
E5061B-005 is the impedance analy-
sis firmware option for the E5061B-
3L5 ENA Series LF-RF network
analyzer. The E5061B-005 combines
DC bias
network analysis (NA) and impedance T R source
analysis (ZA) functions in a single
instrument with comprehensive
component and circuit characteriza- ATT ATT
tion in a broad frequency range, R1 R2
5 Hz to 3 GHz.
Zin Zin T1 T2
E5061BEP-NZA ENA network
analyzer express configuration T R LF OUT Port-1 Port-2
is equivalent to the E5061B with
options 3L5/020/005/720/1E5. For
more information about the Express Gain-phase test port S-parameter test port
(5 Hz to 30 MHz) (5 Hz to 3 GHz, 50 )
ENA, visit (
www.agilent.com/find/
express-e5061b)
Key measurement functions of E5061B-3L5 with Option 005
Impedance analysis functions
Impedance parameters IZI, z , IYI, y, Cp, Cs, Lp, Ls, Rp, Rs, D, Q, R, X, G, B
Measurement methods Port 1 Reflection, Port 2 Reflection, Port 1-2 Series-thru, Port 1-2 Shunt-thru,
Gain-phase Series-thru, Gain-phase Shunt-thru
Equivalent circuit analysis 3-component model (4 models: A/B/C/D), 4-component model (1 model: E)
Source characteristics
Test frequency 5 Hz to 3 GHz (Port-1 and 2), 5 Hz to 30 MHz (Gain-Phase test port), 1 mHz resolution
AC source power (Osc level)