Text preview for : 5328a.pdf part of Agilent 5328a Agilent 5328a.pdf



Back to : 5328a.pdf | Home

TM 11-6625-2941-14&P


TECHNICAL MANUAL




OPERATOR'S, ORGANIZATIONAL, DIRECT SUPPORT,
AND GENERAL SUPPORT MAINTENANCE MANUAL
INCLUDING REPAIR PARTS AND SPECIAL TOOLS LIST
FOR
COUNTER, ELECTRONIC, DIGITAL READOUT
AN/USM-459
(HEWLETT-PACKARD MODEL 5328A/E42)
(NSN 6625-01-061-8928)




HEADQUARTERS, DEPARTMENT OF THE ARMY

MARCH 1979
This manual contains copyright material reproduced by permission of the
Hewlett-Packard Company.
TM 11-6625-2941-14 + P

Technical Manual HEADQUARTERS,
DEPARTMENT OF THE ARMY
No. 11-6625-2941-14 + P Washington, DC 28 March 1979

OPERATOR'S, ORGANIZATIONAL, DIRECT SUPPORT, AND
GENERAL SUPPORT MAINTENANCE MANUAL INCLUDING
REPAIR PARTS AND SPECIAL TOOLS LIST
FOR
COUNTER, ELECTRONIC, DIGITAL READOUT
AN/USM-459
(HEWLETT-PACKARD MODEL 5328A/E42)
(NSN 6625-01-061-8928)


REPORTING OF ERRORS

You can improve this manual by recommending improvements using
DA Form 2028-2 located in the back of the manual. simply
tear out the self-addressed form, fill it out as shown on the
sample, fold it where shown, and drop it in the mail.

If there are no blank DA Forms 2028-2 in the back of
your manual, use the standard DA Form 2028 (Recommended Changes to
Publications and Blank Forms) and forward to the Commander, US
Army Communications and Electronics Materiel Readiness Command,
ATTN: DRSEL-ME-MQ, Fort Monmouth, New Jersey 07703.

In either case a reply will be furnished direct to you.


T h i s m a n u a l i s a n a u t h e n t i c a t i o n o f t h e m a n u f a c t u r e rs c o m m e r c i a l l i t e r a t u r e
which, through usage, has been found to cover the data required to operate
a n d m a i n t a i n t h i s e q u i p m e n t . The manual was not prepared in accordance with
military specifications; therefore, the format has not been structured to
consider categories of maintenance.

TABLE OF CONTENTS

Section Page

0 INTRODUCTION
0-1 Scope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-1
0-2 Indexes of Publications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-1
0-3 Forms and Records . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-1
0-4 Reporting Equipment Improvement Recommendations (EIR) . . 0-1
0-5 Administrative Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-1
0-6 Destruction of Army Electronics Material . . . . . . . . . . . . . . . . . . . . . . . 0-1




i/ii (blank)
Model 5328A
Table of Contents


TABLE OF CONTENTS

Section Page

I GENERAL INFORMATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1-1. Scope of Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1-4. Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1-6. Instrument Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1-8. Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
1-16. Equipment Supplied and Accessories Available . . . . . . . . . . . . . . 1-2
1-18. Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2

II INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2-1. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2-3. Unpacking and Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2-5. Preparation for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2-6. Power Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2-8. Fuse Replacement and Installation . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2-10. Power Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2-13. Operating Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
2-15. Bench Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
2-17. Rack Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
2-19. Packaging for Shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
2-20. Original Packaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
2-24. Other Packaging Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
2-26. Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3

Ill OPERATION (OPERATORS INSTRUCTIONS) . . . . . . . . . . . . . . . . . . . . . . 3-1
3-1. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3-3. Frequency Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3-9. Period Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
3-13. Time Interval Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
3-20. Ratio Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
3-23. Operating Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
3-26. Function of Controls, Indicators, Inputs, and Outputs . . . . . . . . 3-8
3-27. Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
3-29. Power (Line) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
3-31. Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
3-33. Sample Rate Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
3-35. Arming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
3-37. Frequency Resoluton, N Switch . . . . . . . . . . . . . . . . . . . . . . . . 3-9
3-41. Input Channel Section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
3-44. A and B Channel Signal Conditioning . . . . . . . . . . . . . . . . . . . 3-10
3-49. Channel C Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
3-51. "C'' Channel Overload Indicator . . . . . . . . . . . . . . . . . . . . . . . . 3-12
3-53. Hysteresis Band of Trigger Levels . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
3-57. External Frequency Standard Input. . . . . . . . . . . . . . . . . . . . . . . . 3-12
3-59. M a r k e r O u t p .u. t. s . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . 3-13
3-61. Gate/Marker Out . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
3-63. 1 MHz and 10 MHz Frequency Standard Outputs . . . . . . . . . . 3-13
3-65. Trigger Lights . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
3-68. Programming Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
3-71. Setting Add dress Switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
3-75. Measurement Output Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
3-77. Bus Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
3-79. Programming Examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18

iii
Model 5328A
Table of Contents


TABLE OF CONTENTS (Continued)

Section Page

IV THEORY OF OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4-1. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4-3. Overall Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4-5. Basic Counter Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4-7. Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4-15. Period . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
4-18. Ratio . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
4-20. Time Interval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
4-22. Time interval, Resolution, and Averaging Techniques . . . . . . . 4-3
4-24. Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
4-27. Time Interval Averaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
4-34. Sources of Measurement Error . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
4-36.