Text preview for : Resistance_253B DC Current_253B AC Current_253B and Frequency and Period Measurement Errors in DMMs part of Agilent Resistance 253B DC Current 253B AC Current 253B and Frequency and Period Measurement Errors in DMMs Agilent Resistance_253B DC Current_253B AC Current_253B and Frequency and Period Measurement Errors in DMMs 5988-5512EN [4].pdf



Back to : Resistance_253B DC Curren | Home

Resistance; DC Current; AC Current; and
Frequency and Period Measurement
Errors in Digital Multimeters
Application Note




Introduction
When you make measurements with For a discussion of system cabling
a digital multimeter (DMM), common errors and DC voltage measurement
errors will occur. The following errors, see application note part num-
discussion will help you eliminate ber 5988-5511EN. For an overview of
potential measurement errors and AC voltage measurement errors, see
achieve the greatest accuracy with a application note part number 5988-
DMM. This paper covers resistance; 5513EN. (Note: The Agilent 34461A,
DC current; AC current; and frequency a 6