Text preview for : N5242-90004 Installation Note_252C Extd. Power Range & Bias Tees Upgrade Kit to Upgrade N5242A Opt 4 part of Agilent N5242-90004 Installation Note 252C Extd. Power Range & Bias Tees Upgrade Kit to Upgrade N5242A Opt 4 Agilent N5242-90004 Installation Note_252C Extd. Power Range & Bias Tees Upgrade Kit to Upgrade N5242A Opt 400 to Opt 419 [34].pdf



Back to : N5242-90004 Installation | Home

Keysight Extended Power
Range and Bias Tees
Upgrade Kit

To Upgrade PNA-X N5241A, N5242A or N5249A
Option 400 to Option 419
Upgrade Kit Order Numbers: N5241AU-926,
N5242AU-926 and N5249AU-926
Keysight Kit Number: N5242-60103





Installation Note
Installation Note N5242-90004 1
WARRANTY STATEMENT
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED "AS IS," AND IS SUBJECT TO BEING
CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY
APPLICABLE LAW, KEYSIGHT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD
TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE
IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT
SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION
WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED
HEREIN. SHOULD KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY
TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE
WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.




DFARS/Restricted Rights Notice
If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is
delivered and licensed as "Commercial computer software" as defined in DFAR 252.227-7014 (June 1995), or as
a "commercial item" as defined in FAR 2.101(a) or as "Restricted computer software" as defined in FAR 52.227-
19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of
Software is subject to Keysight Technologies' standard commercial license terms, and non-DOD Departments
and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-
19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR
52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.




Safety Notes
The following safety notes are used throughout this document. Familiarize yourself with each of these notes
and its meaning before performing any of the procedures in this document.

WARNING Warning denotes a hazard. It calls attention to a procedure which, if not correctly performed
or adhered to, could result in injury or loss of life. Do not proceed beyond a warning note until
the indicated conditions are fully understood and met.

CAUTION Caution denotes a hazard. It calls attention to a procedure that, if not correctly performed or
adhered to, could result in damage to or destruction of the instrument. Do not proceed beyond a
caution sign until the indicated conditions are fully understood and met.




2 Installation Note N5242-90004
Description of the Upgrade

IMPORTANT In June 2013, the N5241A/AS and N5242A/AS analyzers underwent significant hardware changes.
Some components that have 2.4 mm connectors (bias tees, couplers, and some semi-rigid cables) were
replaced with components that have 3.5 mm connectors.
If your analyzer's serial number prefix is MY/SG/US5310 and below:
Your analyzer was originally shipped with 2.4 mm components. Bias tees with 2.4 mm connectors are no
longer available, so the bias tees and connecting cables included in this kit have 3.5 mm connectors.
Since they are interconnected, the 2.4 mm couplers and connecting cables in your analyzer must be
replaced with the new 3.5 mm items included in this kit.
If your analyzer's serial number prefix is MY/SG/US5321 and above:
Your analyzer was shipped with 3.5 mm components, so it is not necessary to replace the couplers and
connecting cables. These items are included in this kit, but will not be used for your upgrade. Set them
aside for possible use in the future.
Be very careful to use the appropriate hardware in your analyzer. Using the wrong hardware can ruin
analyzer components, resulting in additional customer costs.

This upgrade converts your standard 4-port configurable test set analyzer (Option 400) to an extended power
range analyzer with bias tees by adding: