Text preview for : E5071C_TDR_MIPI_D-PHY_MOI_110 Method of Implementation(MOI) for MIPI D-PHY Interface S-Parameter and part of Agilent E5071C TDR MIPI D-PHY MOI 110 Method of Implementation(MOI) for MIPI D-PHY Interface S-Parameter and Agilent E5071C_TDR_MIPI_D-PHY_MOI_110 Method of Implementation(MOI) for MIPI D-PHY Interface S-Parameter and Impedance Conformance Tests c20141010 [31].pdf
Back to : E5071C_TDR_MIPI_D-PHY_MOI | Home
Keysight MOI for MIPI D-PHY Conformance Tests
Revision 1.10
10-Oct, 2014
Keysight Method of Implementation (MOI) for MIPI
D-PHY Conformance Tests Using Keysight E5071C ENA
Network Analyzer Option TDR
1
Keysight MOI for MIPI D-PHY Conformance Tests
Table of Contents
1. Modification Record .........................................................................................................4
2. Purpose .............................................................................................................................4
3. References ........................................................................................................................4
4. Resource Requirements ....................................................................................................4
5. Test Procedure ..................................................................................................................5
5.1. Outline of Test Procedure ......................................................................................... 5
5.2. Instrument Setup ....................................................................................................... 6
5.2.1. Recalling State File ............................................................................................ 6
5.2.2. Running VBA macro program ........................................................................... 7
5.3. Screen Configuration ................................................................................................ 9
5.4. Calibration and Adjustment .................................................................................... 11
5.4.1. Time Domain Calibration ................................................................................ 11
5.4.2. Frequency Domain Calibration ....................................................................... 13
5.5. HS-TX S-Parameters .............................................................................................. 14
5.5.1. Differential Return Loss (SDD11) ................................................................... 14
5.5.2. Common-Mode Return Loss (SCC11) ............................................................ 15
5.5.3. Mode Conversion Limits (SDC11).................................................................. 16
5.5.4. Single-Ended Output Impedance (ZOS) ........................................................... 17
5.5.5. Single-Ended Output Impedance Mismatch (ZOS) ........................................ 18
5.6. HS-RX S-Parameters .............................................................................................. 19
5.6.1. Differential Return Loss (SDD11) ................................................................... 19
5.6.2. Common-Mode Return Loss (SCC11) ............................................................ 20
5.6.3. Mode Conversion Limit (SDC11) ................................................................... 21
5.6.4. DC Differential Input Impedance (ZID) ........................................................... 22
6. Appendix ........................................................................................................................24
6.1. Manual Setup for Time Domain Measurement ...................................................... 24
6.1.1. Starting Setup .................................................................................................. 24
2
Keysight MOI for MIPI D-PHY Conformance Tests
6.1.2. HS-TX Single-Ended Output Impedance ........................................................ 25
6.1.3. HS-RX DC Differential Input Impedance ....................................................... 26
6.2. Manual Setup for Frequency Domain Measurement .............................................. 27
6.2.1. Channel and Trace Settings ............................................................................. 27
6.2.2. HS-TX/HS-RX Differential Return Loss ........................................................ 27
6.2.3. HS-TX/HS-RX Common-Mode Return Loss ................................................. 28
6.2.4. HS-TX/HS-RX Mode Conversion Limits ....................................................... 28
6.3. Limit Test Settings .................................................................................................. 28
6.3.1. Turning On/Off Fail Sign ................................................................................ 28
6.3.2. Setting the Warning Beeper ............................................................................. 29
6.3.3. Defining the Limit Line ................................................................................... 29
3
Keysight MOI for MIPI D-PHY Conformance Tests
1. Modification Record
Revision Comments Issue Date
1.00 First Release Dec 1, 2011
1.10 Updated procedures to Conformance Test Suite for D-PHY Oct 10, 2014
Physical Layer Version 1.1 Revision 03.
2. Purpose
This document is intended to provide the measurement procedures for the interface
S-parameter and impedance tests defined in the MIPI Alliance Specification for D-PHY. The
instrument operations contained in this document are designed for Keysight E5071C ENA
Network Analyzer Option TDR.
3. References
MIPI Alliance Conformance Test Suite for D-PHY Physical Layer Version 1.1 Revision 03
MIPI Alliance Specification for D-PHY Version 1.1
4. Resource Requirements
1. ENA Series Network Analyzer with Enhanced Time Domain Option
Note: Use the test set option of 4.5 GHz or above, and either of 2-port and 4-port.
Note: E5071C firmware revision A.11.31 or above is installed.
Note: E5071C-TDR application software revision A.01.56 or above is installed.
2. Electronic Calibration Module N4431B (for 4.5/6.5/8.5 GHz models) or N4433A (14/20
GHz models)
3. 3.5 mm cables of 4 GHz bandwidth or above x2
4
Keysight MOI for MIPI D-PHY Conformance Tests
5. Test Procedure
5.1. Outline of Test Procedure
1. Instrument Setup
2. Calibration and Adjustment
Time Domain Calibration with the Setup Wizard in the TDR Application Software
Frequency Domain Calibration with the VBA Macro Program
3. Measurements and Data Analysis
Time Domain Measurements
- HS-TX Single-Ended Output Impedance
- HS-TX Single-Ended Output Impedance Mismatch
- HS-RX DC Differential Input Impedance
Frequency Domain Measurements
- HS-TX Differential Return Loss
- HS-TX Common-Mode Return Loss
- HS-TX Mode Conversion Limits
- HS-RX Differential Return Loss
- HS-RX Common-Mode Return Loss
- HS-RX Mode Conversion Limits
Note: Hard Keys (Keys located on the Front panel of E5071C) are displayed in Blue color and
Bold. (Example: Avg, Analysis)
Note: Soft keys (Keys on the screen) are displayed in Bold. (Example: S11, Real, Transform)
Note: Buttons (in the TDR or VBA) are displayed in Green color and Bold. (Example: Trace,
Rise Time)
Note: Tabs (in the TDR) are displayed in Brown color and Bold. (Example: Setup, Trace
Control)
5
Keysight MOI for MIPI D-PHY Conformance Tests
5.2. Instrument Setup
This section describes the procedures for recalling the state file and VBA macro that support
the instrument setup. Download "E5071C-TDR Test Package for MIPI D-PHY Tx/Rx
Devices" from www.keysight.com/find/ena-tdr_dphy-txrx.
Extract the zip file and transfer the extracted files to the instrument with a USB flash
memory. For manual measurement settings, refer to 6 Appendix.
5.2.1. Recalling State File
1. If TDR setup wizard appears, click Close button on the wizard.
2. Open Setup tab (item1).
3. Click Advanced Mode (item2).
4. A dialog box appears requesting for confirmation. Then click Yes. (Clear the check box
for "Use Advanced Calibration Methods")
5. Click File (item3) and select Recall State to open the Recall State dialog box.
6. Specify the folder and appropriate state file name, and click Open.
E5071C Test Set Option State File Name
240/245/260/265/280/285/440/445/460/465/480/485 MIPI_D-PHYv1.1_TxRx_24x-48x.tdr
2D5/2K5/4D5/4K5 MIPI_D-PHYv1.1_TxRx_2D5-4K5.tdr
6
Keysight MOI for MIPI D-PHY Conformance Tests
5.2.2. Running VBA macro program
1. Installing the VBA macro program
INSTR STATE Obtain the VBA macro program from
Keysight and install it on the E5071C.
Macro Macro Macro
Setup Run Break Press Save/Recall on the front
Save/ System Preset
Recall panel, then press the Explorer soft
key.
Using Explorer, copy the file to
the "D: