Text preview for : Power Supply Connections for Your CET Signal Conditioner Card Application Note 5990-6549EN c20140929 part of Agilent Power Supply Connections for Your CET Signal Conditioner Card Application Note 5990-6549EN c20140929 Agilent Power Supply Connections for Your CET Signal Conditioner Card Application Note 5990-6549EN c20140929 [9].pdf



Back to : Power Supply Connections | Home

Keysight Technologies
Power Supply Connections for Your
CET Signal Conditioner Card


Application Note
Power Requirement for CET Card

The CET signal conditioner card Power is connected via a wire wrap of AWG28 wires to pins 11 and 12 of J6 of
(referenced in this application note the CET card. There is another pair of power connections on J6 at pins 3 and 4.
as CET card) can be powered from These are the power connections for the USB and do not provide power to the
various sources. The recommended CET card itself.
sources are discussed in the follow-
ing sections.

The input supply voltage to the CET
card is +12 V @ 0.3 A. The selected
source to provide the power needs
to be controllable through the test
plan execution. Power to the CET
card must be uninterrupted for the Connector J6
duration of powered testing. If power
to the CET card is disconnected, an
19 1
"unpowered" command must be
issued before turning the power
back on.
20 2

Figure 1. Connections on CET card (J6 connector)




Table 1. Pin layout of J6 connector on CET card
CET card (J6 connector) Wire color Wire type Remarks
Pin 1 White USB D