Text preview for : Method of Implementation (MOI) for HDMI 1.4b Cable Assembly Test E5071C-TDR_HDMI1.4b_MOI_1.12 c20130 part of Agilent Method of Implementation (MOI) for HDMI 1.4b Cable Assembly Test E5071C-TDR HDMI1.4b MOI 1.12 c20130 Agilent Method of Implementation (MOI) for HDMI 1.4b Cable Assembly Test E5071C-TDR_HDMI1.4b_MOI_1.12 c20130425 [49].pdf



Back to : Method of Implementation | Home

Agilent MOI for HDMI 1.4b Cable Assembly Test




Revision 1.12
24-Apr 2013




Agilent Method of Implementation (MOI) for HDMI 1.4b
Cable Assembly Test Using Agilent E5071C ENA Option
TDR




1
Agilent MOI for HDMI 1.4b Cable Assembly Test



Table of Contents
1. Modification Record ....................................................................................................... 4
2. Purpose ........................................................................................................................... 4
3. References ...................................................................................................................... 4
4. Resource Requirements .................................................................................................. 5
5. Test Procedure ................................................................................................................ 6
5.1. Outline of Test Procedure ......................................................................................... 6
5.2. Instrument Setup ....................................................................................................... 7
5.2.1. Recalling State File ............................................................................................ 7
5.2.2. Running VBA macro program ........................................................................... 8
5.3. Screen Area and Cable Connection ........................................................................ 10
5.4. Calibration and Adjustment .................................................................................... 13
5.4.1. Time Domain Calibration ................................................................................ 13
5.4.2. Measure DUT Length ...................................................................................... 15
5.4.3. Frequency Domain Calibration ....................................................................... 16
5.4.4. Perform Auto Port Extension........................................................................... 16
5.5. Measurement and Data Analysis ............................................................................ 17
5.5.1. Data Eye Diagram ........................................................................................... 17
5.5.2. Intra-Pair Skew ................................................................................................ 24
5.5.3. Inter-Pair Skew ................................................................................................ 26
5.5.4. Far End Crosstalk ............................................................................................ 28
5.5.5. Attenuation and Phase ..................................................................................... 30
5.5.6. Differential Impedance .................................................................................... 34
6. Appendix ...................................................................................................................... 37
6.1. Manual Setup for Time Domain Measurement ...................................................... 37
6.1.1. Starting Setup .................................................................................................. 37
6.1.2. Eye Diagram .................................................................................................... 37
6.1.3. Intra-Pair Skew ................................................................................................ 38


2
Agilent MOI for HDMI 1.4b Cable Assembly Test



6.1.4. Inter-Pair Skew ................................................................................................ 40
6.1.5. Differential Impedance .................................................................................... 40
6.2. Manual Setup for Frequency Domain Measurement .............................................. 43
6.2.1. Channel and Trace Settings ............................................................................. 43
6.2.2. Attenuation ...................................................................................................... 44
6.2.3. Phase ................................................................................................................ 44
6.2.4. Far End Crosstalk ............................................................................................ 45
6.2.5. Expand Phase................................................................................................... 45
6.3. Limit Test Settings .................................................................................................. 45
6.3.1. Turning On/Off Fail Sign ................................................................................ 45
6.3.2. Setting the Warning Beeper ............................................................................. 46
6.3.3. Defining the Limit Line ................................................................................... 46




3
Agilent MOI for HDMI 1.4b Cable Assembly Test



1. Modification Record

Revision Comments Issue Date
1.00 Initial release Jun 01, 2011
1.01 Minor modification for errata Jul 13, 2011
1.10 Added Data Eye Diagram Jan 16, 2012
1.11 Updated Resource Requirements on N1080B HDMI fixtures Jul 19, 2012
1.12 Minor modifications Apr 24, 2013


2. Purpose
This document is intended to provide the measurement procedures for High-Definition
Multimedia Interface (HDMI) cable assemblies with Agilent E5071C ENA Option TDR.
The procedures are designed to perform tests equivalent to the methodologies defined in the
HDMI Compliance Test Specification Version 1.4b (CTS). Devices under test are supposed
to be passive HDMI cables. Active cables are out of the scope of this document.


3. References
High-Definition Multimedia Interface Specification Version 1.4b
High-Definition Multimedia Interface Compliance Test Specification Version 1.4b




4
Agilent MOI for HDMI 1.4b Cable Assembly Test



4. Resource Requirements
1. E5071C ENA Series Network Analyzer with Enhanced Time Domain Analysis Option
Note: Ensure that
- Test set option is any one of 480/485/4D5/4K5
- E5071C firmware revision A.10.05 or above is installed.
- E5071C-TDR application software revision A.01.50 or above is installed.
2. Electronic Calibration Module N4431B (for 480/485) or N4433A (for 4D5/4K5)
3. 3.5 mm cables 8 GHz bandwidth or equivalent x4
4. Certified HDMI receptacle fixtures (ex. Agilent N1080B-H05) x2
5. Certified HDMI plug fixtures (ex. Agilent N1080B-H04) x1
Note: The plug fixture is used for the jitter adjustment in the eye diagram test. It should
have similar characteristics to the receptacle fixtures.
6. 50 Ohm terminators to terminate unused fixture connectors x12
7. Female to female adaptors to connect fixture and test cables (if necessary)




5
Agilent MOI for HDMI 1.4b Cable Assembly Test



5. Test Procedure
5.1. Outline of Test Procedure
1. Instrument Setup
2. Calibration and Adjustment
- Time domain calibration with the setup wizard in TDR application software
- Frequency domain calibration with the VBA macro program
3. Measurements and Data Analysis
Time Domain Measurements
- Data Eye Diagram
- Intra-pair Skew Measurements
- Inter-pair Skew Measurements
- Differential Impedance Measurements
Frequency Domain Measurements
- Far End Crosstalk Measurements
- Attenuation and Phase Measurements


Note: Hard Keys (Keys located on the Front panel of E5071C) are displayed in Blue color
and Bold. (Example: Avg, Analysis)
Note: Soft keys (Keys on the screen) are displayed in Bold. (Example: S11, Real, Transform)
Note: Buttons (in the TDR or VBA) are displayed in Green color and Bold. (Example: Trace,
Rise Time)
Note: Tabs (in the TDR) are displayed in Brown color and Bold. (Example: Setup, Trace
Control)




6
Agilent MOI for HDMI 1.4b Cable Assembly Test



5.2. Instrument Setup
This section describes procedures for recalling the state file and VBA macro that support
the instrument setup. Download "HDMI cable assembly test package" from
http://www.agilent.com/find/ena-tdr_hdmi-cabcon.
Extract the zip file and transfer the extracted files to the instrument with a USB flash
memory.
5.2.1. Recalling State File
1. If TDR setup wizard appears, click Close button on the wizard.
2. Open Setup tab (item1).
3. Click More Function (item2).
4. Click Advanced Mode (item3).






5. A dialog box appears requesting for confirmation. Then click Yes. (Clear the check box
for "Use Advanced Calibration Methods")




6. Click File (item4) and select Recall State to open the Recall State dialog box.
7. Specify a folder and a file name, and click Open.


For manual measurement setup, refer to 6 Appendix.




7
Agilent MOI for HDMI 1.4b Cable Assembly Test



5.2.2. Running VBA macro program
1. Installing the VBA macro program

INSTR STATE Obtain the VBA macro program from
Agilent and install it on the E5071C.
Macro Macro Macro
Setup Run Break Press Save/Recall on the front
Save/ System Preset
Recall panel, then press the Explorer soft
key.
Using Explorer, copy the file to
the "D: