Text preview for : Fluke_730A_Army.pdf part of Fluke 332A Army Calibration Procedure. 332A, 332A/AB DC Transfer Standard



Back to : Fluke_730A_Army.pdf | Home

TB 9-4931-702-50
CHANGE 3 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

CALIBRATION PROCEDURE FOR DC TRANSFER STANDARD JOHN FLUKE MODELS 730A AND 730A/AB
Headquarters, Department of the Army, Washington, DC 20 August 2001
Approved for public release; distribution is unlimited.

TB 9-4931-702-50, 10 May 1974, is changed as follows: 1. Remove old pages and insert new pages as indicated below. material is indicated by a vertical bar in the margin of the page. Remove pages 1 and 2 5 and 6 New or changed

Insert pages 1 and 2 5 and 6

2. File this change sheet in front of the publication for reference purposes.

0120707

Distribution: To be distributed in accordance with Std. IDS No. RLC-1500, 11 May, 1992. .

PIN: 009083-003

TB 9-4931-702-50
Change 2 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

CALIBRATION PROCEDURE FOR DC TRANSFER STANDARD, JOHN FLUKE MODELS 730A AND 730A/AB
Headquarters, Department of the Army, Washington, DC 18 February 1986 TB 9-4931-702-50, 10 May 1974, is changed as follows: Page 2. In table 2, item A2, Minimum Use Specifications column, add superscript 1. In table 2, add footnote to table as follows: "1Prior to using the NULL DETECTOR, determine sensitivity on the 3 and 10 µV ranges, using A1 and A4." By Order of the Secretary of the Army: JOHN A. WICKHAM, JR. General, United States Army Chief of Staff Official: MILDRED E. HEDBERG Brigadier General, United States Army The Adjutant General Distribution: To be distributed in accordance with Std. IDA No. RLC-1500, 17 June 1985.

US GOVERNMENT PRINTING OFFICE: 1987 - - 733-293/40535 REGION #4

PIN: 009083-002

TB 9-4931-702-50
Change 1 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

CALIBRATION PROCEDURE FOR DC TRANSFER STANDARD, JOHN FLUKE MODELS 730A AND 730A/AB
Headquarters, Department of the Army, Washington, DC 20 September 1984 TB 9-4931-702-50, 10 May 1974, is changed as follows: Page 3. Figure 1 is superseded as follows:

Figure 1. Absolute voltage - equipment setup.

PIN: 009083-001

Page 4. Figure 2 is superseded as follows:

Figure 2. Output voltage - equipment setup.

Page 4, paragraph 9b. Add the note below, following the second note. NOTE Do not report the adjustments in this paragraph unless the applicable measurement in the performance check exceeds ±10 microvolts. Page 6, paragraph 10b. Add the note below, following the second note: NOTE Do not report the adjustments in this paragraph unless the applicable measurement in the performance check exceeds ±10 microvolts.

By Order of the Secretary of the Army: FRED C. WEYAND General, United States Army Chief of Staff Official: VERNE L. BOWERS Major General, United States Army The Adjutant General Distribution: To be distributed in accordance with Std. IDA No. RLC-1500, 17 June 1985.

U. S. GPO 1975-0 - 564 - 493/1438

TB 9-4931-702-50
DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

CALIBRATION PROCEDURE FOR DC TRANSFER STANDARD, JOHN FLUKE MODELS 730A AND 730A/AB
Headquarters, Department of the Army, Washington, DC 10 May 1974
REPORTING OF ERRORS AND RECOMMENDED CHANGES You can help improve this publication. If you find any mistakes or if you know of a way to improve the procedure, please let us know. Mail your letter or DA Form 2028 to: Commander, U. S. Army Aviation and Missile Command, ATTN: AMSAM-MMCMA-NP, Redstone Arsenal, AL 35898-5230. A reply will be furnished to you. You may also send in your comments electronically to our e-mail address: [email protected], or FAX 256-842-6546/DSN 788-6546 Paragraph Page SECTION I. IDENTIFICATION AND DESCRIPTION Test instrument identification............................ 1 2 Forms, records, and reports ............................... 2 2 Calibration description ....................................... 3 2 II. EQUIPMENT REQUIREMENTS Equipment required........................................... Accessories required .......................................... CALIBRATION PROCESS Preliminary instructions .................................... Loop closure ...................................................... Linearity of output resistance divider and Output voltage check (Model 730A)..................... Linearity of output resistance divider and Output voltage check (Model 730A/AB) ............... Final procedure.................................................. 4 5 6 7 8 9 10 3 3 3 4 5 9 12

III.

CHANGE 3

TB 9-4931-702-50

SECTION I IDENTIFICATION AND DESCRIPTION 1. Test Instrument Identification. This bulletin provides instructions for the calibration of Dc Transfer Standard, John Fluke Models 730A and 730A/AB. The manufacturer's instruction manuals were used as the prime data source in compiling these instructions. The dc transfer standard will be referred to as the "TI" (test instrument) throughout this bulletin. a. Model Variations. Variations among models are described in text. b. Time and Technique. The time required for this calibration is approximately 4 hours, using the dc and low frequency technique. 2. Calibration Data Card, DA Form 2416 a. Forms, records, and reports required for calibration personnel at all levels are prescribed by TM 38-750. DA Form 2416 must be annotated in accordance with TM 38750 for each calibration performed. b. Adjustments to be reported on DA Form 2416 are designated (R) at the end of the sentence in which they appear. When adjustments are in tables, the (R) will follow the designated adjustment. Report only those adjustments made and designated with (R) 3. Calibration Description. TI parameters and performance specifications which pertain to this calibration are listed in table 1. NOTE For calibration of model 730A to be valid, the TI E control must be set to the last 3 digits of the value of standard cell used in the calibration.
Table 1. Calibration description Test Instrument Parameters Performance Specifications Accuracy (ARITH MEAN Of 1.018 E With ±10 µv of calibrated value at end of calibration Output).1 period. Transfer Voltage Accuracy2 10 V: ±10 µV 1.00 V: 1 µV (1.018 E): ±1 µV (1.019 + E): ±1 µV E: 0.0 To 999 µV In 1 µV Steps: ±1µV
1This value is not the manufacturer's specified value. This value is the maximum drift allowed over a 90-day period as determined for Army calibration applications. 2Accuracy to which voltage can be standardized.

2 CHANGE 3

TB 9-4931-702-50

SECTION II EQUIPMENT REQUIREMENTS 4. Equipment Required. Table 2 identifies the specific equipment used in this calibration procedure. This equipment is issued with secondary reference calibration standards set 4931-621-7878 and is to be used in performing this procedure. Alternate items may be used by the calibrating activity when the equipment listed in table 2 is not available. The items selected must be verified to perform satisfactorily prior to use and must bear evidence of current calibration. The equipment must meet or exceed the minimum use specifications listed in table 2. The accuracies listed in table 2 provide a four-to-one accuracy ratio between the standard and TI. Where the four-to-one ratio cannot be met, the actual accuracy of the equipment selected is shown in parenthesis. 5. Accessories Required. The accessories listed in table 3 are issued .with secondary reference calibration standards set 4931-621-7878 are to be used in this calibration procedure. When necessary, these items may be substituted by equivalent items unless specifically prohibited.
Table 2. Minimum Specifications of Equipment Required Minimum Use Manufacturer, Model And Part Number Common Name Specifications DC VOLTAGE Range: 1.0 to 1.1 volts in 1 John Fluke, Model 332A (7911393) Or John STANDARD microvolt steps; 10.0 to 11 Fluke, Model 332B/AF (6625-150-6994) volts in 1-microvolt steps. NULL DETECTOR Resolution: 0.25 ppm1 John Fluke, Model 845AB (6625-445-3333) STANDARD CELL Test report specification Guildine, Model 9154B (MIS-10364) VOLTAGE DIVIDER Accuracy: ±0.25 ppm ESI, Model RV-726 (MIS-10295)

Item A1

A2 A3 A4

1Prior to using the NULL DETECTOR, determine sensitivity on the 3 and 10 µV ranges, using A1 and A4.

Item B1 B2

Common Name SENSITIVITY SWITCH WIRE

Table 3. Accessories Required Description And Part Number Sensitivity and reversing switch; galvanometer key (7913207) No. 18 AWG, shielded pair, solid copper conductor; electrical power cable (MIL-10312).

SECTION III CALIBRATION PROCESS 6. Preliminary Instructions a. The instructions outlined in this paragraph are preparatory to the calibration process. Personnel should become familiar with the entire bulletin before beginning the calibration.

3

TB 9-4931-702-50 b. Items of equipment used in this procedure are referenced within the text by common name and item identification number as listed in tables 2 and 3. For the identification of equipment referenced by item numbers prefixed with A, see table 2, and for prefix B, see table 3. WARNING HIGH VOLTAGE is used during the performance of this calibration. DEATH ON CONTACT may result if personnel fail to observe safety precautions. NOTE Unless otherwise specified, verify the results of each test and take corrective action whenever, the test requirement is not met before continuing with the calibration. 7. Loop Closure Check NOTE If this is initial calibration of the TI, omit this paragraph an begin with paragraph 8 below. (1) Connect equipment as shown in figure 1.

Figure 1. Absolute voltage - equipment setup.

4

TB 9-4931-702-50 (2) Ensure that the sensitivity switch (B1) is open. CAUTION Avoid damaging or destroying the standard cell (A3) by insuring that the TI output is set to the correct voltage and adjusted as described in the procedure before operating the sensitivity switch. (3) Energize equipment and allow sufficient time for warm-up and stabilization. (4) Set TI controls as follows: (a) Set function selectors and E controls to standard cell value recorded at last calibration. (b) Depress ARITH MEAN pushbutton. (5) Check and record the deviation of the TI ARZTH MEAN output from value recorded at last calibration. Deviation of the two values will be within ±10 microvolts. If not, the calibrations performed during the loop may be invalid. An explanation of appropriate action taken will be recorded to verify loop completion. NOTE The values recorded during the loop closure check and explanations of appropriate action taken with an out-oftolerance reading will be maintained for 3 years. 8. Linearity of Output Resistance Divider and Output Voltage Check (Model 730A) a. Performance Check (1) Connect equipment as shown in figure 2. (2) Position TI controls as listed below: (a) Supply No. 1 function switch to 10V and set E control to the last 3 digits of standard cell (A3) certified value. (b) EXT COMPARISON/OUTPUT 1 depressed. (3) Set voltage divider (A4) to 1.0000000. (4) Set dc voltage standard (A1) to 10.999999. (5) Operate dc voltage standard, sensitivity switch (B1), and null detector (A2) to obtain null on null detector. CHANGE 3 5

TB 9-4931-702-50 (6) Open sensitivity switch. (7) Set voltage divider to .1000000. (8) Set TI supply No. 1 function switch to 1V.

Figure 2. Output Voltage - equipment setup

(9) Operate sensitivity switch. microvolt of null, perform b(1) below. (10) Open sensitivity switch.

If null detector indication is not within ±1

(11) Set voltage divider to 1.0000000. (12) Set dc voltage standard to 1.099999. (13) Operate dc voltage standard, sensitivity switch, and null detector to obtain null on null detector. (14) Open sensitivity switch. (15) Set voltage divider to standard cell value. (16) Position TI supply No. 1 functions switch to 1.018 +E and E control to last 3 digits of standard cell value (should already be).

6

CHANGE 3

TB 9-4931-702-50 (17) Operate sensitivity-switch. microvolt of null, perform b(2) below. (18) Open sensitivity switch. (19) Set voltage divider to 1.019---0. standard cell value.) (The 3 blanks represent last 3 digits of If null detector indication is not within ±1

(20) Set TI supply No. 1 function switch to 1.019 ±E. (21) Operate sensitivity switch. microvolt of null. perform b(3) below. (22) Open sensitivity switch. (23) Set voltage divider to 1.0199990. (24) Set TI supply No. 1 E control to 999. (25) Operate sensitivity switch. microvolt of null, perform b(4) below. (26) Open sensitivity switch. (27) Set voltage divider to .0009990. (28) Set TI supply No. 1 function switch to E. (29) Operate sensitivity switch. microvolt of null, perform b(5) below. (30) Open sensitivity switch. (31) Repeat technique of (1) through (30) above for supplies 2, 3, and 4. (32) Connect equipment as shown in figure 1. (33) Set TI supply No. 1 function switch and E control for output equal to standard cell certified value. (34) Operate the sensitivity switch. If null detector indication is not within ±1 microvolt of null, perform b(6) below. (35) Open sensitivity switch. (36) Repeat technique of (33) through (35) above for supplies 2, 3, and 4. 7 If null detector indication is not within ±1 If null detector indication is not within ±1 If null detector indication is not within ±1

TB 9-4931-702-50 (37) Depress ARITH MEAN pushbutton with all four supplies set at standard cell value. If null detector indication is not within ±1 microvolt of null, repeat (1) through (36) above. Record ARITH MEAN output value for use in loop closure check. (38) Affix label to TI showing standard cell value used for calibration and a statement to the effect that the E control must be set to the last 3 digits of the standard cell value for calibration to be valid. b. Adjustments NOTE Remove TI protective cover only to make adjustments. Replace for indications. NOTE In (1) through (4) and (6) below, the designation of the adjustments for each reference supply is the same. When reporting these adjustments, the component designator must be prefaced with the proper supply designator. For example: if working with supply No. 1, R25 should be reported as A3A1R25; if working with supply No. 3, R25 should be reported as A3A3R25 NOTE Do not report the adjustments in this paragraph unless the applicable measurement in the performance check exceeds ±10 microvolts. (1) Adjust supply No. 1. "1V CAL" control (R25, fig. 3) to obtain null on null detector (R) (2) Adjust supply No. 1, "1.018 CAL" control (R17, fig. 3) to obtain null on null detector (R). (3) Adjust supply No. 1, "1.019 CAL" control (R18. fig. 3) to obtain null on null detector (R) (4) Adjust supply No. 1, " 1.018/1.019 +E CAL" control (R19, fig. 3) to obtain null on null detector (R) (5) Adjust supply No. 1, "E CAL" control (R5, fig. 3) to obtain null on null detector (R). NOTE In (5) above all four "E CAL" controls are designated R5. They are four different adjustments and all are located on 8

TB 9-4931-702-50 the main PCB board A3 (fig. 3). When reporting those adjustments, the component designator must be prefaced with supply number begin adjusted. For example: if working,, with supply No. 1, report supply No. 1 R5; if working, with supply No. 3, report supply No. R5.

Figure 3. Test instrument (model 730A) - internal view.

(R)

(6) Adjust supply No. 1 "CAL" control (R13, fig. 3) to obtain null on null detector

9

TB 9-4931-702-50 9. Linearity of Output Resistance Divider and Output Voltage Check (Model 730A/AB) a. Performance Check (1) Connect equipment as shown in figure 2. (2) Position 11 controls as listed below: (a) Supply No. 1 function switch to 10V. (b) Supply No. 1 E control to 000. (c) EXT COMPARISON/OUTPUT 1 depressed. (3) Set voltage divider (A4) to 1.0000000. (4) Set dc voltage standard (A1) to 10.999999. (5) Operate dc voltage standard, sensitivity switch (B1), and null detector (A2) to obtain null on null detector. (6) Open sensitivity switch. (7) Set voltage divider to .1000000. (8) Set TI supply No. 1 function switch to 1V. (9) Operate sensitivity switch. If indication is not within ±1 microvolt of null, perform b(1) below. (10) Open sensitivity switch. (11) Set voltage divider to 1.0000000. (12) Set dc voltage standard to 1.099999. (13) Operate dc voltage standard, sensitivity switch, and null detector to obtain null on null detector. (14) Open sensitivity switch. (15) Set voltage divider to 1.018000. (16) Position TI supply No. 1 function switch to 1.018 +E.

10

TB 9-4931-702-50 (17) Operate sensitivity switch. microvolt of null, perform b(2) below. (18) Open sensitivity switch. (19) Set voltage divider to 1.0190000. (20) Set TI supply No. 1 function switch to 1.019 +E. (21) Operate sensitivity switch. microvolt of null, perform b(3) below. (22) Open sensitivity switch. (23) Set voltage divider to 1.0199990. (24) Set TI supply No. 1 E control to 999. (25) Operate sensitivity switch. microvolt of null, perform b(4) below. (26) Open sensitivity switch. (27) Set voltage divider to .0009990. (28) Set TI supply No. 1 function switch to E. (29) Operate sensitivity switch. microvolt of null, perform b(5) below. (30) Open sensitivity switch. (31) Repeat technique of (1) through (30) above for supplies 2, 3, and 4. (32) Connect equipment as shown in figure 1. (33) Set TI supply No. function switch and E control for output equal to standard cell (A3) certified value. (34) Operate sensitivity switch. microvolt of null, perform b(6) below. (35) Open sensitivity switch. (36) Repeat technique of (33) through (35) above for supplies No. 2, 3, and 4. If null detector indication is not within ±1 If null detector indication is not within ±1 If null detector indication is not within ±1 If null detector indication is not within ±1 If null detector indication is not within ±1

11

TB 9-4931-702-50 (37) Depress ARITH MEAN pushbutton with all four supplies set at standard cell value. If null detector indication is not within ±1 microvolt of null, repeat (1) through (36) above. Record ARITH MEAN output value for use in loop closure check. (38) Affix label to TI showing standard cell value used for calibration. b. Adjustments NOTE Remove TI protective cover only to make adjustments. Replace for indications. NOTE In (1) through (4) and (6) below, the designation of the adjustments for each reference supply is the same. When reporting these adjustments, the component designator must be prefaced with the proper supply designator. For example: if working with supply N 1, R25 should be reported as o. A3A1R25; if working with supply No. 3, R25 should be reported as A3A3R25 NOTE Do not report the adjustments in this paragraph unless the applicable measurement in the performance check exceeds ±10 microvolts. (1) Adjust supply No. 1 "1V CAL" control (R19, fig. 4) to obtain null on null detector (R). (2) Adjust supply No. 1 "1.018 CAL" control (R17, fig. 4) to obtain null on null detector (R). (3) Adjust supply No. 1 "1.019 CAL" control (R15, fig. 4) to obtain null on null detector (R). (4) Adjust supply No. 1 "1.018/1.019 +E CAL" control (R12, fig. 4) to obtain null on null detector (R). (5) Adjust supply No. 1 "E CAL" control (R4, fig. 4) to obtain null on null detector (R). (For supply No. 2 adjust R8 (R), supply No. 3 adjust R12 (R), and supply No. 4 adjust R16 (R) (R). (6) Adjust supply No. 1 "CAL" control (R11, fig. 4) to obtain null on null detector

12

TB 9-4931-702-50 10. Final Procedure a. Deenergize and disconnect all equipment. b. In accordance with TM 38-750, annotate and affix DA Label 80 (U.S. Army Calibration System). When the TI cannot be adjusted within tolerance, annotate and affix DA Form 2417 (Unserviceable or Limited Use)

Figure 4. Test instrument (model 730A/AB) - internal view.

13

TB 9-4931-702-50 By Order of the Secretary of the Army: CREIGHTON W. ABRAMS General, United States Army Chief Of Staff Official: VERNE L. BOWERS Major General, United States Army The Adjutant General DISTRIBUTION: To be distributed in accordance with Std. IDS No. RLC-1500, 13 March 1974..

US GOVERNMENT PRINTING OFFICE: 1974 - 747699/256

PIN NO: 009083-003