Text preview for : 2306-910-01C.pdf part of Keithley 2306-910-01C Keithley 2306 2306-910-01C.pdf



Back to : 2306-910-01C.pdf | Home

Model 2306-VS Voltage Protection
Keithley Instruments, Inc. Condition with External Triggering
28775 Aurora Road
Cleveland, Ohio 44139 Release Note
(440) 248-0400
www.keithley.com




If only one channel is used for external triggering functionality, and if the sense and source lines are left
unconnected on the unused channel, Voltage Protection (VPT) condition may occur since a channel on the
Model 2306-VS with no load may drift high or low. If it drifts high, then a VPT condition will occur since VPT is
not dependent on output state (it is dependent on set voltage value and VPT settings). When using external
triggering functionality, a VPT condition on either channel will cause the channel using external triggering to
stop. This is done to allow the VPT condition to be detected in cases where the Model 2306-VS is constantly
being externally triggered for the desired functionality. To avoid a VPT condition, connect the sense and
source lines on the unused channel. See the TRIG[1]: EXT:STEP:VPT and the TRIG2:EXT:STEP:VPT com-
mands for guidance on disabling VPT checking on the channel being externally triggered only.




2306-910-01 Rev. C / 10-04