Text preview for : 3458A-20.pdf part of Agilent 3458A-20 Agilent 3458A service 3458A-20.pdf
Back to : 3458A-20.pdf | Home
MODIIFIICATIION RECOMMENDED
MOD F CAT ON RECOMMENDED
3458A-20
S E R V I C E N O T E
Supersedes:
NONE
3458A Multimeter
Serial Numbers: MY45043514 / MY45044113 and SG45040233 / SG45040350
Bad SRAM causing Cal Ram batteries to fail prematurely
Parts Required:
P/N Description Qty.
03458-66547 A5 Outguard controller 1
OR
03458-66548 A5 Outguard controller (opt. 001) 1
ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:
MODIFICATION RECOMMENDED
ACTION X ON SPECIFIED FAILURE STANDARDS
CATEGORY: [[]] AGREEABLE TIME LABOR: 1.0 Hours
LOCATION [[]] CUSTOMER INSTALLABLE SERVICE X RETURN USED X RETURN
CATEGORY: [[]] ON-SITE INVENTORY: [[]] SCRAP PARTS: [[]] SCRAP
X SERVICE CENTER [[]] SEE TEXT [[]] SEE TEXT
[[]] CHANNEL PARTNER
AVAILABILITY: PRODUCT'S SUPPORT LIFE NO CHARGE AVAILABLE UNTIL: December 2010
AUTHOR: LH PRODUCT LINE: SP
ADDITIONAL INFORMATION: