Text preview for : TEK 067-0599-00 Operation.pdf part of Tektronix TEK 067-0599-00 Operation Tektronix TEK 067-0599-00 Operation.pdf
Back to : TEK 067-0599-00 Operation | Home
067-0599-00 CALIBRATION FIXTURE
576Cdibrotion
Unit
The 067-0599-00 Calibration Pixture was designed specifically for
calibrating and testing the performance of the 576 Curve Tracer.
Installed in place of the Standard Test Fixture plug-in, the 067-0599-00
provides accurate calibration qLtages and current5 for the horizontal and
vertical deflection circuits and accurate loads for the collector supply and
8tep generator circuits through the external input and output lines at the
test fixture interface. External monitoring of the step generator output
ia available through iln EXTERNAL MONITOR BNC jack providing a 1 V/STEP
(into 10,000 Mj?) output.
The Vectical, Display Offset, Horizonta.1 snd Step Generator switches
of ehe 576 have marching switches on the 067-0599-00 for easy, direct-
reading, Comparison-Style adjusting and checking, using the 576's own
display.
The 067-0599-00 iS totally dependent on the 576 for regulated low voltage
power supplies and AC line voltage.
062-1101-00
- . <
. . .--..e -m.-.. .
TEK .INTER-OFFICE COMlvllJNlCATlON
.-
.
Sohn Martin 94-540 3:.' t Juno 25, 1991
TO
Frank Gray, SO-PAT
FWW
GIDEP permit request
M3KC.T
In response to the request to grant permission to the
Government Industry Exchange Pro ram (GPDEP) to reproduce
Tektronix operator, service and P nstruction manuals,
Tektronix, Inc. hereby grant6 such permission for distribution
of such documents to any GTDEP user that is a full participant
in the Metrology Data Interchange Data Base of GIDEP provided
that all copies of the original work include the entire
copyright notice and ownership statement exactly a-& it appears
in the original, together with the Legend "Reproduded with
pemieeion,a
This permission has been approved by the Intellectual
Committee of Tektronix and a copy of this memo may
aeon% 1[ded to GIDEP to provide'the requested permission.
g:&dy
Group Pat&t Coun&al
CONTENTS
Section 1 Characteristics
Electrical
Environmental
Physical
Mechanical
Controls and Connectors
Section 2 Operating Instructions
Section 3 Circuit Description
Section 4 Maintenance
General Information
s
daiibrstion
Section 5 Schematic Diagrams
Section 6 Electrical Parts List
Section 7 Mechanical Parts List
Mechanical Parts List Illustrations
Type 067-0599-00
SECTION 1
CHARACTERISTICS
ELECTRICAL
Chasacteristics Performance Limits
Voltage Accuracies
CALIBRGTOR RANGES 25 mV, 50 mV, 100 mV, 125 mV, 200 mV:
all within 0.04%
50 mV VARIABLE Range At least -4% fo at least +4%
HORIZONTAL VOLTS
COLLECTOR Within 0.5%
BASE Within 0.5%
STEP GENERATOR Within 0.5%
Current Accuracies
VERTICAL
COLLECTOR Within 0.5%
Within 0.5%
Within 0.5%
Resistance Ratios
DISPLAY OFFSET MU LTIPLIER Within 0.04%
4%'
q`v COLL VOLTAGE + 10 INTO 10 Mf2 Within 0.8%
Resistive Loads
1 K COLLECTOR SHORT 1 I& within 0.5%
lKt18K 19 kra, within 5%
Characteristics--Type 067-0599-00
ELECTRICAL (cant)
Characteristics I Performance Limits
Resistance Loads (cant) I
.l n 0.1 R, within 3%
40 v LOAD 3.6 kn, within 5%
I 75 Cl, within 1%
EXTERNAL MONITOR Output 1 V/STEP (INTO 10,000 IQ),
(With STEP GENERATOR LOADS within 0.5%
Switch in EXT ONLY or STEP
GEN)
POWER REQUIREMENTS
Characteristics Performance Requirements
Power derived through test fixture
interface of the 576.
ENVIRONMENTAL
Characteristics Performance Requirements
I
Temperature
Operating Range Performance requirements listed
I apply over the range of 20