Text preview for : HP 3614A_252C 15A_252C 16A_252C 17A Operating & service.pdf part of Agilent HP 3614A 252C 15A 252C 16A 252C 17A Operating & service Agilent HP 3614A_252C 15A_252C 16A_252C 17A Operating & service.pdf



Back to : HP 3614A_252C 15A_252C 16 | Home

I


Agilent E361xA 60W BENCH SERIES DC POWER SUPPLIES



OPERATING AND SERVICE MANUAL FOR MODELS:



Agilent E3614A, Serials KR83503035 and above
Agilent E3615A, Serials KR83506197 and above
Agilent E3616A, Serials KR83502651 and above
Agilent E3617A, Serials KR83502522 and above



For instruments with higher Serial Numbers than above,
a change page may be included.




Manual Part No. 5959-5310 April 2000
Edition 8
SAFETY SUMMARY
The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument.
Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design,
manufacture, and intended use of the instrument. Agilent Technologies assumes no liability for the customer's failure to comply
with these requirements.



BEFORE APPLYING POWER. SAFETY SYMBOLS
Verify that the product is set to match the available line voltage
Instruction manual symbol; the product will
and that the correct fuse is installed.
be marked with this symbol when it is neces-

GROUND THE INSTRUMENT. ! sary for the user to refer to the instruction
manual.
This product is a Safety Class I instrument (provided with a pro-
tective earth terminal). To minimize shock hazard, the instrument
chassis and cabinet must be connected to an electrical ground. or Indicate earth(ground) terminal.
The instrument must be connected to the ac power supply mains
through a three-conductor power cable, with the third wire firmly
connected to an electrical ground(safety ground) at the power
outlet. Any interruption of the protective(grounding) conductor or The WARNING sign denotes a hazard. It
disconnection of the protective earth terminal will cause a poten- calls attention to a procedure, practice,
WARNING or the like, which, if not correctly performed or
tial shock hazard that could result in personal injury. If the instru-
adhered to, could result inpersonal injury. Do
ment is to be energized via an external autotransformer for
not proceed beyond a WARNING sign until
voltage reduction, be certain that the autotransformer common the indicated conditions are fully understood
terminal is connected to the neutral(earthed pole) of the ac power and met.
lines (supply mains).

DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE. The CAUTION sign denotes a hazard. It calls
Do not operate the instrument in the presence of flammable attention to an operating procedure, or the
gases or fumes. like, which, if not correctly performed or
CAUTION adhered to, could result in damage to or
destruction of part or all of the product. Do
KEEP AWAY FROM LIVE CIRCUITS.
not proceed beyond CAUTION sign until the
Operating personnel must not remove instrument covers. Compo-
indicated conditions are fully understood and
nent replacement and internal adjustments must be made by
met.
qualified service personnel. Do not replace components with
power cable connected. Under certain conditions, dangerous volt-
ages may exist even with the power cable removed. To avoid inju- The NOTE sign denotes important infor-
ries, always disconnect power, discharge circuits and remove NOTE mation. It calls attention to a procedure, prac-
external voltage sources before touching components. tice, condition or the like, which is essential to
highlight.
DO NOT SERVICE OR ADJUST ALONE.
Do not attempt internal service or adjustment unless another per-
son, capable of rendering first aid and resuscitation, is present. DO NOT SUBSTITUTE PARTS OR MODIFY INSTRUMENT.
Because of the danger of introducing additional hazards, do not
install substitute parts or perform any unauthorized modification
to the instrument. Return the instrument to a Agilent Technologies
Sales and Service Office for service and repair to ensure that
safety features are maintained.




Instruments that appear damaged or defective should be made inoperative and secured against unintended
operation until they can be repaired by qualified service personnel.




1-2
Table of Contents


SAFETY SUMMARY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
GENERAL INFORMATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
SAFETY REQUIREMENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
INSTRUMENT AND MANUAL IDENTIFICATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
OPTIONS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
ACCESSORY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
INSTALLATION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
INITIAL INSPECTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Mechanical Check. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Electrical Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
INSTALLATION DATA. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Location and Cooling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Outline Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Rack Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
INPUT POWER REQUIREMENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Line Voltage Option Conversion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Power Cord . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
OPERATING INSTRUCTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
TURN-ON CHECKOUT PROCEDURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
OPERATING MODES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
LOCAL OPERATING MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Constant Voltage Operaton. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Constant Current Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Overvoltage Protection (OVP). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
CONNECTING LOADS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
OPERATION BEYOND RATED OUTPUT. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
REMOTE OPERATING MODES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
Remote Voltage Sensing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
Remote Analog Voltage Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
MULTIPLE-SUPPLY OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
NORMAL PARALLEL OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
AUTO-PARALLEL OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
NORMAL SERIES OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11
AUTO-SERIES OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-12
AUTO-TRACKING OPERATON . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-13
LOAD CONSIDERATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-14
PULSE LOADING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-14
REVERSE CURRENT LOADING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-14
OUTPUT CAPACITANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-14
REVERSE VOLTAGE LOADING. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-14
BATTERY CHARGING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-14


1-3
GENERAL INFORMATION OPTIONS
Options OE3 and OE9 determine which line voltage is
INTRODUCTION selected at the factory. The standard unit is configured for 115
This manual describes all models in the Agilent E361xA 60W Vac