Text preview for : FieldFox Microwave Analyzers - White Paper 5991-1300EN c20140919 [8].pdf part of Agilent FieldFox Microwave Analyzers - White Paper 5991-1300EN c20140919 [8] Agilent FieldFox Microwave Analyzers - White Paper 5991-1300EN c20140919 [8].pdf
Back to : FieldFox Microwave Analyz | Home
Keysight Technologies
FieldFox Microwave Analyzers
4/6/6.5/9/14/18/26.5 GHz
Introduction
Today, accurate RF and microwave measurements are becoming increasingly necessary in harsh
and hard-to-reach environments. These scenarios span a wide range of conditions: day or night,
rain or shine, hot or cold; aboard a ship, in an aircraft, or in a vehicle.
In these situations, a handheld instrument must be capable of making the required measurements
with sufficient levels of performance and accuracy. Those attributes are important; however, given
the expected operating conditions, physical and environmental specifications are equally important
to the instrument and the user (Figure 1). This white paper presents general suggestions and specific
examples regarding the essential attributes of handheld analyzers that will be used in harsh conditions.
Figure 1. The ideal RF or microwave handheld analyzer must be capable of making accurate measurements in
tough working conditions.
Establishing a reference for "ruggedness"
United States military specification US MIL-PRF-28800F provides a set of benchmarks for test
equipment that will be used in the testing and calibrating of electrical and electronic equipment.
The spec includes four classes that range from the "extremes of world climatic variation" (Class 1)
to controlled, protected operational environments (Class 4).
The 88-page specification is quite detailed, and section 3.8 uses 13 major categories to define the
environmental requirements. Six of these are especially important in handheld analyzers: