This manual change describes the procedure when use the Agilent 16353A/16353H
Standard Resistor Set instead of the Agilent 16340A RC-Box in 4339B Performance Test.
When use the Agilent 16340A RC-Box, refer to the Chapter 9 (Maintenance) in this
manual.
Manual Change
Test Equipment




Test Equipment
Table 1 lists the recommended equipment for performing 4339B maintenance.
Table 1 Required Equipment

Equipment Requirements Recommended Qty. Use*1
Model

DC Voltmeter Voltage Range: 0V to 1000V 3458A 1 P
Accuracy: <0.04%
Standard Resistor No substitute 16353A/16353H 1 P, F
Set
10 k Resistor 42038A 1 P
RC Box Adapter No substitute PN 04339-65005 1 P, F
V Measurement No substitute PN 04339-65006 1 P
Adapter
BNC(F)-TXA 1 P
Ground Open (F)
Adapter
BNC(F)-(F) PN 1250-1830 1 P
Adapter
Cable Banana-Banana Jumper 11058A 1 P
Cable Dual Banana-BNC(m) Cable PN 11001-60001 1 P
Handler Interface No Substitute PN 04339-65007 1 F
Tester
*1.P: Performance Tests, F: Functional Tests
Manual Change
Performance Tests




Performance Tests

Introduction
This section provides the test procedures used to verify that the 4339B's specifications
listed in Chapter 8, General Information, of this manual are met. All tests can be performed
without access to the interior of the instrument. The performance tests can also be used to
perform incoming inspection, and to verify that the 4339B meets its performance
specifications after troubleshooting or adjustment. If the performance tests indicate that the
4339B is NOT operating within the specified limits, check your test setup, then proceed
with troubleshooting if necessary.

NOTE Allow the 4339B to warm up for at least 30minutes before you execute any of the
performance tests.


NOTE Perform all performance tests in ambient conditions of 23