Text preview for : 85p2cmds.PDF part of Tektronix 85p2cmds Tektronix TEK CTS850 Test Set Reference 85p2cmds.PDF



Back to : 85p2cmds.PDF | Home

CTS 850 Cross-Reference listing Remote Commands

10/23/98


Matt Fichtenbaum

There are now two separate files. Each begins with the various major menus (Test Setups, Tx #1,
Tx #2, Utility, etc.) and lists, for each horizontal menu button, the assocaited HIF page(s) and the
sequence numbers of those pages in the listing. This "index" is new.

Then comes the main listing, a picture of each HIF screen listing the menu items on the screen
together with their screen labels and either (1) the remote commands associated with each item, or
(2) the vertical menu button choices associated with each item (new).

Document file 85p2cmds.doc lists Remote Commands; Document file 85p2vert.doc lists the
Vertical Menu Choices.




CTS 850 Cross-Reference listing Remote Commands 1
INDEX
** denotes pages that have no menus and are not shown

----P_RESULTS_MAJOR_MENU1----
TEST SDH PDH JITTE ERROR PERF -more
SUMMA RESUL RESUL WANDE ANALY ANALY 1 of

----------------------
Button Page
1. ** P_RESULTS_TEST_SUMMARY_PARAM1
2. ** P_RESULTS_ALL_RESULTS_PARAM1
** P_RESULTS_ALL_RESULTS_SDH_SECTION_DEFECTS_PARAM1
3. ** P_RESULTS_ALL_RESULTS_PARAM1
** P_RESULTS_ALL_RESULTS_SDH_SECTION_DEFECTS_PARAM1
4. ** P_RESULTS_ALL_RESULTS_PARAM10
5. ** P_RESULTS_ALL_RESULTS_PARAM1
** P_RESULTS_ALL_RESULTS_SDH_SECTION_DEFECTS_PARAM1
6. 3 P_PERFORMANCE_ANALYSIS_M2101_PARAM1
=========================================

----P_RESULTS_MAJOR_MENU2----
TEST SDH PDH HISTO SAVE RECAL -more
SUMMA RESUL RESUL GRAPH RESUL RESUL 2 of

----------------------
Button Page
1. ** P_RESULTS_TEST_SUMMARY_PARAM1
2. ** P_RESULTS_ALL_RESULTS_PARAM1
** P_RESULTS_ALL_RESULTS_SDH_SECTION_DEFECTS_PARAM1
3. ** P_RESULTS_ALL_RESULTS_PARAM1
** P_RESULTS_ALL_RESULTS_SDH_SECTION_DEFECTS_PARAM1
4. ** P_RESULTS_HISTORY_GRAPHS_PARAM1
5. 7 P_RESULTS_SAVE_RESULTS_PARAM1
6. 6 P_RESULTS_RECALL_RESULTS_PARAM1
=========================================

----P_RX_MAJOR_MENU----
RECEI SIGNA JITTE TRANS PATH
SETTI STATU WANDE OVERH OVERH

----------------------
Button Page
1. 19 P_RX_SETTINGS_PARAM1
20 P_RX_SETTINGS_TRIB_PARAM1
2. ** P_RX_SIGNAL_STATUS_PARAM1
** P_RX_SIGNAL_STATUS_PARAM2
3. 11 P_RX_JITTER_WANDER_PARAM1
** P_TX_JITTER_PARAM1_NO_OPTION
4. 22 P_RX_TRANSPORT_OVERHEAD_PARAM1
5. 12 P_RX_PATH_OVERHEAD_PARAM1
13 P_RX_PATH_OVERHEAD_V5_PARAM1
14 P_RX_PATH_OVERHEAD_VC12_PARAM1
15 P_RX_PATH_OVERHEAD_VC3_PARAM1
=========================================

----P_RX_SDH_MAJOR_MENU1----
RECEI 64k R SIGNA ANALY JITTE -more
SETTI SETTI STATU CONFI WANDE 1 of

----------------------
Button Page
1. 17 P_RX_SDH_SETTINGS_PARAM1
18 P_RX_SDH_SETTINGS_TRIB_PARAM1
2. 16 P_RX_SDH_K64_SETTINGS_PARAM1



CTS 850 Cross-Reference listing Remote Commands 2
3. ** P_RX_SIGNAL_STATUS_PARAM1
** P_RX_SIGNAL_STATUS_PARAM2
4. 8 P_RX_ANALYSIS_CONFIGURATION_SETTINGS_PARAM1
5. 11 P_RX_JITTER_WANDER_PARAM1
** P_TX_JITTER_PARAM1_NO_OPTION
=========================================

----P_RX_SDH_MAJOR_MENU2----
RECEI 64k R SECTI PATH Rx CA -more
SETTI SETTI OVERH OVERH VOICE 2 of

----------------------
Button Page
1. 17 P_RX_SDH_SETTINGS_PARAM1
18 P_RX_SDH_SETTINGS_TRIB_PARAM1
2. 16 P_RX_SDH_K64_SETTINGS_PARAM1
3. 22 P_RX_TRANSPORT_OVERHEAD_PARAM1
4. 12 P_RX_PATH_OVERHEAD_PARAM1
13 P_RX_PATH_OVERHEAD_V5_PARAM1
14 P_RX_PATH_OVERHEAD_VC12_PARAM1
15 P_RX_PATH_OVERHEAD_VC3_PARAM1
5. 9 P_RX_CAS_VOICE_SETTINGS_PARAM1
10 P_RX_CAS_VOICE_SETTINGS_PARAM2
=========================================

----P_TEST_SETUP_MAJOR_MENU----
TEST RECAL RECAL SAVE SAVE JITTE
INSTRUMENT PASS/ INSTRUMENT PASS/
CONTR SETUP TESTS SETUP TESTS TESTS
----------------------
Button Page
1. 36 P_TEST_SETUP_TEST_CONTROL_PARAM1
2. 32 P_TEST_SETUP_RECALL_INSTRUMENT_SETUPS_PARAM1
3. 33 P_TEST_SETUP_RECALL_PASS_FAIL_TESTS_PARAM1
4. 34 P_TEST_SETUP_SAVE_INSTRUMENT_SETUPS_PARAM1
5. 35 P_TEST_SETUP_SAVE_PASS_FAIL_TESTS_PARAM1
6. 25 P_TEST_JITTER_OUTPUT_PARAM1
26 P_TEST_JITTER_POINTER_PARAM1
27 P_TEST_JITTER_RECALL_RESULTS
28 P_TEST_JITTER_SAVE_RESULTS
29 P_TEST_JITTER_SPECTRUM_PARAM1
30 P_TEST_JITTER_TOLERANCE_PARAM1
31 P_TEST_JITTER_TRANSFER_PARAM1
** P_TEST_JITTER_VIEW_DATA_TOLERANCE
** P_TEST_JITTER_VIEW_DATA_TRANSFER
** P_TEST_JITTER_VIEW_GRAPH_SPECTRUM
** P_TEST_JITTER_VIEW_GRAPH_TOLERANCE
** P_TEST_JITTER_VIEW_GRAPH_TRANSFER
** P_TX_JITTER_PARAM1_NO_OPTION
=========================================

----P_TX_MAJOR_MENU----
TRANS ERROR POINT JITTE APS TRANS PATH
SETTI & ALA & TIM WANDE COMMA OVERH OVERH

----------------------
Button Page
1. 49 P_TX_SDH_SETTINGS_PARAM1
50 P_TX_SDH_SETTINGS_TRIB_PARAM1
51 P_TX_SETTINGS_PARAM1
52 P_TX_SETTINGS_THROUGH_MODE_PARAM1
53 P_TX_SETTINGS_TRIB_PARAM1
2. 39 P_TX_ERRORS_ALARMS_PARAM1
3. 45 P_TX_POINTERS_TIMING_FREQUENCY_PARAM1
46 P_TX_POINTERS_TIMING_PARAM1
47 P_TX_POINTERS_TIMING_SEQUENCE_PARAM1
4. 40 P_TX_JITTER_PARAM1



CTS 850 Cross-Reference listing Remote Commands 3
** P_TX_JITTER_PARAM1_NO_OPTION
5. 38 P_TX_APS_COMMANDS_PARAM1
6. 54 P_TX_TRANSPORT_OVERHEAD_PARAM1
7. 41 P_TX_PATH_OVERHEAD_PARAM1
42 P_TX_PATH_OVERHEAD_V5_PARAM1
43 P_TX_PATH_OVERHEAD_VC12_PARAM1
44 P_TX_PATH_OVERHEAD_VC3_PARAM1
=========================================

----P_TX_SDH_MAJOR_MENU1----
TRANS 64k T DEFEC POINT JITTE APS -more
SETTI SETTI ANOMA & TIM WANDE COMMA 1 of

----------------------
Button Page
1. 49 P_TX_SDH_SETTINGS_PARAM1
50 P_TX_SDH_SETTINGS_TRIB_PARAM1
52 P_TX_SETTINGS_THROUGH_MODE_PARAM1
53 P_TX_SETTINGS_TRIB_PARAM1
2. 48 P_TX_SDH_K64_SETTINGS_PARAM1
3. 39 P_TX_ERRORS_ALARMS_PARAM1
4. 45 P_TX_POINTERS_TIMING_FREQUENCY_PARAM1
46 P_TX_POINTERS_TIMING_PARAM1
47 P_TX_POINTERS_TIMING_SEQUENCE_PARAM1
5. 40 P_TX_JITTER_PARAM1
** P_TX_JITTER_PARAM1_NO_OPTION
6. 38 P_TX_APS_COMMANDS_PARAM1
=========================================

----P_TX_SDH_MAJOR_MENU2----
TRANS SECTI PATH TRACE SIGNA OVERH -more
SETTI OVERH OVERH SETTI LABEL PRBS 2 of

----------------------
Button Page
1. 49 P_TX_SDH_SETTINGS_PARAM1
50 P_TX_SDH_SETTINGS_TRIB_PARAM1
52 P_TX_SETTINGS_THROUGH_MODE_PARAM1
53 P_TX_SETTINGS_TRIB_PARAM1
2. 54 P_TX_TRANSPORT_OVERHEAD_PARAM1
3. 41 P_TX_PATH_OVERHEAD_PARAM1
42 P_TX_PATH_OVERHEAD_V5_PARAM1
43 P_TX_PATH_OVERHEAD_VC12_PARAM1
44 P_TX_PATH_OVERHEAD_VC3_PARAM1
4. 37 P_TRACE_SETTINGS_PARAM1
5. 24 P_SIGNAL_LABELS_PARAM1
6. 1 P_OVERHEAD_PRBS_TEST_PARAM1
=========================================

----P_UTILITY_MAJOR_MENU----
MISC PRINT REMOT INSTR SELF JITTE
SETTI SETUP CONTR CONFI TEST CAL

----------------------
Button Page
1. 58 P_UTILITY_MISC_SETTINGS_PARAM1
2. 59 P_UTILITY_PRINTER_SETUP_PARAM1
3. 60 P_UTILITY_REMOTE_CONTROL_PARAM1
4. 57 P_UTILITY_INSTRUMENT_CONFIG_PARAM1
5. 61 P_UTILITY_SELF_TEST_PARAM1
6. 55 P_UTILITY_CALIBRATION_PARAM1
** P_UTILITY_CALIBRATION_M45_PARAM1
=========================================




CTS 850 Cross-Reference listing Remote Commands 4
1.

----P_OVERHEAD_PRBS_TEST_PARAM1----
Tx Overhead PRBS Test ...... None 2^15-1 Normal
Rx Overhead PRBS Test ...... None 2^15-1 Normal
Pattern Lock ... Unlocked

Errors ... BER ...
----------------------
Item OVERHEAD_TEST_SELECT_TX_BYTE ("Tx Overhead PRBS Test ...")
:source:data:telecom:overhead:prbs
Item OVERHEAD_TEST_SELECT_TX_PATTERN ("Tx Overhead PRBS Test ...")
:source:data:telecom:overhead:prbs
Item OVERHEAD_TEST_SELECT_TX_INVERT ("Tx Overhead PRBS Test ...")
:source:data:telecom:overhead:prbs
Item OVERHEAD_TEST_SELECT_RX_BYTE ("Rx Overhead PRBS Test ...")
:sense:data:telecom:overhead:prbs
Item OVERHEAD_TEST_SELECT_RX_PATTERN ("Rx Overhead PRBS Test ...")
:sense:data:telecom:overhead:prbs
Item OVERHEAD_TEST_SELECT_RX_INVERT ("Rx Overhead PRBS Test ...")
:sense:data:telecom:overhead:prbs
=============================================




CTS 850 Cross-Reference listing Remote Commands 5
2.

----P_PERFORMANCE_ANALYSIS_G826_PARAM1----
Analysis Type .............. G.826

SDH Allocation .............
PDH Allocation .............
Include UAS ................ No Limit ..

Current Results
Near End Far End Near End Far End
RS PASS 140 PASS PASS
MS PASS PASS 34 PASS PASS
HP PASS PASS 8 PASS PASS
LP PASS PASS 2 PASS PASS
----------------------
Item PERFORMANCE_ANALYSIS_SELECT_ANALYSIS_G826_TYPE ("Analysis Type ...")
(No commands found)
Item SENSE_DATA_MEASURE_ANALYSIS_G826_SDH_ALLOCATION ("SDH Allocation ...")
(No commands found)
Item SENSE_DATA_MEASURE_ANALYSIS_G826_PDH_ALLOCATION ("PDH Allocation ...")
(No commands found)
Item SENSE_DATA_MEASURE_ANALYSIS_G826_UAS_ENABLE ("Include UAS ...")
:sense:data:telecom:measure:analysis:g826:uaseconds:enable
Item SENSE_DATA_MEASURE_ANALYSIS_G826_UAS_LIMIT ("Limit ...")
(No commands found)
=============================================




CTS 850 Cross-Reference listing Remote Commands 6
3.

----P_PERFORMANCE_ANALYSIS_M2101_PARAM1----
Analysis Type .............. M.2101.1

Test Type .................. BIS
Allocation .................
Include UAS ................ No Limit ..
APO Multiplier .............
Section ES . SES ..
Path ES . SES ..

Current Results Acceptable
Near End Far End
RS ACCEPTABLE
MS ACCEPTABLE ACCEPTABLE
HP ACCEPTABLE ACCEPTABLE
LP ACCEPTABLE ACCEPTABLE
----------------------
Item PERFORMANCE_ANALYSIS_SELECT_ANALYSIS_M2101_TYPE ("Analysis Type ...")
(No commands found)
Item SENSE_DATA_MEASURE_ANALYSIS_M2101_TEST_TYPE ("Test Type ...")
:sense:data:telecom:measure:analysis:m2101:test:type
Item SENSE_DATA_MEASURE_ANALYSIS_M2101_ALLOCATION_HIF ("Allocation ...")
(No commands found)
Item SENSE_DATA_MEASURE_ANALYSIS_M2101_UAS_ENABLE ("Include UAS ...")
:sense:data:telecom:measure:analysis:m2101:uaseconds:enable
Item SENSE_DATA_MEASURE_ANALYSIS_M2101_UAS_LIMIT_HIF ("Limit ...")
(No commands found)
Item SENSE_DATA_MEASURE_ANALYSIS_M2101_SECTION_ES_APOM_HIF ("ES ...")
(No commands found)
Item SENSE_DATA_MEASURE_ANALYSIS_M2101_SECTION_SES_APOM_HIF ("SES ...")
(No commands found)
Item SENSE_DATA_MEASURE_ANALYSIS_M2101_PATH_ES_APOM_HIF ("ES ...")
(No commands found)
Item SENSE_DATA_MEASURE_ANALYSIS_M2101_PATH_SES_APOM_HIF ("SES ...")
(No commands found)
=============================================




CTS 850 Cross-Reference listing Remote Commands 7
4.

----P_RESULTS_ANALYSIS_SDH_G826_B1BIP----
Section Analysis (B1 BIP)
Near End B1-BIP B2-BIP B3-BIP TU-BIP

Far End MS-REI HP-REI LP-REI
B1-BIP Analysis COUNT RATIO
Errored Blocks
Errored Seconds
Background Block Errors
Severely Errored Seconds
Consecutive SES Periods
Unavailable Seconds

Analysis PASS


----------------------
Item ..._B1BIP
(No commands found)
Item ..._B2BIP
(No commands found)
Item ..._MSREI
(No commands found)
Item ..._B3BIP
(No commands found)
Item ..._HPREI
(No commands found)
Item ..._TUBIP
(No commands found)
Item ..._LPREI
(No commands found)
=============================================




CTS 850 Cross-Reference listing Remote Commands 8
5.

----P_RESULTS_ANALYSIS_SDH_M2101_B1BIP----
Section Analysis (B1 BIP)
Near End B1-BIP B2-BIP B3-BIP TU-BIP

Far End MS-REI HP-REI LP-REI
B1-BIP Analysis (per G.829) COUNT RATIO
Errored Blocks
Errored Seconds
Background Block Errors
Severely Errored Seconds
Consecutive SES Periods
Unavailable Seconds

Analysis ACCEPTABLE


----------------------
Item ..._B1BIP
(No commands found)
Item ..._B2BIP
(No commands found)
Item ..._MSREI
(No commands found)
Item ..._B3BIP
(No commands found)
Item ..._HPREI
(No commands found)
Item ..._TUBIP
(No commands found)
Item ..._LPREI
(No commands found)
=============================================




CTS 850 Cross-Reference listing Remote Commands 9
6.

----P_RESULTS_RECALL_RESULTS_PARAM1----
Memory Start Time Durati
Current 00:00:00 01-Jan-00 Running...r>

Disk Description




<




----------------------
Item RESULTS0_NAME
(No commands found)
=============================================




CTS 850 Cross-Reference listing Remote Commands 10
7.

----P_RESULTS_SAVE_RESULTS_PARAM1----
The results of the last test are stored in memory.
For a permanent record, they must be saved to disk.

Name .......................
Description ................

Save To Disk ........... Select Action

Memory Start Time/Date Duration
Current 00:00:00 01-Jan-00 Running...




----------------------
Item ..._NAME ("Name ...")
(No commands found)
Item SENSE_DATA_MEASURE_INFORMATION_DESCRIPTION ("Description ...")
(No commands found)
Item ..._SAVE_DISK ("\011 Save To Disk ...")
(No commands found)
=============================================




CTS 850 Cross-Reference listing Remote Commands 11
8.

----P_RX_ANALYSIS_CONFIGURATION_SETTINGS_PARAM1----
FAS Error Threshold .....................
Pointer ss-Bit Mismatch Action .......... Loss of Pointer
Trace Mismatch Detection ................ Disabled
HP Unequipped Detection ................. Disabled
LP Unequipped Detection ................. Disabled
HP Signal Label Mismatch Detection.. Disabled
LP Signal Label Mismatch Detection... Disabled
Severely Errored Second
CRC Error Threshold .................... 300

Rx Error Trigger Output ................. None




----------------------
Item ..._FAS_THRESHOLD ("FAS Error Threshold ...")
(No commands found)
Item ..._POINTER_MISMATCH ("Pointer ss-Bit Mismatch Action ...")
:sense:data:telecom:measure:config:pointer:mismatch
Item ..._TRACE_MISMATCH ("Trace Mismatch Detection ...")
:sense:data:telecom:measure:config:trace:mismatch
Item SENSE_DATA_POVERHEAD_HPUNEQUIPPED ("HP Unequipped Detection ...")
:sense:data:telecom:poverhead:hpunequipped
Item SENSE_DATA_POVERHEAD_LPUNEQUIPPED ("LP Unequipped Detection ...")
:sense:data:telecom:poverhead:lpunequipped
Item SENSE_DATA_POVERHEAD_HPPLM ("HP Signal Label Mismatch Detection..")
:sense:data:telecom:poverhead:hpplm
Item SENSE_DATA_POVERHEAD_LPPLM ("LP Signal Label Mismatch Detection...")
:sense:data:telecom:poverhead:lpplm
Item ..._SES_THRESHOLD ("CRC Error Threshold ...")
:sense:data:telecom:measure:tributary:analysis:g826:threshold:m2:seseconds
:sense:data:telecom:measure:tributary:analysis:g826:threshold:m45:seseconds
Item ..._ERROR_TRIGGER ("Rx Error Trigger Output ...")
:sense:data:telecom:erroutput:source
=============================================




CTS 850 Cross-Reference listing Remote Commands 12
9.

----P_RX_CAS_VOICE_SETTINGS_PARAM1----
CAS Display Mode............ Single
Selected Timeslot...........
Speaker..................... Off

TS abcd
CAS Bit Values..............




----------------------
Item ..._CAS_MODE ("CAS Display Mode...")
:sense:data:telecom:cas:mode
Item ..._TIMESLOT ("Selected Timeslot...")
(No commands found)
Item ..._VOLUME ("Speaker...")
:sense:data:telecom:tributary:k64:speaker
=============================================




CTS 850 Cross-Reference listing Remote Commands 13
10.

----P_RX_CAS_VOICE_SETTINGS_PARAM2----
CAS Display Mode............ Single

TS abcd TS abcd TS abcd TS abcd
1 9 17 25
2 10 18 26
3 11 19 27
4 12 20 28
5 13 21 29
6 14 22 30
7 15 23 31
8 16 ---- 24




----------------------
Item ..._CAS_MODE ("CAS Display Mode...")
:sense:data:telecom:cas:mode
=============================================




CTS 850 Cross-Reference listing Remote Commands 14
11.

----P_RX_JITTER_WANDER_PARAM1----
Jitter Mode ................ Peak-Peak
Jitter Input Source ........ Line
Jitter Input Clock Rate .... 52MHz (STM-0)
Jitter Range ............... Normal (6UI)
Jitter Measurement Filter .. Wideband
(100Hz - 400kHz)
Fullband Highpass .......... 10.0Hz
Jitter Hit Threshold .......
Pointer Hit Threshold ......




----------------------
Item RX_JITTER_WANDER_MODE ("Jitter Mode ...")
:sense:data:telecom:jitter:mode
Item RX_JITTER_WANDER_SOURCE ("Jitter Input Source ...")
:sense:data:telecom:jitter:source
Item RX_JITTER_WANDER_CLOCK_RATE ("Jitter Input Clock Rate ...")
:sense:data:telecom:jitter:clock:rate
Item RX_JITTER_WANDER_RANGE ("Jitter Range ...")
:sense:data:telecom:jitter:range
Item RX_JITTER_WANDER_FILTER_TYPE ("Jitter Measurement Filter ...")
:diagnostic:select
:sense:data:telecom:jitter:filter:highband
:sense:data:telecom:jitter:filter:type
Item RX_JITTER_WANDER_FILTER_SUBBAND_HPASS ("Fullband Highpass ...")
:sense:data:telecom:jitter:filter:fullband:hpass
Item RX_JITTER_WANDER_HIT_THRESHOLD ("Jitter Hit Threshold ...")
(No commands found)
Item RX_JITTER_WANDER_PTR_EVENT_THRESHOLD ("Pointer Hit Threshold ...")
(No commands found)
=============================================




CTS 850 Cross-Reference listing Remote Commands 15
12.

----P_RX_PATH_OVERHEAD_PARAM1----
Overhead View .............. STS Path Overhead
External Drop .............. None
Pause Control .............. Updates Active
C2 Signal Label ............ (Undefined)
J1 HP Trace Format ......... 64-byte

Path Overhead Path Trace Message
J1: B3:
C2:
G1:
F2:
H4:
Z3:
Z4:
Z5:
----------------------
Item ..._DISPLAY ("Overhead View ...")
(No commands found)
Item ..._EXTERNAL_DROP ("External Drop ...")
:sense:data:telecom:overhead:drop
:sense:data:telecom:poverhead:drop
Item RX_PATH_OVERHEAD_CONTROL ("Pause Control ...")
:initiate
Item RX_PATH_OVERHEAD_J1_TRACE_FORMAT ("J1 HP Trace Format ...")
:sense:data:telecom:poverhead:trace:format
=============================================




CTS 850 Cross-Reference listing Remote Commands 16
13.

----P_RX_PATH_OVERHEAD_V5_PARAM1----
Receiving STS-1 with STS-1 : 1 under test
Overhead View .............. STS Path Overhead
Pause Control .............. Updates Active

Overhead
V5:




----------------------
Item RX_PATH_OVERHEAD_SELECT_V5
(No commands found)
Item RX_PATH_OVERHEAD_V5_CONTROL ("Pause Control ...")
:initiate
=============================================




CTS 850 Cross-Reference listing Remote Commands 17
14.

----P_RX_PATH_OVERHEAD_VC12_PARAM1----
Overhead View .............. STS Path Overhead

Pause Control .............. Updates Active
V5 Signal Label ............ (Unequipped)

Path Overhead Path Trace Message
V5:
J2: N2:
K4:




----------------------
Item RX_PATH_OVERHEAD_SELECT_VC12 ("Overhead View ...")
(No commands found)
Item RX_PATH_OVERHEAD_VC12_CONTROL ("Pause Control ...")
:initiate
=============================================




CTS 850 Cross-Reference listing Remote Commands 18
15.

----P_RX_PATH_OVERHEAD_VC3_PARAM1----
Overhead View .............. STS Path Overhead

Pause Control .............. Updates Active
C2 Signal Label ............

Path Overhead Path Trace Message
J1: B3:
C2:
G1:
F2:
H4:
Z3:
Z4:
Z5:

----------------------
Item RX_PATH_OVERHEAD_SELECT_VC3 ("Overhead View ...")
(No commands found)
Item RX_PATH_OVERHEAD_VC3_CONTROL ("Pause Control ...")
:initiate
=============================================




CTS 850 Cross-Reference listing Remote Commands 19
16.

----P_RX_SDH_K64_SETTINGS_PARAM1----
2 Mb/s Framing.............. Framed
Configuration............... N x 64k
Number of Timeslots.........
Starting Timeslot...........
Test Pattern................ 2^23-1 Normal (ITU)

Active Timeslots

<< < << < << < < < < << < << < << < << < < << << < < <
1 5 10 15 20 25 30




----------------------
Item ..._FRAMING ("2 Mb/s Framing...")
:sense:data:telecom:tributary:framing
Item ..._CONFIGURATION ("Configuration...")
(No commands found)
Item ..._WIDTH ("Number of Timeslots...")
(No commands found)
Item ..._TIMESLOT ("Starting Timeslot...")
(No commands found)
Item ..._TEST_PATTERN ("Test Pattern...")
:sense:data:telecom:payload:pattern
:sense:data:telecom:tributary:pattern
Item ..._TEST_PATTERN_POLARITY ("Test Pattern...")
:sense:data:telecom:payload:prbs:invert
Item ..._UBYTE ("Test Pattern...")
:sense:data:telecom:payload:pattern
:sense:data:telecom:tributary:pattern
=============================================




CTS 850 Cross-Reference listing Remote Commands 20
17.

----P_RX_SDH_SETTINGS_PARAM1----
Receive Rate ............... STM-4 Optical
Receive Level .............. Cross Connect
AU Under Test .............. 1
SDH Structure .............. AU-4 Bulk
TU Under Test .............. :1 (Group 1, TUG-2:1 TU-12:1
Payload Structure .......... 140 Mb/suipped)Framed
Test Pattern ............. 2^23-1 Normal (ITU)
34Mb/s Active Channel..... 1
8Mb/s Active Channel..... 1
2Mb/s Active Channel..... 1
Payload Drop................ Drop DS1 (AMI)
Tx/Rx Setup................. Independent




----------------------
Item ..._RATE ("Receive Rate ...")
:input1:telecom:rate
:input1:telecom:type
:input3:telecom:rate
:sense:data:telecom:source
Item ..._QUALIFIER ("Receive Rate ...")
:input1:telecom:type
Item ..._LEVEL ("Receive Level ...")
:input1:telecom:level
Item ..._AU ("AU Under Test ...")
(No commands found)
Item ..._STRUCTURE ("SDH Structure ...")
:sense:data:telecom:payload:mapping
:sense:data:telecom:structure
:sense:data:telecom:tributary:mapping
Item ..._TU_GROUP ("TU Under Test ...")
(No commands found)
Item ..._TU2 ("TU Under Test ...")
(No commands found)
Item ..._TU12 ("TU Under Test ...")
(No commands found)
Item ..._BULK_PAYLOAD ("Payload Structure ...")
:sense:data:telecom:payload:mapping
Item ..._TU_PAYLOAD ("Payload Structure ...")
:sense:data:telecom:tributary:payload
Item ..._TU_PAYLOAD_FRAMING ("Payload Structure ...")
:sense:data:telecom:tributary:framing
Item ..._TEST_PATTERN ("Test Pattern ...")
:sense:data:telecom:payload:pattern
:sense:data:telecom:tributary:pattern
Item ..._TEST_PATTERN_POLARITY ("Test Pattern ...")
:sense:data:telecom:payload:prbs:invert
Item ..._UBYTE ("Test Pattern ...")
:sense:data:telecom:payload:pattern
:sense:data:telecom:tributary:pattern
Item ..._M34_CHANNEL ("34Mb/s Active Channel...")
:sense:data:telecom:tributary:dmx34:channel
Item ..._M8_CHANNEL ("8Mb/s Active Channel...")
:sense:data:telecom:tributary:dmx8:channel
Item ..._M2_CHANNEL ("2Mb/s Active Channel...")
:sense:data:telecom:tributary:dmx2:channel
Item ..._PAYLOAD_DROP ("Payload Drop...")
:output2:telecom:code
:output3:telecom:code
:sense:data:telecom:tributary:drop
Item ..._TX_RX ("Tx/Rx Setup...")
:instrument:coupling



CTS 850 Cross-Reference listing Remote Commands 21
:source:data:telecom:source
=============================================




CTS 850 Cross-Reference listing Remote Commands 22
18.

----P_RX_SDH_SETTINGS_TRIB_PARAM1----
Receive Rate ............... STM-4 Balanced
Receive Level............... Normal (Cross Connect)

Payload Structure .......... 140 Mb/s Framed
Test Pattern ............. 2^23-1 Normal (ITU)
34Mb/s Active Channel..... 1
8Mb/s Active Channel..... 1
2Mb/s Active Channel..... 1
Payload Drop................ Drop DS1 (AMI)
Tx/Rx Setup................. Independent




----------------------
Item ..._RATE ("Receive Rate ...")
:input1:telecom:rate
:input1:telecom:type
:input3:telecom:rate
:sense:data:telecom:source
Item ..._QUALIFIER ("Receive Rate ...")
:input2:telecom:code
:input2:telecom:terminator
:input3:telecom:code
Item ..._LEVEL ("Receive Level...")
:input2:telecom:level
:input3:telecom:level
Item ..._PAYLOAD ("Payload Structure ...")
:sense:data:telecom:tributary:payload
Item ..._PAYLOAD_FRAMING ("Payload Structure ...")
:sense:data:telecom:tributary:framing
Item ..._TEST_PATTERN ("Test Pattern ...")
:sense:data:telecom:payload:pattern
:sense:data:telecom:tributary:pattern
Item ..._TEST_PATTERN_POLARITY ("Test Pattern ...")
:sense:data:telecom:payload:prbs:invert
Item ..._UBYTE ("Test Pattern ...")
:sense:data:telecom:payload:pattern
:sense:data:telecom:tributary:pattern
Item ..._M34_CHANNEL ("34Mb/s Active Channel...")
:sense:data:telecom:tributary:dmx34:channel
Item ..._M8_CHANNEL ("8Mb/s Active Channel...")
:sense:data:telecom:tributary:dmx8:channel
Item ..._M2_CHANNEL ("2Mb/s Active Channel...")
:sense:data:telecom:tributary:dmx2:channel
Item ..._PAYLOAD_DROP ("Payload Drop...")
:output2:telecom:code
:output3:telecom:code
:sense:data:telecom:tributary:drop
Item ..._TX_RX ("Tx/Rx Setup...")
:instrument:coupling
:source:data:telecom:source
=============================================




CTS 850 Cross-Reference listing Remote Commands 23
19.

----P_RX_SETTINGS_PARAM1----
Receive Rate ............... STS-1
Receive Level .............. Cross Connect
Signal Structure ........... STS-1
STS Under Test ............. 1
Mapping .................... Bulk Fill (Equipped)
DS3 Framing ................ DS3 C-bit 1, VT 1) TU-12:1
DS1 Under Test ............. DS1 1nframed
DS1 Framing ................ DS1 Unframed
Tributary Drop ............. Drop DS1 (AMI)
Test Pattern ............... 2^23-1
Tx/Rx Settings ............. Independent




----------------------
Item ..._RATE ("Receive Rate ...")
:input1:telecom:rate
:input1:telecom:type
:input2:telecom:terminator
:input3:telecom:rate
:sense:data:telecom:source
Item ..._LEVEL ("Receive Level ...")
:input1:telecom:level
Item ..._SIGNAL_STRUCTURE ("Signal Structure ...")
:sense:data:telecom:structure
Item ..._SPE ("STS Under Test ...")
(No commands found)
Item ..._SPE_MAPPING ("Mapping ...")
:sense:data:telecom:payload:mapping
:sense:data:telecom:tributary:mapping
Item ..._VT_GROUP ("DS3 Framing ...")
(No commands found)
Item ..._TU_TU12_CHANNEL ("DS3 Framing ...")
(No commands found)
Item ..._VT_FRAMING ("DS1 Under Test ...")
:sense:data:telecom:tributary:framing
Item ..._DEMAP_FRAMING ("DS3 Framing ...")
:sense:data:telecom:tributary:framing
Item ..._DEMUX_CHAN_READOUT ("DS1 Under Test ...")
(No commands found)
Item SENSE_DATA_TRIBUTARY_DEMUX_FRAMING ("DS1 Framing ...")
:sense:data:telecom:tributary:demultiplex:framing
Item ..._TRIBUTARY_DROP ("Tributary Drop ...")
:output2:telecom:code
:output3:telecom:code
:sense:data:telecom:tributary:drop
Item ..._TEST_PATTERN ("Test Pattern ...")
:sense:data:telecom:payload:pattern
:sense:data:telecom:tributary:pattern
Item ..._UBYTE ("Test Pattern ...")
:sense:data:telecom:payload:pattern
:sense:data:telecom:tributary:pattern
Item ..._TX_RX ("Tx/Rx Settings ...")
:instrument:coupling
:source:data:telecom:source
=============================================




CTS 850 Cross-Reference listing Remote Commands 24
20.

----P_RX_SETTINGS_TRIB_PARAM1----
Receive Rate ............... STS-1
Receive Level .............. Normal (Cross Connect)

DS3 Framing ................ DS3 C-bitmed
DS1 Under Test ............. DS1 1
DS1 Framing ................ DS1 Unframed

Test Pattern ............... 2^23-1
Tx/Rx Settings ............. Independent




----------------------
Item RX_SETTINGS_SELECT_TRIB_RATE ("Receive Rate ...")
:input1:telecom:rate
:input1:telecom:type
:input2:telecom:terminator
:input3:telecom:rate
:sense:data:telecom:source
Item RX_SETTINGS_SELECT_TRIBUTARY_LEVEL ("Receive Level ...")
:input2:telecom:level
:input3:telecom:level
Item RX_SETTINGS_SELECT_TRIBUTARY_FRAMING ("Framing ...")
:sense:data:telecom:tributary:framing
Item RX_SETTINGS_SELECT_TRIBUTARY_DEMAP_FRAMING ("Framing ...")
:sense:data:telecom:tributary:framing
Item RX_SETTINGS_SELECT_TRIBUTARY_DEMUX_CHAN_READOUT ("DS1 Under Test ...")
(No commands found)
Item RX_SETTINGS_SELECT_TRIBUTARY_DEMUX_FRAMING ("DS1 Framing ...")
:sense:data:telecom:tributary:demultiplex:framing
Item RX_SETTINGS_SELECT_TRIB_TEST_PATTERN ("Test Pattern ...")
:sense:data:telecom:payload:pattern
:sense:data:telecom:tributary:pattern
Item RX_SETTINGS_SELECT_TRIB_UBYTE ("Test Pattern ...")
:sense:data:telecom:payload:pattern
:sense:data:telecom:tributary:pattern
Item RX_SETTINGS_SELECT_TRIB_TX_RX ("Tx/Rx Settings ...")
:instrument:coupling
:source:data:telecom:source
=============================================




CTS 850 Cross-Reference listing Remote Commands 25
21.

----P_RX_SET_BYTE_TRACE_PARAM1----




----------------------
=============================================




CTS 850 Cross-Reference listing Remote Commands 26
22.

----P_RX_TRANSPORT_OVERHEAD_PARAM1----
Showing Overhead for ....... STS-1 :1
External Drop .............. None
Pause Control .............. Updates Active
J0 RS Trace ................ Tek CTS 850
S1 Sync. Status Message ....
Column ? Column ? Column ?
A1: A2: J0:
B1: E1: F1:
D1: D2: D3:
H1: H2: H3:
B2: K1: K2:
D4: D5: D6:
D7: D8: D9:
D10: D11: D12:
S1: Z2: E2:

----------------------
Item ..._COLUMNS_HIF ("Showing Overhead for ...")
(No commands found)
Item ..._EXTERNAL_DROP ("External Drop ...")
:sense:data:telecom:overhead:drop
:sense:data:telecom:poverhead:drop
Item RX_TRANSPORT_OVERHEAD_CONTROL ("Pause Control ...")
:initiate
Item RX_TRANSPORT_OVERHEAD_TRACE_J0 ("J0 RS Trace ...")
:sense:data:telecom:section:trace:mode
=============================================




CTS 850 Cross-Reference listing Remote Commands 27
23.

----P_RX_VIEW_BYTE_TRACE_PARAM1----




----------------------
=============================================




CTS 850 Cross-Reference listing Remote Commands 28
24.

----P_SIGNAL_LABELS_PARAM1----
C2 HP-Signal Label Measurement: Disabled
Expected ......
Rx ............ (Undefined)
Tx ............
C2 LP-Signal Label Measurement: Disabled
Expected ......
Rx.............
Tx.............
V5 LP-Signal Label Measurement: Disabled
Expected ...... 000 (Unequipped)
Rx............. Tx............. 000




----------------------
Item TX_SIGNAL_LABELS_C2_HP_EXPECTED_VALUE ("Expected ...")
(No commands found)
Item TX_SIGNAL_LABELS_C2_LP_EXPECTED_VALUE ("Expected ...")
(No commands found)
Item TX_SIGNAL_LABELS_V5_LP_EXPECTED_VALUE ("Expected ...")
(No commands found)
=============================================




CTS 850 Cross-Reference listing Remote Commands 29
25.

----P_TEST_JITTER_OUTPUT_PARAM1----
Test Type .................. Output Jitter
Test Control ............... Edit Setup
Test Duration .............. 1 min 0 sec

No Data


Measured Max Allowed
Filter Jitter (G.823)
Wideband
Highband

Press START/STOP to Begin Jitter Transfer Test



----------------------
Item ..._TYPE ("Test Type ...")
:sense:data:telecom:test:jitter:type
Item ..._CONTROL ("Test Control ...")
:sense:data:telecom:test:jitter:tolerance:alldata
Item ..._DURATION ("Test Duration ...")
(No commands found)
=============================================




CTS 850 Cross-Reference listing Remote Commands 30
26.

----P_TEST_JITTER_POINTER_PARAM1----
Test Type .................. Output Jitter
Test Control ............... Edit Setup
Sequence Type (G.783) ...... a) Single Alternating

No Data


Measured Max Allowed
Filter Jitter (G.783)
Wideband
Highband

(Go to Transmit POINTERS & TIMING to setup sequence)
Press START/STOP to Begin Jitter Transfer Test


----------------------
Item ..._TYPE ("Test Type ...")
:sense:data:telecom:test:jitter:type
Item ..._CONTROL ("Test Control ...")
:sense:data:telecom:test:jitter:tolerance:alldata
Item SENSE_DATA_TEST_JITTER_POINTER_SEQUENCE_TYPE ("Sequence Type (G.783) ...")
:sense:data:telecom:test:jitter:pointer:sequence:type
=============================================




CTS 850 Cross-Reference listing Remote Commands 31
27.

----P_TEST_JITTER_RECALL_RESULTS----
Test Type .................. Output Jitter
Test Control ............... Edit Setup
Choose Test ................ Select Action

Disk Description




<




----------------------
Item TEST_SETUP_JITTER_RECALL_RESULTS_TYPE ("Test Type ...")
:sense:data:telecom:test:jitter:type
Item TEST_SETUP_JITTER_RECALL_RESULTS ("Test Control ...")
:sense:data:telecom:test:jitter:tolerance:alldata
Item TEST_JITTER_CHOOSE_TEST ("Choose Test ...")
(No commands found)
Item JITTER_RESULTS0_FILENAME
(No commands found)
Item JITTER_RESULTS1_FILENAME
(No commands found)
Item JITTER_RESULTS2_FILENAME
(No commands found)
Item JITTER_RESULTS3_FILENAME
(No commands found)
Item JITTER_RESULTS4_FILENAME
(No commands found)
Item JITTER_RESULTS5_FILENAME
(No commands found)
=============================================




CTS 850 Cross-Reference listing Remote Commands 32
28.

----P_TEST_JITTER_SAVE_RESULTS----
Test Type .................. Output Jitter
Test Control ............... Edit Setup
File Name ..................
Description ................
Disk Operation ............. Select Action




----------------------
Item TEST_SETUP_JITTER_SAVE_RESULTS_TYPE ("Test Type ...")
:sense:data:telecom:test:jitter:type
Item TEST_SETUP_JITTER_SAVE_RESULTS ("Test Control ...")
:sense:data:telecom:test:jitter:tolerance:alldata
Item TEST_JITTER_SAVE_RESULTS_NAME ("File Name ...")
(No commands found)
Item TEST_JITTER_SAVE_RESULTS_DESCRIPTION ("Description ...")
(No commands found)
Item TEST_JITTER_SAVE_RESULTS_DISK ("Disk Operation ...")
(No commands found)
=============================================




CTS 850 Cross-Reference listing Remote Commands 33
29.

----P_TEST_JITTER_SPECTRUM_PARAM1----
Test Type .................. Output Jitter
Test Control ............... Edit Setup
Frequency Range ............ Spectral 0.1Hz - 1kHz

Press START/STOP to Begin Jitter Transfer Test




----------------------
Item ..._TYPE ("Test Type ...")
:sense:data:telecom:test:jitter:type
Item ..._CONTROL ("Test Control ...")
:sense:data:telecom:test:jitter:tolerance:alldata
Item SENSE_DATA_TEST_JITTER_SPECTRUM_RANGE ("Frequency Range ...")
(No commands found)
=============================================




CTS 850 Cross-Reference listing Remote Commands 34
30.

----P_TEST_JITTER_TOLERANCE_PARAM1----
Test Type .................. Output Jitter
Test Control ............... Edit Setup
Mask Type .................. G.825
Start Frequency ............
End Frequency ..............
Number Freq Samples ........
Tolerance Criteria ......... Onset Of Errors Method
Record Error Threshold ..... SELECT ACTION
Recommended BER (O.171) >
Current BER (Threshold) In Progress...

Press START/STOP to Begin Jitter Transfer Test




----------------------
Item SENSE_DATA_TEST_JITTER_TYPE ("Test Type ...")
:sense:data:telecom:test:jitter:type
Item SENSE_DATA_TEST_JITTER_CONTROL ("Test Control ...")
:sense:data:telecom:test:jitter:tolerance:alldata
Item SENSE_DATA_TEST_JITTER_TOLERANCE_MASK ("Mask Type ...")
:sense:data:telecom:test:jitter:tolerance:mask
Item SENSE_DATA_TEST_JITTER_TOLERANCE_FREQUENCY_START ("Start Frequency ...")
(No commands found)
Item SENSE_DATA_TEST_JITTER_TOLERANCE_FREQUENCY_END ("End Frequency ...")
(No commands found)
Item SENSE_DATA_TEST_JITTER_TOLERANCE_SAMPLES ("Number Freq Samples ...")
(No commands found)
Item SENSE_DATA_TEST_JITTER_TOLERANCE_METHOD ("Tolerance Criteria ...")
:sense:data:telecom:test:jitter:tolerance:method
Item SENSE_DATA_TEST_JITTER_TOLERANCE_MEASURE ("Record Error Threshold ...")
:sense:data:telecom:test:jitter:tolerance:measure
=============================================




CTS 850 Cross-Reference listing Remote Commands 35
31.

----P_TEST_JITTER_TRANSFER_PARAM1----
Test Type .................. Output Jitter
Test Control ............... Edit Setup
Mask Type .................. GR-253
Start Frequency ............
End Frequency ..............
Number Freq Samples ........
Action ..................... Perform Jitter Transfer

The calibration is performed to establish a 0dB amplitude
reference trace of the CTS (using an internal loopback).

Calibration Status: Unsupported Settings

Press START/STOP to Begin Jitter Transfer Test


----------------------
Item ..._TYPE ("Test Type ...")
:sense:data:telecom:test:jitter:type
Item ..._CONTROL ("Test Control ...")
:sense:data:telecom:test:jitter:tolerance:alldata
Item SENSE_DATA_TEST_JITTER_TRANSFER_MASK ("Mask Type ...")
:sense:data:telecom:test:jitter:transfer:mask
Item SENSE_DATA_TEST_JITTER_TRANSFER_FREQUENCY_START ("Start Frequency ...")
(No commands found)
Item SENSE_DATA_TEST_JITTER_TRANSFER_FREQUENCY_END ("End Frequency ...")
(No commands found)
Item SENSE_DATA_TEST_JITTER_TRANSFER_SAMPLES ("Number Freq Samples ...")
(No commands found)
Item SENSE_DATA_TEST_JITTER_TRANSFER_STATE ("Action ...")
:sense:data:telecom:test:jitter:transfer:state
=============================================




CTS 850 Cross-Reference listing Remote Commands 36
32.

----P_TEST_SETUP_RECALL_INSTRUMENT_SETUPS_PARAM1----
Memory Description
0. DEFAULT FACTORY SETTINGS
1. EMPTY
2. EMPTY
3. EMPTY
4. EMPTY
5. EMPTY

Disk Description




<

----------------------
Item INSTRUMENT_SETUP0_NAME ("0.")
(No commands found)
Item INSTRUMENT_SETUP1_NAME ("1.")
(No commands found)
Item INSTRUMENT_SETUP2_NAME ("2.")
(No commands found)
Item INSTRUMENT_SETUP3_NAME ("3.")
(No commands found)
Item INSTRUMENT_SETUP4_NAME ("4.")
(No commands found)
Item INSTRUMENT_SETUP5_NAME ("5.")
(No commands found)
=============================================




CTS 850 Cross-Reference listing Remote Commands 37
33.

----P_TEST_SETUP_RECALL_PASS_FAIL_TESTS_PARAM1----
Disk Description




<


----------------------
Item PASS_FAIL_TEST0_NAME
(No commands found)
Item PASS_FAIL_TEST1_NAME
(No commands found)
Item PASS_FAIL_TEST2_NAME
(No commands found)
Item PASS_FAIL_TEST3_NAME
(No commands found)
Item PASS_FAIL_TEST4_NAME
(No commands found)
Item PASS_FAIL_TEST5_NAME
(No commands found)
Item PASS_FAIL_TEST6_NAME
(No commands found)
Item PASS_FAIL_TEST7_NAME
(No commands found)
Item PASS_FAIL_TEST8_NAME
(No commands found)
Item PASS_FAIL_TEST9_NAME
(No commands found)
Item PASS_FAIL_TEST10_NAME
(No commands found)
Item PASS_FAIL_TEST11_NAME
(No commands found)
Item PASS_FAIL_TEST12_NAME
(No commands found)
=============================================




CTS 850 Cross-Reference listing Remote Commands 38
34.

----P_TEST_SETUP_SAVE_INSTRUMENT_SETUPS_PARAM1----
Set up the instrument as required. Then fill out the
parameters below and save to disk or memory.

Name .......................
Description ................

Save to Memory ......... Select Action
Save to Disk ........... Select Action




----------------------
Item ..._NAME ("Name ...")
(No commands found)
Item SYSTEM_DESCRIPTION ("Description ...")
(No commands found)
Item ..._SAVE_MEM ("\012 Save to Memory ...")
(No commands found)
Item ..._SAVE_DISK ("\011 Save to Disk ...")
(No commands found)
=============================================




CTS 850 Cross-Reference listing Remote Commands 39
35.

----P_TEST_SETUP_SAVE_PASS_FAIL_TESTS_PARAM1----
Set up the instrument as required. Then fill out the
parameters below and save your test to disk.

Name .......................
Description ................
Operator Start Prompt ......
Test Duration ..............


Fail If


Operator End Prompt ........
On Test Completion ......... Do Nothing
Save to Disk ........... Select Action

----------------------
Item TEST_SETUP_SAVE_PASS_FAIL_TESTS_NAME ("Name ...")
(No commands found)
Item SENSE_DATA_MEASURE_STESTS_DESCRIPTION ("Description ...")
(No commands found)
Item SENSE_DATA_MEASURE_STESTS_START_PROMPT ("Operator Start Prompt ...")
(No commands found)
Item TEST_SETUP_SAVE_PASS_FAIL_TESTS_DURATION ("Test Duration ...")
(No commands found)
Item TEST_SETUP_SAVE_PASS_FAIL_TESTS_EVAL_CLASS1
:sense:data:telecom:measure:stests:threshold
Item TEST_SETUP_SAVE_PASS_FAIL_TESTS_EVAL_TYPE1
(No commands found)
Item TEST_SETUP_SAVE_PASS_FAIL_TESTS_EVAL_THRESHOLD1
(No commands found)
Item TEST_SETUP_SAVE_PASS_FAIL_TESTS_EVAL_CLASS2
:sense:data:telecom:measure:stests:threshold
Item TEST_SETUP_SAVE_PASS_FAIL_TESTS_EVAL_TYPE2
(No commands found)
Item TEST_SETUP_SAVE_PASS_FAIL_TESTS_EVAL_THRESHOLD2
(No commands found)
Item TEST_SETUP_SAVE_PASS_FAIL_TESTS_EVAL_CLASS3
:sense:data:telecom:measure:stests:threshold
Item TEST_SETUP_SAVE_PASS_FAIL_TESTS_EVAL_TYPE3
(No commands found)
Item TEST_SETUP_SAVE_PASS_FAIL_TESTS_EVAL_THRESHOLD3
(No commands found)
Item TEST_SETUP_SAVE_PASS_FAIL_TESTS_EVAL_CLASS4
:sense:data:telecom:measure:stests:threshold
Item TEST_SETUP_SAVE_PASS_FAIL_TESTS_EVAL_TYPE4
(No commands found)
Item TEST_SETUP_SAVE_PASS_FAIL_TESTS_EVAL_THRESHOLD4
(No commands found)
Item SENSE_DATA_MEASURE_STESTS_END_PROMPT ("Operator End Prompt ...")
(No commands found)
Item SENSE_DATA_MEASURE_STESTS_ON_COMPLETE ("On Test Completion ...")
:sense:data:telecom:measure:stests:done
Item TEST_SETUP_SAVE_PASS_FAIL_TESTS_DISK_ACTION ("\011 Save to Disk ...")
(No commands found)
=============================================




CTS 850 Cross-Reference listing Remote Commands 40
36.

----P_TEST_SETUP_TEST_CONTROL_PARAM1----
Test Duration ..............
History Resolution ......... Normal (1 min samples)

At 1 minute resolution the maximum history length
is at least 5 days. Upon reaching the 5 day limit
new data will begin overwriting the old data.




----------------------
Item UTILITY_TEST_DURATION ("Test Duration ...")
(No commands found)
Item ..._HIST_RES ("History Resolution ...")
:sense:data:telecom:test:history:resolution
=============================================




CTS 850 Cross-Reference listing Remote Commands 41
37.

----P_TRACE_SETTINGS_PARAM1----
Trace Select ... RS Trace (J0) Measurement: Disabled

Expected .....


Rx ...........


Tx ...........




----------------------
Item OVERHEAD_TEST_SELECT_TRACE ("Trace Select ...")
(No commands found)
Item OVERHEAD_TEST_EXPECTED_TRACE_1
(No commands found)
=============================================




CTS 850 Cross-Reference listing Remote Commands 42
38.

----P_TX_APS_COMMANDS_PARAM1----
APS Mode ................... Span Network
USER SETUP TRANSMIT RECEIVE
K1 Full Byte ...............
Bits 1-4: Switch Request ... ? ? ?
Bits 5-8: Channel Requesting ? ? ?
K2 Full Byte ...............
Bits 1-4: Bridged Channel .. ? ? ?
Bit 5: Architecture ........ ? ? ?
Bits 6-8: Status ........... ? ? ?

Transmit Setup ............. Select Action




----------------------
Item SOURCE_DATA_APS_MODE ("APS Mode ...")
(No commands found)
Item TX_APS_K1_BYTE ("K1 Full Byte ...")
(No commands found)
Item TX_APS_K1_BITS_1_4 ("Bits 1-4: Switch Request ...")
(No commands found)
Item TX_APS_K1_BITS_5_8 ("Bits 5-8: Channel Requesting ...")
(No commands found)
Item TX_APS_K2_BYTE ("K2 Full Byte ...")
(No commands found)
Item TX_APS_K2_BITS_1_4 ("Bits 1-4: Bridged Channel ...")
(No commands found)
Item TX_APS_K2_BIT_5 ("Bit 5: Architecture ...")
(No commands found)
Item TX_APS_K2_BITS_6_8 ("Bits 6-8: Status ...")
(No commands found)
Item TX_APS_TRANSMIT_SETUP ("Transmit Setup ...")
(No commands found)
=============================================




CTS 850 Cross-Reference listing Remote Commands 43
39.

----P_TX_ERRORS_ALARMS_PARAM1----
Error Type set to .......... Section B1 BIP
Frame Bit Error Burst ...... Error in Frames
PDH Anomaly Layer .......... 140 Mb/s

Transmit Alarm set to ...... None
PDH Alarm Layer ............ 140 Mb/s

Transmit Failure set to .... None
PDH Failure Layer .......... 140 Mb/s
Press INSERT ERROR to insert a single error
into the transmitted signal.




----------------------
Item ..._TYPE ("Error Type set to ...")
:source:data:telecom:error:enable
:source:data:telecom:error:type
:source:data:telecom:tributary:error
Item ..._RATE ("Error Rate set to ...")
:source:data:telecom:error:enable
Item ..._FAS_BURST_COUNT ("Error Rate set to ...")
(No commands found)
Item ..._FAS_BURST_SIZE ("Error in")
(No commands found)
Item ..._ERROR_LAYER ("PDH Anomaly Layer ...")
:source:data:telecom:tributary:anomaly:layer
Item ..._ALARM ("Transmit Alarm set to ...")
:source:data:telecom:alarm
:source:data:telecom:tributary:alarm
Item ..._ALARM_LAYER ("PDH Alarm Layer ...")
:source:data:telecom:tributary:defect:layer
Item ..._FAILURE ("Transmit Failure set to ...")
:source:data:telecom:failure:type
:source:data:telecom:tributary:failure
Item ..._FAILURE_LAYER ("PDH Failure Layer ...")
:source:data:telecom:tributary:fault:layer
=============================================




CTS 850 Cross-Reference listing Remote Commands 44
40.

----P_TX_JITTER_PARAM1----
Jitter/Wander Generation ... Off
Jitter Output .............. Line
Jitter Clock Rate .......... 52MHz (STM-0)
Jitter Clock Offset ........
Jitter Frequency ...........
Jitter Amplitude ...........

For Jitter Generation Set Jitter Frequency > 10 Hz
For Wander Generation Set Jitter Frequency < 10 Hz




----------------------
Item TX_JITTER_MODE ("Jitter/Wander Generation ...")
:source:data:telecom:jitter:mode
Item TX_JITTER_SOURCE ("Jitter Output ...")
:source:data:telecom:jitter:source
Item TX_JITTER_CLOCK_RATE ("Jitter Clock Rate ...")
:source:data:telecom:jitter:clock:rate
Item TX_JITTER_CLOCK_OFFSET ("Jitter Clock Offset ...")
(No commands found)
Item TX_JITTER_FREQUENCY ("Jitter Frequency ...")
:diagnostic:select
:sense:data:telecom:jitter:filter:highband
:sense:data:telecom:jitter:filter:type
Item TX_JITTER_AMPLITUDE ("Jitter Amplitude ...")
(No commands found)
=============================================




CTS 850 Cross-Reference listing Remote Commands 45
41.

----P_TX_PATH_OVERHEAD_PARAM1----
Transmitting STS-1 : 1 under test
Overhead View .............. STS Path Overhead
External Add ............... None
C2 Signal Label ............
H4 Multiframe Sequence ..... 4 Bytes

Path Overhead Path Trace Message
J1: B3:
C2:
G1:
F2:
H4:
Z3:
Z4:
Z5:
----------------------
Item ..._DISPLAY ("Overhead View ...")
(No commands found)
Item ..._EXTERNAL_ADD ("External Add ...")
:source:data:telecom:overhead:insert
:source:data:telecom:poverhead:insert
Item TX_PATH_C2_OVERHEAD_LABEL ("C2 Signal Label ...")
(No commands found)
Item TX_PATH_OVERHEAD_H4_SEQUENCE ("H4 Multiframe Sequence ...")
(No commands found)
Item SOURCE_DATA_POVERHEAD_BYTENAME_J1
:source:data:telecom:overhead:preset
:source:data:telecom:poverhead:preset
Item SOURCE_DATA_POVERHEAD_BYTENAME_B3
(No commands found)
Item SOURCE_DATA_POVERHEAD_BYTENAME_C2
:source:data:telecom:overhead:preset
:source:data:telecom:poverhead:preset
Item SOURCE_DATA_POVERHEAD_BYTENAME_G1 ("\014\014\014\014")
(No commands found)
Item SOURCE_DATA_POVERHEAD_BYTENAME_F2
:source:data:telecom:overhead:preset
:source:data:telecom:poverhead:preset
Item SOURCE_DATA_POVERHEAD_BYTENAME_H4
(No commands found)
Item SOURCE_DATA_POVERHEAD_BYTENAME_Z3
:source:data:telecom:overhead:preset
:source:data:telecom:poverhead:preset
Item SOURCE_DATA_POVERHEAD_BYTENAME_Z4
:source:data:telecom:overhead:preset
:source:data:telecom:poverhead:preset
Item SOURCE_DATA_POVERHEAD_BYTENAME_Z5
:source:data:telecom:overhead:preset
:source:data:telecom:poverhead:preset
=============================================




CTS 850 Cross-Reference listing Remote Commands 46
42.

----P_TX_PATH_OVERHEAD_V5_PARAM1----
Transmitting STS-1 : 1 under test
Overhead View .............. STS Path Overhead


V5:




----------------------
Item TX_PATH_OVERHEAD_SELECT_V5
(No commands found)
Item TX_PATH_OVERHEAD_TRIB_VT_SIG_LABEL ("\014\014\014\014")
(No commands found)
=============================================




CTS 850 Cross-Reference listing Remote Commands 47
43.

----P_TX_PATH_OVERHEAD_VC12_PARAM1----
Transmitting STS-1 : 1 under test
Overhead View .............. STS Path Overhead

V5 Signal Label ............

Path Overhead Path Trace Message
V5:
J2: N2:
K4:




----------------------
Item TX_PATH_OVERHEAD_SELECT_VC12
(No commands found)
Item ..._V5 ("V5 Signal Label ...")
(No commands found)
Item TX_PATH_OVERHEAD_TRIB_TU12_SIG_LABEL
(No commands found)
Item TX_PATH_OVERHEAD_TRIB_VALUE_J2
(No commands found)
Item TX_PATH_OVERHEAD_TRIB_VALUE_Z6
:source:data:telecom:overhead:preset
:source:data:telecom:poverhead:preset
Item TX_PATH_OVERHEAD_TRIB_VALUE_Z7
:source:data:telecom:overhead:preset
:source:data:telecom:poverhead:preset
=============================================




CTS 850 Cross-Reference listing Remote Commands 48
44.

----P_TX_PATH_OVERHEAD_VC3_PARAM1----
Transmitting STS-1 : 1 under test
Overhead View .............. STS Path Overhead

C2 Signal Label ............
H4 Multiframe Sequence ..... 4 Bytes

Path Overhead Path Trace Message
J1: B3:
C2:
G1:
F2:
H4:
Z3:
Z4:
Z5:
----------------------
Item TX_PATH_OVERHEAD_SELECT_VC3 ("Overhead View ...")
(No commands found)
Item ..._C2 ("C2 Signal Label ...")
(No commands found)
Item ..._H4 ("H4 Multiframe Sequence ...")
(No commands found)
Item TX_PATH_OVERHEAD_TRIB_VALUE_J1
:source:data:telecom:overhead:preset
:source:data:telecom:poverhead:preset
Item TX_PATH_OVERHEAD_TRIB_VALUE_C2
:source:data:telecom:overhead:preset
:source:data:telecom:poverhead:preset
Item TX_PATH_OVERHEAD_TRIB_VALUE_G1 ("\014\014\014\014")
(No commands found)
Item TX_PATH_OVERHEAD_TRIB_VALUE_F2
:source:data:telecom:overhead:preset
:source:data:telecom:poverhead:preset
Item TX_PATH_OVERHEAD_TRIB_VALUE_H4
(No commands found)
Item TX_PATH_OVERHEAD_TRIB_VALUE_Z3
:source:data:telecom:overhead:preset
:source:data:telecom:poverhead:preset
Item TX_PATH_OVERHEAD_TRIB_VALUE_Z4
:source:data:telecom:overhead:preset
:source:data:telecom:poverhead:preset
Item TX_PATH_OVERHEAD_TRIB_VALUE_Z5
:source:data:telecom:overhead:preset
:source:data:telecom:poverhead:preset
=============================================




CTS 850 Cross-Reference listing Remote Commands 49
45.

----P_TX_POINTERS_TIMING_FREQUENCY_PARAM1----
Pointer / Timing Mode ...... Pointer Movements
Offset Mode ................ Offset Line (Payload fixed)
Frequency Offset ...........

Press POINTER ACTION for a single movement.
It will alternate between increment and decrement.




----------------------
Item TX_POINTERS_TIMING_MODE_FO ("Pointer / Timing Mode ...")
:source:data:telecom:pointer:mode
Item SOURCE_CLOCK_OFFSET_MODE ("Offset Mode ...")
:source:clock:offset:mode
Item TX_POINTERS_TIMING_OFFSET ("Frequency Offset ...")
(No commands found)
=============================================




CTS 850 Cross-Reference listing Remote Commands 50
46.

----P_TX_POINTERS_TIMING_PARAM1----
Pointer / Timing Mode ...... Pointer Movements
Pointer Type ............... STS Pointer
Pointer Control ............ Set Value
Pointer Rate ...............
Pointer Direction .......... Increment
Pointer ss-Bits ............ 00

Press POINTER ACTION for a single movement.
It will alternate between increment and decrement.




----------------------
Item TX_POINTERS_TIMING_MODE ("Pointer / Timing Mode ...")
:source:data:telecom:pointer:mode
Item TX_POINTERS_TIMING_TYPE ("Pointer Type ...")
(No commands found)
Item TX_POINTERS_TIMING_CONTROL ("Pointer Control ...")
:source:data:telecom:pointer:mode
:source:data:telecom:tributary:pointer:mode
Item TX_POINTERS_TIMING_NBURST ("Pointer Rate ...")
(No commands found)
Item TX_POINTERS_TIMING_MOVEMENT_VALUE ("Pointer Rate ...")
(No commands found)
Item TX_POINTERS_TIMING_MOVEMENT_NDF ("Pointer Direction ...")
:source:data:telecom:pointer:ndflag
:source:data:telecom:tributary:pointer:ndflag
Item TX_POINTERS_TIMING_MOVEMENT_SBITS ("Pointer ss-Bits ...")
(No commands found)
Item TX_POINTERS_TIMING_POINTER_RATE ("Pointer Rate ...")
(No commands found)
Item TX_POINTERS_TIMING_DIRECTION ("Pointer Direction ...")
:source:data:telecom:pointer:direction
:source:data:telecom:tributary:pointer:direction
=============================================




CTS 850 Cross-Reference listing Remote Commands 51
47.

----P_TX_POINTERS_TIMING_SEQUENCE_PARAM1----
Pointer / Timing Mode ...... Pointer Movements
Pointer Type ............... STS Pointer
Sequence Type (G.783)....... a) Single Alternating
Pointer Direction .......... Increment
Pointer Rate ...............
Initialization Time ........
Cool Down Time .............
Tributary Mapping Offset ...
Status: Current State: Not Running
Time Between Anomalies:
Time Until Next Anomaly:

Press POINTER ACTION to START Pointer Sequences



----------------------
Item TX_POINTERS_TIMING_MODE_PS ("Pointer / Timing Mode ...")
:source:data:telecom:pointer:mode
Item TX_POINTERS_TIMING_TYPE_PS ("Pointer Type ...")
(No commands found)
Item SOURCE_DATA_POINTER_SEQUENCE_TYPE ("Sequence Type (G.783)...")
:source:data:telecom:pointer:sequence:type
:source:data:telecom:tributary:pointer:sequence:type
Item SOURCE_DATA_TRIBUTARY_POINTER_SEQUENCE_TYPE ("Sequence Type (G.783)...")
:source:data:telecom:pointer:sequence:type
:source:data:telecom:tributary:pointer:sequence:type
Item SOURCE_DATA_POINTER_SEQUENCE_DIRECTION ("Pointer Direction ...")
(No commands found)
Item SOURCE_DATA_TRIBUTARY_POINTER_SEQUENCE_DIRECTION ("Pointer Direction ...")
(No commands found)
Item TX_POINTERS_TIMING_SEQUENCE_RATE ("Pointer Rate ...")
(No commands found)
Item TX_POINTERS_TIMING_SEQUENCE_IPERIOD_READOUT ("Initialization Time ...")
(No commands found)
Item TX_POINTERS_TIMING_TRIBUTARY_SEQUENCE_IPERIOD_READOUT ("Initialization Time
...")
(No commands found)
Item TX_POINTERS_TIMING_SEQUENCE_CPERIOD_READOUT ("Cool Down Time ...")
(No commands found)
Item TX_POINTERS_TIMING_TRIBUTARY_SEQUENCE_CPERIOD_READOUT ("Cool Down Time ...")
(No commands found)
Item TX_POINTERS_TIMING_SEQUENCE_MAPPING_OFFSET ("Tributary Mapping Offset ...")
(No commands found)
=============================================




CTS 850 Cross-Reference listing Remote Commands 52
48.

----P_TX_SDH_K64_SETTINGS_PARAM1----
2 Mb/s Framing.............. Framed
Configuration............... N x 64k
Number of Timeslots.........
Starting Timeslot...........
Test Pattern................ 2^23-1 Normal (ITU)

Active Timeslots

<< < << < << < < < < << < << < << < << < < << << < < <
1 5 10 15 20 25 30




----------------------
Item ..._FRAMING ("2 Mb/s Framing...")
:input2:telecom:terminator
:source:data:telecom:tributary:add
:source:data:telecom:tributary:framing
Item ..._CONFIGURATION ("Configuration...")
(No commands found)
Item ..._WIDTH ("Number of Timeslots...")
(No commands found)
Item ..._TIMESLOT ("Starting Timeslot...")
(No commands found)
Item ..._TEST_PATTERN ("Test Pattern...")
:source:data:telecom:payload:pattern
:source:data:telecom:tributary:pattern
Item ..._TEST_PATTERN_POLARITY ("Test Pattern...")
:source:data:telecom:payload:prbs:invert
Item ..._UBYTE ("Test Pattern...")
:source:data:telecom:payload:pattern
:source:data:telecom:tributary:pattern
=============================================




CTS 850 Cross-Reference listing Remote Commands 53
49.

----P_TX_SDH_SETTINGS_PARAM1----
Transmit Rate .............. STM-4 Optical
Transmit Clock ............. Internal
AU Under Test .............. 1igh
SDH Structure .............. AU-4 Bulk
TU Under Test .............. :1 (Group 1, TUG-2:1 TU-12:1
TU Background Fill ....... Idle (11010101)
Payload Structure .......... 140 Mb/suipped)Framed
Add Signal Type & Level .... HDB3>2^23-1 Normal (X-C)TU)
34Mb/s Active Channel..... 1 Background 2^15-1
8Mb/s Active Channel..... 1 Background 2^15-1
2Mb/s Active Channel..... 1 Background 2^15-1
Tx/Rx Setup ................ Independent




----------------------