Text preview for : HighVoltg_CVAppNote1.pdf part of Keithley HighVoltg CVAppNote1 Keithley Appnotes HighVoltg_CVAppNote1.pdf



Back to : HighVoltg_CVAppNote1.pdf | Home

Number 2972

Using the Model 4200-CVU-PWR C-V Power Package to Make
Application Note High Voltage and High Current C-V Measurements with the
Series Model 4200-SCS Semiconductor Characterization System



Introduction terminals and it is best that the AC ammeter be connected to the
gate of the device if applicable. If this is the case, then the HCUR
Traditional capacitance-voltage (C-V) testing of semiconductor and HPOT (or CVH1) terminals of the 4200-CVU are connected
materials is typically limited to about 30V and 10mA DC to the other side of the device or to the chuck. The DC bias is
bias. However, many applications, such as characterizing C-V supplied by the 4200-SMU, with the Force and Sense from the
parameters of LD MOS structures, low k interlayer dielectrics, SMU connected to the Force and Sense terminals of the bias
MEMs devices, organic TFT displays, and photodiodes, require tee. The SMU LO terminals and the CVU CVH1 terminals are
higher voltage or higher current C-V measurements. For these referenced internally to the system output common.
applications, a separate high voltage DC power supply and a
capacitance meter are required to make the measurements.
4200-CVU
The Model 4200-CVU-PWR C-V Power Package for the Model C
H CUR
4200-SCS allows making C-V measurements with a DC voltage SMA Tee
bias of up to