TB 9-4931-702-50
CHANGE 3
DEPARTMENT OF THE ARMY TECHNICAL BULLETIN
CALIBRATION PROCEDURE FOR
DC TRANSFER STANDARD
JOHN FLUKE MODELS
730A AND 730A/AB
Headquarters, Department of the Army, Washington, DC
20 August 2001
Approved for public release; distribution is unlimited.

TB 9-4931-702-50, 10 May 1974, is changed as follows:
1. Remove old pages and insert new pages as indicated below. New or changed
material is indicated by a vertical bar in the margin of the page.
Remove pages Insert pages
1 and 2 1 and 2
5 and 6 5 and 6

2. File this change sheet in front of the publication for reference purposes.




0120707

Distribution:
To be distributed in accordance with Std. IDS No. RLC-1500, 11 May, 1992.

.



PIN: 009083-003
TB 9-4931-702-50
Change 2
DEPARTMENT OF THE ARMY TECHNICAL BULLETIN
CALIBRATION PROCEDURE FOR
DC TRANSFER STANDARD,
JOHN FLUKE MODELS 730A
AND 730A/AB
Headquarters, Department of the Army, Washington, DC
18 February 1986
TB 9-4931-702-50, 10 May 1974, is changed as follows:

Page 2. In table 2, item A2, Minimum Use Specifications column, add superscript 1.

In table 2, add footnote to table as follows:

"1Prior to using the NULL DETECTOR, determine sensitivity on the 3 and 10