DC MODULE CONNECTORS Installation Guide
Standard Connector (P/N 5060-3386)
Relay Connector (P/N 5060-3387)
General Information +Sense used to connect the + remote sense lead. Set the
sense switch to Remote when this connection is
Load connections are made on the connector supplied
used.
with each module. The connector is fastened to the back
of the mainframe. Modules can then be installed or -Sense used to connect the - remote sense lead. Set the
removed from the mainframe without disturbing the sense switch to Remote when this connection is
wiring on the connector. used.
The connector has a switch to select either local or remote GND used for signal ground (GND) connections.
voltage sensing, depending on how the load is connected.
Connect the Wires
A relay connector is identified by a check mark on the Loosen the barrier block screws and connect the wires.
Relay label on the connector. Relay connectors include The barrier block accepts spade or ring terminals. You
load and sense disconnect and polarity reversal relays. can also connect a stripped wire to the barrier block.
Twist the wire around under the screw terminal and
Making Load Connections tighten the screw.
Load connections are the same on both the standard and
the relay connector.
Open the Connector
Open the connector by loosening the four corner screws
(M3 X 10 mm).
TIE
WRAP
Route the wires out of the bottom of the connector as
shown in the figure above. Use the tie-wrap supplied
CAUTION