Text preview for : 077084700(2290-10-SHVBH).pdf part of Keithley 077084700(2290-10-SHVBH) Keithley 2290 077084700(2290-10-SHVBH).pdf
Back to : 077084700(2290-10-SHVBH). | Home
Model 2290-10-SHVBH
Keithley Instruments, Inc. SHV Female Bulkhead Connector
28775 Aurora Road
Cleveland, Ohio 44139
1-888-KEITHLEY
http://www.keithley.com
Description
The Model 2290-10 can contain hazardous voltages. Be certain that the high voltage is
completely discharged before removing or connecting the high-voltage cable. High-voltage
cables can store charge if they are disconnected from the supply while high voltage is turned
on and can cause personal injury or death if not handled properly.
Do not connect the HIGH-VOLTAGE output to exposed circuitry. Any load connected to the
HIGH-VOLTAGE output should be enclosed in a metal shield that is connected to safety earth
ground using #18AWG or larger wire.
The high-voltage 10 kV SHVBH SHV female bulkhead connector connects to the Model 2290-10-SHVUC male to
unterminated cable (refer to document number 077084500 for more information).
High-voltage installation:
1. Turn off power and allow all voltage to discharge.
2. Install the Model 2290-10-SHVBH female bulkhead connector to the Model 2290-10-SHVUC male to
unterminated cable (see next figure).
Figure 1: Model 2290-20 SHV female bulkhead connector
Electrical characteristics
Nominal impedance non-constant
Voltage rating 10 kV DC
Insulation resistance 10E6 M
077084700 *077084700* 1
Model-10-SHVBH Female Bulkhead Connector
Material characteristics
A. Insulator: high-density polyethylene, ASTM-D4976, natural
B. O-rings: silicone rubber, A-A-59588
C. Lock-washer: phosphor bronze
D. All other parts: brass, ASTM-B16
E. Center contact: gold plate per ASTM-B488
F. Leakage may not exceed 10x-7 cc/second of tracer gas at atmospheric pressure
2 077084700