Text preview for : 2600B-800A-readme.pdf part of Keithley 2600B-800A-readme Keithley 2600 2600B-800A-readme.pdf



Back to : 2600B-800A-readme.pdf | Home

Appendix G

Model 2400 emulation

In this appendix:

Model 2400 emulation ............................................................. G-1




Model 2400 emulation
The Series 2600B provides for emulation of the Model 2400 command set using a personality script
named Persona2400. When run, this script takes control of the remote command interfaces and
interprets any commands received.


Loading, running, and configuring Model 2400 emulation

Before running or configuring the script, it must be loaded into internal memory. Also, you cannot run
or load a script while a script is already running.




If the Persona2400 script has been deleted from the Series 2600B without using the script's
DeleteScript menu item (which is contained in the Configure2400 user test), you must execute the
userstring.delete("AutoRun2400") command before reloading the Persona2400 script.
To load the script into the internal memory:
1. Plug the USB drive (provided with the Series 2600B) into the front panel USB port.
2. Load the Persona2400 script (2600B-800A.tsp) and save the script to the Series 2600B
internal nonvolatile memory.
a. From the front panel, press the MENU key, and then select SCRIPT > LOAD > USB1.
c. Use the navigation wheel to select the Persona2400.tsp script. If prompted to overwrite an existing
script, select YES to overwrite (it may take a few seconds to complete loading the script).
d. Turn the navigation wheel to select SAVE-INTERNAL and then press the navigation wheel (or
the ENTER key).
e. Select YES and then press the navigation wheel (or the ENTER key) to save the script internally (it
may take a few seconds to complete the save).
f. Press the EXIT key as needed to leave the menu structure.
Appendix G: Model 2400 emulation Series 2600B System SourceMeter