Text preview for : 077084400(2290-5-SHVBH).pdf part of Keithley 077084400(2290-5-SHVBH) Keithley 2290 077084400(2290-5-SHVBH).pdf
Back to : 077084400(2290-5-SHVBH).p | Home
Model 2290-5-SHVBH
Keithley Instruments, Inc. SHV Male Bulkehead Connector
28775 Aurora Road
Cleveland, Ohio 44139
1-888-KEITHLEY
http://www.keithley.com
Description
The Model 2290-5 can contain hazardous voltages. Be certain that the high voltage is
completely discharged before removing or connecting the high-voltage cable. High-voltage
cables can store charge if they are disconnected from the supply while high voltage is turned
on and can cause personal injury or death if not handled properly.
Do not connect the HIGH-VOLTAGE output to exposed circuitry. Any load connected to the
HIGH-VOLTAGE output should be enclosed in a metal shield that is connected to safety earth
ground using #18AWG or larger wire.
The high-voltage 5 kV SHV male bulkhead connector is used to connect external equipment.
High-voltage installation:
1. Turn off power and allow all voltage to discharge.
2. Install the Model 2290-5-SHVBH SHV male bulkhead connector (see next figures).
Figure 1: Model 2290-5 SHV male bulkhead connector
Figure 2: Model 2290-5 SHV male bulkhead connector mounting dimensions
077084400 *077084400* 1
Model 2290-5-SHVBH Connector
Characteristics
Nominal impedance non-constant
Maximum operating voltage (RMS) 3500 RMS
5000 V DC
Insulation resistance 10E6 M
Temperature range -65