WAVESURFER OSCILLOSCOPE
Errata Sheet for Remote Control Manual
Most of the commands contained in the Remote Control Manual for the WaveMasterTM DSO and other
X-StreamTM instruments apply also to the WaveSurfer oscilloscope. There are also two new WaveSurfer
scope commands that are not in the manual. The Remote Control Manual can be found in the Technical
Library at LeCroy's Web site http://www.lecroy.com.
Please refer to the following table for additions, exclusions, and modifications.

Command Status Notes
COMBINE_CHANNELS Not Applicable
Allowed modes: CUST [,STAT], FAIL, HREL [,ABS, DELTA,
CURSOR_MEASURE Modified
SLOPE], OFF, PASS, VREL [,ABS, DELTA]
Allowed traces: F1, F5, F6, F7, F8, TA, C1, C2, C3, C4, M1, M2,
CURSOR_SET Modified M3, M4
F5 to F8 correspond to Z1 to Z4, respectively.
CURSOR_VALUE? Modified Allowed traces: F1, F5, F6, F7, F8, TA, C1, C2, C3, C4
CURSORS Not Applicable
Allowed function is F1 only.
DEFINE Modified
Not all function parameters are allowed.
FUNCTION_RESET Modified Allowed functions: F1 or TA.
GRID Modified Allowed arguments: SINGLE, XYONLY, XYSINGLE
Allowed prefixes:

HOR_MAGNIFY Modified TA or F1 for the Math trace
Z1 to Z4 for F5 to F8
F2 to F4, TB to TD are not valid
Allowed prefixes:

HOR_POSITION Modified TA or F1 for the Math trace
Z1 to Z4 for F5 to F8
F2 to F4, TB to TD are not valid
This command sets the input impedance.
Syntax:
IMPEDANCE New
:IMPEDANCDE
:= A1M, D1M, D50, GND, 0VL
For external trigger: D50, D1M, GND
Allowed prefixes:
C1 to C4 for channel traces
INSP? Modified M1 to M4 for memory traces
TA or F1 for the Math trace
Z1 to Z4 for F5 to F8


1
WAVESURFER OSCILLOSCOPE
F2 to F4, TB to TD are not valid
MEMORY_SIZE Not Applicable
Allowed parameter types: CUST, OFF
PARAMETER Modified
Allowed parameter readout: STAT, OFF
PARAMETER_CLR Modified Pass/Fail is not applicable.
PARAMETER_CUSTOM Modified Only parameter columns 1 to 6 allowed.
Pass/Fail is not applicable.
PARAMETER_DELETE Modified
Only parameter columns 1 to 6 allowed.
Allowed parameter type: CUST
PARAMETER_STATISTICS Modified
Allowed parameters: P1 to P6
Allowed traces: F1, F5, F6, F7, F8, TA, C1, C2, C3, C4
PARAMETER_VALUE? Modified
Not allowed: CUST7, CUST8
argument is not valid.
PASS_FAIL Modified
settings allowed: AllTrue, AllFalse
Allowed traces: F1, F5, F6, F7, F8, TA, C1, C2, C3, C4, M1, M2,
PER_CURSOR_SET Modified
M3, M4
PERSIST_SAT Modified Trace selection is not allowed.
PERSIST_SETUP Modified Allowed mode: ALL
SAMPLE_CLOCK Modified Allowed states: INTERNAL, ECL, TTL, 0V
SEQUENCE Not Applicable
STORE_SETUP Modified F2 to F4, TB to TD are not valid
TRACE Modified F2 to F4, TB to TD are not valid
ETM10 is not valid
TRIG_COUPLING Modified
New values: AC, DC, HF, HFREJ, LFREJ
TRIG_LEVEL Modified ETM10 is not valid
TRIG_SELECT Modified New types: TV, RUNT, SLEW
TRIG_SLOPE Modified ETM10 is not valid
The TRIG_WINDOW command sets the window amplitude in
volts on the current Edge trigger source. The window is
centered around the Edge trigger level.
TRIG_WINDOW New
Syntax:
TRIG_WINDOW
:= 0 to 25 V
VERT_MAGNIFY Modified F2 to F4, TB to TD are not valid
VERT_POSITION Modified F2 to F4, TB to TD are not valid
WAVEFORM Modified F2 to F4, TB to TD are not valid




2