Text preview for : 5988-2410EN.pdf part of HP 5988-2410EN HP Publikacje 5988-2410EN.pdf



Back to : 5988-2410EN.pdf | Home

Agilent
E8267C/E8257C/E8247C PSG

Application Note




Obtain flat-port power
with Agilent's PSG
user flatness correction
or external leveling functions
E8247C PSG CW signal generator



Agilent E8244A




E8257C PSG analog signal generator




Agilent E8254A


E8267C PSG vector signal generator




2
Table of Contents Introduction


Using External Leveling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 The PSG synthesized signal genera-
tors provide extremely flat power
Leveling with a Millimeter-Wave Source Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7 at your test port for testing power
sensitive devices such as amplifiers,
Creating and Applying User Flatness Correction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9 mixers, diodes, or detectors. The user
flatness correction feature of the
Creating a User Flatness Correction Array Using a MM-Wave Source Module . . . . . . . .14 PSGs compensates for attenuation
and power variations created by
Performing the User Flatness Correction Manually . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18 components between the source and
the test device.
Saving the User Flatness Correction Data to the Memory Catalog . . . . . . . . . . . . . . . . . .19
User flatness correction allows the
digital adjustment of RF output
amplitude for up to 1601 frequency
points in any frequency or sweep
mode (for example, start/stop, CW,
power sweep, and more.). Using
a power meter to calibrate the
measurement system, a power level
correction table is created for the
frequencies where power level varia-
tions or loss occur. These frequen-
cies may be defined in sequential
linear steps or arbitrarily spaced.
To allow different correction arrays
for different tests setup or different
frequency ranges, you may save
individual user flatness correction
tables to the signal generators
memory catalog and recall them
on demand.

This application note illustrates how
to utilize the user flatness correction
feature by providing step-by-step
instructions for several common
measurement examples.




3
Using External Leveling


The PSG signal generator can be Figure 3-1.
leveled externally by connecting an External detector Signal generator
external sensor at the point where leveling with a
MENUS
E8254A 250 kHz