Manual for Model Number 3324A
Manual printed on October 1990
Manual Part Number 03324-90001

Make all ERRATA corrections.

Check the following table for your instrument serial prefix/serial number/EDC and make the listed
changes to your manual

New Item

Serial Prefix or Manual
Serial Number Changes

ERRATA

3009G00806 1
3009G02706 2




14 July, 1998 Page 1 of 10
02'(/ $
BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB
INDEX OF MANUAL CHANGES

MANUAL FRAME
CHANGE
ERRATA
1 W4,5
2 MP30




__________________________________________________________________________________
07/14/98 Page 2 of 10
02'(/ $
BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB
ERRATA

On Page 1-9, General Information (Specifications)

Frequency Sweep
.
.
Linear Sweep (settable for each interval)
.
.
ADD: Minimum sweep rate: 0.2 Hz/sec.

On Page 2-9, Installation
step 6: change to read:

...Refer to Fig. 2.7

On Page 2-10, Installation
step 8: change to read:

...Refer to Fig. 2.7

On Page 4-1, Replaceable Parts, change to read:

Introduction
.
.
Repair philosophy .... However Appendix B ....


Performance Test

Triangle Linearity

Page 12-37 in the 03324-90011 Manual
Page C-26 in the 03324-90001 Manual

step 13:
change to read:
.
.
.
y=a1x+ao


step 14:
change to read:

For each delay (x), subtract the calculated voltage (y') from

the measured voltage (y) [y-y'= y]. Find the largest positive

or negative voltage difference ( y max). Using the following

Formular, computer the % linearity:

% linearity = (|y max | / 4 Volt) x 100%


__________________________________________________________________________________
07/14/98 Page 3 of 10
02'(/ $
BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

Performance Test

Amplitude Accuracy

Page 12-24 and 12-49, in the 03324-90011 Manual

Page C-18 and C-39, in the 03324-90001 Manual

step 32, or Test Record List

change to read:

Amplitude 1 V New Limits!
DC Oddset 1 mV

Test Freq Function Minimum Maximum
13 99.9Hz Square 0.958 V 1.042 V
15 99.9Hz Triangle 0.953 V 1.047 V
18 99.9Hz + Ramp 0.953 V 1.047 V
19 99.9Hz - Ramp 0.953 V 1.047 V
21 1kHz Square 0.958 V 1.042 V
22 2kHz Triangle 0.953 V 1.047 V
23 500Hz + Ramp 0.953 V 1.047 V
24 500Hz - Ramp 0.953 V 1.047 V
25 100kHz Square 0.880 V 1.120 V
26 10kHz Triangle 0.918 V 1.082 V
27 10kHz + Ramp 0.868 V 1.132 V
28 10kHz - Ramp 0.868 V 1.132 V


Specifications

page 1-8 in the 03324-90001 manual

DC Offset

change to read:
Accuracy:

a) DC only:

DC Offset Range Accuracy
1.5000 V to 5.0000 V +/- 50 mV
0.5000 V to 1.4999 V +/- 17 mV
0.0150 V to 0.4999 V +/- 5 mV
0.0500 V to 0.1499 V +/- 1.7 mv
0.0150 V to 0.0499 V +/- 0.52 mV
0.0050 V to 0.0149 V +/- 0.19 mV
0.0015 V to 0.0049 V +/- 0.07 mV
0.0005 V to 0.0014 V +/- 0.037 mV


The accuracy given to each range is the same for each value in this range.




__________________________________________________________________________________
07/14/98 Page 4 of 10
02'(/ $
BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

b) DC + AC, < 1 MHz:

Sinewave + Squarewave

Amplitude Range Accuracy
1.0000 V to 10.0000 V +/- 60 mV
0.3000 V to 0.0000 V +/- 20 mV
0.1000 V to 0.2999 V +/- 6 mV
0.0300 V to 0.0999 V +/- 2 mV
0.0100 V to 0.0299 V +/- 0.6 mV
0.0030 V to 0.0099 V +/- 0.2 mV
0.0010 V to 0.0029 V +/- 0.06 mV

The accuracy given to each range is the same for each value in this range


Ramp + Triangle, upto 11 kHz

Amplitude Range Accurancy
1.0000 V to 10.0000 V +7- 120 mV
0.3000 V to 0.9999 V +/- 40 mV
0.1000 V to 0.2999 V +/- 12 mV
0.0300 V to 0.0999 V +/- 4 mV
0.0100 V to 0.0299 V +/- 1.2 mV
0.0030 V to 0.0099 V +/- 0.4 mV
0.0010 V to 0.0029 V +/- 0.12 mV

The accuracy given to each range is the same for each value in this range.


c) DC + AC, < 1 MKz

Sinewave up to 21 MHz

Squarewave up to 11 MHz


Amplitude Range Accuracy
1.0000 V to 10.0000 V +/- 150 mV
0.3000 V to 0.9999 V +/- 48 mV
0.1000 V to 0.2999 V +/- 15 mV
0.0300 V to 0.0999 V +/- 4.8 mV
0.0100 V to 0.0299 V +/- 1.5 mV
0.0030 V to 0.0099 V +/- 0.49 mV
0.0010 V to 0.0029 V +/- 0.15 mV

The accurancy given to each range is the same for each value in this range.




__________________________________________________________________________________
07/14/98 Page 5 of 10
02'(/ $
BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

Section: General Information

Add: on page 1-3, after "Tools and Test Equipment:

Recalibration, cleaning and lubrication

The Performance Tests should be performed at 12-month intervals, if necessary using the
adjustment procedure described in Chapter 3 in this manual should any of the
performance tests fail.
At the same intervals the instrument should be inspected mechanically and for cleanliness.
Dust is best removed with a small vacuum cleaner or finebrush. Accumulation of dirt on the
front panel or covers can be removed with a soft cloth if necessary slightly moistened with
water.

The instrument requires no lubrication.


on page 1-14, General Characteristics

add: Calibration cycle: 12 months recommended

page 3-7, Adjustments

Amplitude

Negative setting

change to read:

step 9: ... Specification requirement is -10V +/- 0.2V.

Positive setting

change to read:

step 2: ... Specification requirement is 10V +/- 0.2V.


page Contents-4, Table of Contents

add after line "General Fault Finding":

Access - Cover Removal ....... 5-27


page 5-27, General Fault Finding:

change the heading "Access" to read:

Access - Cover Removal

page 5-32, Service

removal and refitting procedure

add: Cover removal is described on page 5-27, Access - Cover Removal.




__________________________________________________________________________________
07/14/98 Page 6 of 10
02'(/ $
BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

page 4-1, Replaceable Parts

Introduction

change to read:

... However Appendix B to this manual is reserved for an optional information package,
CLIP p/n 03324-90031, which contains a full parts listing and schematics for those
customers who require it.

In Table 4-2, Page 4-5, add the following lines after the linebegining "MP28":
MP29 Screw M M3X0.58 mm 0515-0897
Securing screw for board assemblies.




__________________________________________________________________________________
07/14/98 Page 7 of 10
02'(/ $
BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

Page 1-14 add:

'HFODUDWLRQ RI &RQIRUPLW\

(similar to ISO/IEC Guide 22)


Manufacturer: Hewlett-Packard GmbH
Boeblingen Instruments Division
Herrenberger Str. 130
D-71034 Boeblingen Germany

We declare that the product


+3 $ )XQFWLRQ6ZHHS *HQHUDWRU


conforms to the following standards:

Safety: IEC 348 (1978)

EMC: EN 55011 (1991)/CISPR 11 Group1, Class B
EN 50082-1 (1991)
IEC 801-2 ESD: 4kV cd, 8kV ad
IEC 801-3 Radiated Immunity: 3V/m
IEC 801-4 Fast Transients: 0.5kV, 1kV
Supplementary Information:

During the measurements against EN 55011, the I/O ports were terminated with their
nominal impedance, the HP-IB connection was terminated with the cable HP 10833B.
When the product is connected to other devices, the user must ensure that the connecting
cables and the other devices are adequately shielded to prevent radiation.




Boeblingen, 29th April 1993
Robert Hofgaertner
Quality Assurance Manager




__________________________________________________________________________________
07/14/98 Page 8 of 10
02'(/ $
BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

Page C-17, Performance Tests

change to read:
.
.

Test Frequenz Function Minimum Maximum
25 101kHz Square 2.850 V 3.150 V
.
.

Page C-38 Test Record

Amplitude: 3 V (p-p)
change to read:

.
. Min Measured Max
Square, 101kHz 2.850 V ------- 3.150 V
.
.


page 1-14 Operating Environment:
add errata:
Altitude up to 2000m
Installation Category II
Pollution Degree 2
Warning: To prevent electrical shock, disconnent the HP model 3324A from mains before cleaning. Use a dry cloth or
one slightly dampened with water to clean the external case parts. Do not attempt to clean internally.


========================================================================

! CHANGE TO NEW SERIAL NUMBER FORMAT !

last serial number old format 3009G02755
first serial number new format DE30202756

========================================================================




__________________________________________________________________________________
07/14/98 Page 9 of 10
02'(/ $
BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

MANUAL CHANGE 1

On Page 4-6, change the Table of Replaceable Parts to read:

W4 03324-61601 CBL AY PWR SS
W5 03324-61602 CBL AY PWR UC


MANUAL CHANGE 2

On Page 4-4 Repl.Parts, Table 4-2 add:

MP30 7121-5585 Ce-Label




__________________________________________________________________________________
07/14/98 Page 10 of 10