Text preview for : 077084300(2290-5-SHV).pdf part of Keithley 077084300(2290-5-SHV) Keithley 2290 077084300(2290-5-SHV).pdf



Back to : 077084300(2290-5-SHV).pdf | Home

Model 2290-5-SHV
Keithley Instruments, Inc. SHV Plug to SHV Plug Cable
28775 Aurora Road
Cleveland, Ohio 44139
1-888-KEITHLEY
http://www.keithley.com



Description

The Model 2290-5 can contain hazardous voltages. Be certain that the high voltage is
completely discharged before removing or connecting the high-voltage cable. High-voltage
cables can store charge if they are disconnected from the supply while high voltage is turned
on and can cause personal injury or death if not handled properly.
Do not connect the HIGH-VOLTAGE output to exposed circuitry. Any load connected to the
HIGH-VOLTAGE output should be enclosed in a metal shield that is connected to safety earth
ground using #18AWG or larger wire.

The high-voltage output cable connects the Model 2290-5 source output to external equipment.
High-voltage cable installation:
1. Turn off power and allow all voltage to discharge.
2. Install Model 2290-5-SHV high-voltage cable (see next figure).
Figure 1: Model 2290-5 SHV cable




077084300 *077084300* 1
Model 2290-5-SHV plug to SHV plug cable


Characteristics
Length 10 feet (3.05 meters)

Type RG-8A/U coax

Maximum operating voltage (RMS) 5,000 volts

Impedance 50 Ohms




2 077084300