File name calk2000_ver.txtKeithley Model 2000 - detected, S/N ???????, Version: A20 /A02
Environment temperature = 31.86 C, humidity = 35.66 RH, pressure = 1005.7023 hPa
HP 3458A detected
TEMP? = 38.0 C
Ambient temp = +24.7C
HP 3458A DCV -1.873613851E-06 VDC, K2000 DUT -1.017560509E-06 VDC, Delta 841280.036 ppm
MFC error = 186.828 ppm
HP 3458A DCV 9.999802543E-03 VDC, K2000 DUT 1.000123133E-02 VDC, Delta -142.861 ppm
MFC error = 33.143 ppm
HP 3458A DCV 9.999981272E-02 VDC, K2000 DUT 1.000001956E-01 VDC, Delta -3.828 ppm
MFC error = -16.392 ppm
HP 3458A DCV 9.999972156E-01 VDC, K2000 DUT 1.000001285E+00 VDC, Delta -4.069 ppm
MFC error = -4.707 ppm
HP 3458A DCV 9.999972689E+00 VDC, K2000 DUT 1.000000753E+01 VDC, Delta -3.484 ppm
MFC error = -5.978 ppm
HP 3458A DCV 1.000008679E+02 VDC, K2000 DUT 9.999968567E+01 VDC, Delta 11.823 ppm
MFC error = 18.251 ppm
HP 3458A DCV 1.000005285E+03 VDC, K2000 DUT 1.000015667E+03 VDC, Delta -10.381 ppm
MFC error = 162.782 ppm
Measured REF DCI 5 : 9.999833622E-05 ADC, stored DCI = 0.099984
MFC error = 7.402 ppm
Measured REF DCI 5 : 9.999984695E-04 ADC, stored DCI = 0.999993
MFC error = 14.221 ppm
Measured REF DCI 5 : 9.999985530E-03 ADC, stored DCI = 9.999858
MFC error = -14.197 ppm
Measured REF DCI 5 : 9.999997257E-02 ADC, stored DCI = 100.001420
MFC error = -81.391 ppm
Measured REF DCI 5 : 9.999965331E-01 ADC, stored DCI = 1.000081
MFC error = 28.861 ppm
MFC error = -307.161 ppm
MFC error = -312.260 ppm
MFC error = -57.844 ppm
MFC error = -20.743 ppm
MFC error = 78.178 ppm
MFC error = 1042.746 ppm
Calibrations count = 3
Calibration last date = 2016,07,21
:CAL:PROT:CODE 'KI002000'
CAL MFC error = -4.694 ppm
:CAL:PROT:DC:STEP3 9.999975
CAL MFC error = -19.045 ppm
:CAL:PROT:DC:STEP4 -9.999991
CAL MFC error = 3.023 ppm
:CAL:PROT:DC:STEP5 100.001520
:CAL:PROT:DC:STEP6 999.687755
:CAL:PROT:DC:STEP7 9999.418338
:CAL:PROT:DC:STEP8 99997.989270
:CAL:PROT:DC:STEP9 1000079.693333
:CAL:PROT:DC:STEP10 10010415.926667
:CAL:PROT:DC:STEP11 0.010000
:CAL:PROT:DC:STEP12 0.100000
:CAL:PROT:DC:STEP13 0.999997
K2000 CAL complete
HP 3458A DCV -1.000000000E+38 VDC, K2000 DUT -4.230067813E-01 VDC, Delta 236402829488445149402248186038634766753857536.000 ppm |