datasheet,schematic,electronic components, service manual,repairs,tv,monitor,service menu,pcb design
Schematics 4 Free
Service manuals, schematics, documentation, programs, electronics, hobby ....


registersend pass
Bulgarian - schematics repairs service manuals SearchBrowseUploadWanted

Now downloading free:Agilent HP 8904A Service Notes 1

Agilent HP 8904A Service Notes 1 free download

Various electronics service manuals

File information:
File name:HP 8904A Service Notes 1.pdf
[preview HP 8904A Service Notes 1]
Size:44 kB
Extension:pdf
Mfg:Agilent
Model:HP 8904A Service Notes 1 🔎
Original:HP 8904A Service Notes 1 🔎
Descr: Agilent HP 8904A Service Notes 1.pdf
Group:Electronics > Other
Uploaded:25-08-2020
User:Anonymous
Multipart:No multipart

Information about the files in archive:
Decompress result:OK
Extracted files:1
File name HP 8904A Service Notes 1.pdf

8904A-04 S E R V I C E N O T E SUPERSEDES: None 8904A Multifunction Synthesizer (Option 006/H16) Serial Numbers: 2948A04593 / 2948A04693 Replacement of Audio Transformer to Correct Inverted Waveform Parts Required: 9100-4833, Audio Transformer Situation: Some 8904A Multifunction Synthesizer (Option 006/H16) with serial numbers listed above may have inverted OUTPUT 1 waveforms. This is because the primary of Audio Transformer T1 was wound backwards, resulting in a 180 degree phase shift from the pri- mary to the secondary. OUTPUT 2 waveforms are not affected. The most noticeable waveform is the ramp because it is asymmetrical. Sinewave, square and triangle waveforms are inverted but are less noticeable because they are symmetrical. Normally the ramp waveform starts at the negative peak and ramps up to the positive peak. Suspect units will exhibit an inverted ramp waveform, i.e., the ramp starts at posi- tive peak and ramps down to the negative peak. NOTE: Option 006/H16 is not indicated on the front panel LCD display when ann instru- ment preset is performed. Verify if Option 006/H16 is installed by inspecting the option labelon the rear panel of the unit. Continued DATE: 01 January 1992 ADMINISTRATIVE INFORMATION SERVICE NOTE CLASSIFICATION: MODIFICATION RECOMMENDED ACTION IMMEDIATELY STANDARDS: CATEGORY: ON SPECIFIED FAILURE LABOR: 1.0 Hours AGREEABLE TIME LOCATION CUSTOMER INSTALLABLE SERVICE RETURN USED RETURN CATEGORY: ON-SITE INVENTORY: SCRAP PARTS: SCRAP SERVICE CENTER SEE TEXT SEE TEXT AVAILABILITY: PRODUCT'S SUPPORT LIFE RESPONSIBLE UNTIL: 1000 UNTIL: 01 January 1997 AUTHOR: LHL ENTITY: 1000 ADDITIONAL INFORMATION:

>> View document online <<



>> Download document << eServiceInfo Context Help



Was this file useful ? Share Your thoughts with the other users.

User ratings and reviews for this file:

DateUserRatingComment

Average rating for this file: 0.00 ( from 0 votes)


Similar Service Manuals :
Agilent ReleaseNotes SW ReleaseNotes SW Desktop A050220121101 - Agilent README.txt README DEE A.05.00.2012.0630 - Agilent nc param - Agilent ReleaseNotes SW ReleaseNotes SW A050320130124 - Agilent ReleaseNotes FW A.05.00.2012.0710.txt ReleaseNotes FW A.05.00.2012.0710 - Agilent Release Note FW A.05.04.2013.0328 ReleaseNotes FW A0504 20130328 - Agilent ReleaseNotes SW ReleaseNotes SW A.05.00.2012.0630 -
 FB -  Links -  Info / Contacts -  Forum -   Last SM download : Bose 2150

script execution: 0.07 s