File name ReleaseNotes_FW A.05.00.2012.0710.txt ReleaseNotes_FW_A.05.00.2012.0710.txt== Agilent B1500A/B1505A Firmware/FLEX control Software A.05.00.2012.0710 Release Notes ==
== NEW FEATURES ==
The following new features are added in this Release.
- B1511A support for B1505A
- B1513B support for B1505A
- B1514A support for B1505A
- N1265A support for B1505A
- N1266A support for B1505A
- N1268A support for B1505A
== FIXED DEFECTS ==
The following defects are fixed in this Release.
- [8242]: [Trigger IO] Support of MM27 and MM28
- [8734]: [Self calibration] Self calibration should abort on abnormal measurement value.
== REVISION HISTORY ==
- [8439]: Step voltage of linear search is different from setup value on HVSMU.
- [8454]: [MM16] Unexpected error 224 may occur for issuing XE due to uncleared WNX setting.
- [8468]: Dual HCSMU force 2.3 A constant current causes overflow status.
- [8490]: [MM16] The first step after a polarity change may be output in wrong polarity in case of reaching current compliance.
- [8493]: [*LRN?] Sub-channel number of wrong offset is returned for B1525A.
- [8528]: Output from HPSMU kept at pulse peak value in multi channel pulsed spot measurement.
- [8595]: [MM10] HVSMU may not output negative current for voltage compliance setting above 1500V.
- [8602] FLEX GUI may return a wrong response for ERR? after receiving multiple DCL.
- [8403] [Enh. Req.] weighted factor mode for HRSMU High-resolution ADC integration time
- [8400] [FLEX] Does not generate SRQ after emergency.
- [8247] B1505A + N1259A: Interlock error occurs intermittently even the lid of N1259A is closed.
- [8278] [HVSMU] The limit of voltage range gain adjustment is too small.
- [8309] [MM9] Unexpected error 307 occurs for omitting measurement channel for quasi-pulse measurement mode.
- [8324] Transitional current may be observed due to insufficient wait time for negative current setting of SMU.
- [8133] [Module Selector] Output relay of SMU once assigned with ERHPA is on after execution of *TST? although in case of ERMOD 0.
- [8134] [DHC] Inconsistency may occur for self test/calibration results of Dual HCSMU.
- [8140] [HCSMU] Cannot write SPI EEPROM.
- [8146] [Module Selector] ERHPA/ERHPP must be rejected in case of High Voltage Status.
- [8147] [Module Selector] Response of *LRN? may be inconsistent for slots controlled with ERHPP.
- [8160] [DFSMU] Current pulse may have error of less than a few percents due to inconsistent current bias polarity.
- [8165] [SPGU] Improve the algorithm of power amp idle current calibartion.
- [8166] [SPGU] Don't report error message in *CAL? when module failed.
- [8168] Firmware hungs up in case of detecting high voltage status whil |