File name ReleaseNotes_FW.A.05.60.2014.0926 Release Note_FW A.05.60.2014.0926.txt== Keysight B150xA Series Firmware/FLEX control Software A.05.60.2014.0910 Release Notes ==
== NEW FEATURES ==
The following features are added in this release.
- B1513C support
- B1506A support
- B1507A support
== IMPROVEMENTS ==
The following improvements are added in this release.
- [9005] [FLEX Service] error handling for each category
- [9183] [CPU Support FPGA] Enhancement of monitoring interlock and high voltage status
== FIXED DEFECTS ==
The following defects are fixed in this release.
- [8638] [N1265A Thermometer Support] Board temperature may not be read.
- [9007] [FLEX Service] Cannot save GPIB log file in specified folder.
- [9027] Sending MCPNT command with wrong channel number causes main frame shutdown.
- [9062] [B1520A] Freq sweep measurement value has large dips after OSL cal.
- [9071] [B1511B Support] ASU serial number reading is to be supported.
- [9168] [N1265A Support] ICMR test limit for 1500AR is to be 3.5A.
- [9189] [Dual HCSMU Support] Unexpected error 204 occurs for CL command in case of INTLKVTH 0.
- [9237] [B1513x] Temp. limit in self test is wrong.
- [9271] [DR14I-010] B1500A shutdown may occur for HRSMU@10pA measurement noise.
- [9281] SMU may apply high voltage after interlock open during application measurement.
- [9287] [N1265A] Calibration data of ERPFUHCCAL? is not stored in EEPROM.
- [9312] I/V Sweep measurement speed may differ depending on ranging and polarity.
- [9338] [CMU] OSL correction data measurement may return incorrect result.
== REVISION HISTORY ==
- B1511B support for B1500A
- B1511B/E5288A support for B1500A
- B1514A support for B1500A
- B1511B support for B1505A
- Maximum number of sweep point is enhanced to 10,001 for I/V sweep measurements of MM2, MM4, MM5, MM16 and MM28.
- [6958] [DR13I-004] B1500A FlexPrg Initialization Failed
- [9000] Compliance range for DV command is not checked for SMU type.
- [8982] Fault tolerance for hardware failure is to be enhanced for interlock and emergency handlings.
- [8925] [N1268A Support] Unexpected error 1004 may occur depending on pulse base and peak settings.
- [8937] Pulse fall timing below 20us of MCSMU causes unexpected error 233 in case of measuring with HPSMU/MPSMU.
- [8943] [SPGU] ALWG output delay is not compensated correctly.
- [8946] Sending DSMPLSETUP, DSMPLARM, DSMPLFLUSH or SOVC to non-SMU channel causes firmware crash.
- [8952] [HCSMU/HVSMU/MCSMU Support] 0A force offset correction may not work for negative current sweep.
- [8851] [N1265A + N1266A Support] Diagnosis using HVSMU is to be performed only in case of direct connection of HVSMU.
- [8869] [N1268A Support] Wrong error 694 is reported for missing N1268A instead of appropriate error 940.
- [8872] [N1266 |