File name HP 81110A with 81111A Perf. Test p97.pdf5
5Agilent 81110A/'11A
Performance Test
1
Agilent 81110A/'11A Performance Test
Introduction
Use the tests in this chapter if you want to check that the Agilent
81110A Pulse Generator Frame with the Agilent 81111A 165
MHz Output Channel(s) is working correctly. Before starting any
testing allow all test equipment to warm up for at least 30 minutes.
Conventions Used
When referring to actions that you perform during the tests, the
following conventions are used:
FUNCTION This indicates that a labelled button must be pressed
[MODE/TRG] This shows that a soft-key must be pressed. A soft-
key is an unlabelled button whose label is shown on the display,
and which can vary according to the job that the button is doing
CONTINUOUS PULSES This is an option shown on the dis-
play, and is selected by use of the vernier keys. It is shown in
upper or lower case to match the case displayed.
Test Results Tables
Tables for entering the results of the tests are included at the end
of this chapter. The tests are numbered and reference numbers for
each Test Result (TR) are given in a small table at the end of each
test. The reference number shows you where the actual results
should be entered in the Test Results Tables.
The Test Results tables at the end of the chapter should be photo-
copied, and the Test Results entered on the copies. Then, if the
tests need to be repeated, the tables can be copied again.
2 Agilent 81110A/'11A Performance Test
Agilent 81110A/'11A Performance Test
If Channel 2 has been fitted to your instrument, make an extra
copy of the Test Results tables for entry of the results of tests on
that channel. In this case, however, it is not necessary to repeat the
Period tests, as these are common to both channels.
Agilent 81110A/'11A Performance Test 3
Agilent 81110A/'11A Performance Test
Recommended Test Equipment and Accessories
The following tables list the recommended test equipment you
need to perform all the tests in this chapter. You can use alterna-
tive instruments if they meet the critical specifications given. The
test set-ups and procedures assume you are using the recom-
mended equipment.
Test Equipment Model Critical Specifications
Oscilloscope Agilent 54121T 20 GHz, 10 bit vertical resolution, Histogram
or
Oscilloscope Agilent 54750A + |