File information:
File name:LECROY AP034-OM-E-06 Proformace.pdf
[preview LECROY AP034-OM-E-06 Proformace]
Size:93 kB
Extension:pdf
Mfg:LeCroy
Model:LECROY AP034-OM-E-06 Proformace 🔎
Original:LECROY AP034-OM-E-06 Proformace 🔎
Descr: LeCroy LECROY AP034-OM-E-06 Proformace.pdf
Group:Electronics > Other
Uploaded:26-09-2019
User:Anonymous
Multipart:No multipart

Information about the files in archive:
Decompress result:OK
Extracted files:1
File name LECROY AP034-OM-E-06 Proformace.pdf

3HUIRUPDQFH 9HULILFDWLRQ 3HUIRUPDQFH 9HULILFDWLRQ This procedure can be used to verify the warranted characteristics of the AP034 Active Differential Probe. The recommended calibration interval for the model AP034 Active Differential Probe is one year. The complete performance verification procedure should be performed as the first step of annual calibration. You can record test results on a photocopy of the Test Record provided at the end of this section. You can do the performance verification without removing the instrument covers, exposing yourself to hazardous voltages. Adjustment should only be attempted if a parameter measured in the Performance Verification Procedure is outside of the specified limits. Adjustment should only be performed by qualified personnel. TEST EQUIPMENT REQUIRED Table 3 lists the test equipment and accessories (or their equivalents) that are required to verify the performance of the AP034 Active Differential Probe. This procedure is designed to minimize the number of calibrated test instruments required. Only the parameters listed in boldface in the "Minimum Requirements" column must be calibrated to the accuracy indicated. Because the input and output connector types may vary on different brands and models of test instruments, additional adapters or cables may be required. AP034-OM-E Rev D ISSUED: January 2000

>> View document online <<



>> Download document << eServiceInfo Context Help



Was this file useful ? Share Your thoughts with the other users.

User ratings and reviews for this file:

DateUserRatingComment

Average rating for this file: 0.00 ( from 0 votes)


Similar Service Manuals :
LeCroy LeCroy 94XX Remote Control Manual - LeCroy 93XXC-OM-E17 - LeCroy 93XXC-OM-E19 - LeCroy 93XXC-OM-E20 - LeCroy 93XXC-OM-E23 - LeCroy LECROY WS Series Operation - LeCroy LECROY WJ Series Getting Started -
 FB -  Links -  Info / Contacts -  Forum -   Last SM download : Galaxy Pluto

script execution: 0.01 s