File information: | |
File name: | 3458A-18A.pdf [preview 3458A-18A] |
Size: | 144 kB |
Extension: | |
Mfg: | Agilent |
Model: | 3458A-18A 🔎 3458A18A |
Original: | 3458A-18A 🔎 |
Descr: | Agilent 3458A service 3458A-18A.pdf |
Group: | Electronics > Other |
Uploaded: | 11-01-2020 |
User: | Anonymous |
Multipart: | No multipart |
Information about the files in archive: | ||
Decompress result: | OK | |
Extracted files: | 1 | |
File name 3458A-18A.pdf MODIIFIICATIION RECOMMENDED MOD F CAT ON RECOMMENDED 3458A-18A S E R V I C E N O T E Supersedes: 3458A-18 3458A Multimeter Serial Numbers: US28031400 / US28032927 Some recently produced 3458As may be "out of DCV specifications" due to time drift issues. One issue is long-term drift of the internal reference that may cause the instrument to be out of DCV specification in time intervals less than 1 year. The second issue is short-term drift that may cause some instruments to be out of specifications within 24 hours following an ACAL. To Be Performed By: Either qualified customers or Agilent Service Personnel Parts Required: P/N Description Qty. 03458-69503 PCA A/D & Inguard Logic 1 (rebuilt A3 assembly) ADMINISTRATIVE INFORMATION SERVICE NOTE CLASSIFICATION: MODIFICATION RECOMMENDED ACTION [[]] IMMEDIATELY STANDARDS: CATEGORY: X ON SPECIFIED FAILURE LABOR: 3.0 Hours [[]] AGREEABLE TIME LOCATION X CUSTOMER INSTALLABLE SERVICE [[]] RETURN USED [[]] RETURN CATEGORY: [[]] ON-SITE INVENTORY: [[]] SCRAP PARTS: [[]] SCRAP X SERVICE CENTER X SEE TEXT X SEE TEXT AVAILABILITY: PRODUCT'S SUPPORT LIFE AGILENT RESPONSIBLE UNTIL: March 31, 2006 AUTHOR: DLL PRODUCT LINE: 9E ADDITIONAL INFORMATION: |