File name 2430PulseMode.pdf Why does my 2430 fail to Source High Energy Pulses using the Pulse Mode on the 10A Range? In the 10A range, the duty cycle is limited to 8% with a maximum pulse width of 2.5ms. To achieve a duty cycle of 8% with the maximum allowable pulse width of 2.5 ms, the Pulse OFF Time should be set for no less than 28.75ms. Duty cycle = Pulse On Time___________ Pulse On time + Pulse Off Time Pulse Off time = (2.5ms / 0.08) |