datasheet,schematic,electronic components, service manual,repairs,tv,monitor,service menu,pcb design
Schematics 4 Free
Service manuals, schematics, documentation, programs, electronics, hobby ....


registersend pass
Bulgarian - schematics repairs service manuals SearchBrowseUploadWanted

Now downloading free:Keithley Interlock 6487

Keithley Interlock 6487 free download

Various electronics service manuals

File information:
File name:Interlock_6487.pdf
[preview Interlock 6487]
Size:35 kB
Extension:pdf
Mfg:Keithley
Model:Interlock 6487 🔎
Original:Interlock 6487 🔎
Descr: Keithley 6487 Interlock_6487.pdf
Group:Electronics > Other
Uploaded:01-03-2020
User:Anonymous
Multipart:No multipart

Information about the files in archive:
Decompress result:OK
Extracted files:1
File name Interlock_6487.pdf

A GREATER MEASURE OF CONFIDENCE Product(s) Picoammeter: 6487 Question: Why do I get a "CLOSE INTLCK" error when trying to turn ON the source? Answer: Model 6487 has 3 voltage source ranges: 10V, 50V, and 500V. The default power up sequence enables the 10V source range. On this range you should never see "CLOSE INTLCK " error on turning the OPERATE ON. On the 50V and 500V ranges you could possibly see "CLOSE INTLCK" error when turning the OPERATE ON. The Interlock prevents the voltage source being operated on the 50V and 500V range, to help assure Safe operation. The Interlock is always operational on these source ranges. The sole purpose of this is to prevent any contact with the live circuitry when making measurements if using a Test Fixture with a lid. A normally open switch is connected to pins 1 and 2 of the connector as shown. When the switch is OPEN, Interlock is asserted and voltage source cannot be placed in OPERATE on the 50 V and 500 V source ranges, hence you get the error " CLOSE INTLOCK". When the lid is closed, this shorts pins 1 and 2, allowing you to source voltages. The model 6487 is shipped with a standard CS-459 connector mounted on the Interlock I/P connector. The CS-459 shorts pins 1 and 2. If not using a Test Fixture, make sure the CS-459 connector is mounted on the Interlock I/P Connector. Otherwise, when you try to turn the output ON (50 or 500 volt range) the Instrument will give a "CLOSE INTLCK" error.

>> View document online <<



>> Download document << eServiceInfo Context Help



Was this file useful ? Share Your thoughts with the other users.

User ratings and reviews for this file:

DateUserRatingComment

Average rating for this file: 0.00 ( from 0 votes)


Similar Service Manuals :
Keithley 98932C(KPCMCIA_RS422_485S) - Keithley 6514RevA_DocSpec - Keithley Selector LowV-LowR - Keithley 2765 Thermistors - Keithley 052207 4ACS - Keithley 79480C(DASCard) - Keithley 3321RevB DocSpec -
 FB -  Links -  Info / Contacts -  Forum -   Last SM download : Sanwa SANWA+CX506A

script execution: 0.03 s