File information: | |
File name: | Bupl305e.pdf [preview Bupl305e] |
Size: | 241 kB |
Extension: | |
Mfg: | Rohde & Schwarz |
Model: | Bupl305e 🔎 |
Original: | Bupl305e 🔎 |
Descr: | Rohde & Schwarz R&S UPL Document Set Bupl305e.pdf |
Group: | Electronics > Other |
Uploaded: | 01-03-2020 |
User: | Anonymous |
Multipart: | No multipart |
Information about the files in archive: | ||
Decompress result: | OK | |
Extracted files: | 1 | |
File name Bupl305e.pdf R&S UPL Previous/Actual Data Comparison Supplement to the R&S UPL Manual The R&S UPL software version 3.02/3.03 has been upgraded to 3.04/3.05. The modifications in 3.04 and 3.05 are described below. Brief Overview GENERATOR panel: Previously ... Now ... ... frequency sweeps were only possible with ... this error is eliminated. an activated equalizer when you selected at least three sweep points. With only two For manual control, see sweep points, a wrong output level was set. 2.5.4.1.3 Equalization of SINE, SINE BURST, DFD, MULTISINE, RANDOM For IEC/IEEEbus command, see 3.10.1.4.2 Sweep Settings for Gen. SIN, STER, BURS, S2P, MDIS,DFD, DC ... in the RANDOM Domain TIME function, ... this noise generator has been improved in such a acoustically interfering periodic artfacts were way that there is no periodicity anymore. audible but irrelevant to measurements. For manual control, see 2.5.4.9 RANDOM For IEC/IEEEbus command, see 3.10.1.5.7 RANDOM ... negative phase parameters were allowed in ... this error is eliminated. the MULTISINUS function but were inter- nally limited to 0 |
Date | User | Rating | Comment |