Text preview for : High_Resistance_Measurements.PDF part of Fluke High_Resistance_Measurements Fluke appnotes High_Resistance_Measurements.PDF
Back to : High_Resistance_Measureme | Home
HIGH RESISTANCE MEASUREMENTS
Karl-Peter Lallmann
Fluke Deutschland GmbH
Heinrich-Hertz-Strasse 11
34123 Kassel / Germany
Phone:++49 561 9594 651
Fax: ++49 561 9594 609
E-mail: [email protected]
Abstract
This paper describes the accredited calibration methodology used in the Calibration Laboratory operated by
Fluke Germany. It is based on the use of a High-Ohms Resistance Bridge made by Measurement International
together with appropriate Resistance Standards from 10 k to 10 M from Guildline and Russian made
Resistance Standards from 100 M to 1G .
Indroduction
The calibration of Standard Resistors in the range from 10 k to 1 G is based on the use of a High-Ohms Resis-
tance Bridge, model 6000B, made by Measurement International/Canada (MI) and the use of Standard Resistors
from 10 k to 1 G.
The Model 6000B from MI is a binary voltage divider which measures the ratio between the connected known and
unknown resistors with very low uncertainty. The ratio follows the following equation:
RX V1 - V 2 V 1/E - V 2 /E r1 - r 2
V= = = =
RN V 3 - V 4 V 3 /E - V 4 /E r 3 - r 4
The value of the unknown resistor is calculated as follows:
R X = R N