File name Release Note_HW A.05.03.2013.0128 ReleaseNotes_FW_A0503_20180128.txt== Agilent B1500A/B1505A Firmware/FLEX control Software A.05.03.2013.0128 Release Notes ==
== FIXED DEFECTS ==
The following defects are fixed in this Release.
- [8925] [N1268A Support] Unexpected error 1004 may occur depending on pulse base and peak settings.
- [8937] Pulse fall timing below 20us of MCSMU causes unexpected error 233 in case of measuring with HPSMU/MPSMU.
- [8943] [SPGU] ALWG output delay is not compensated correctly.
- [8946] Sending DSMPLSETUP, DSMPLARM, DSMPLFLUSH or SOVC to non-SMU channel causes firmware crash.
- [8952] [HCSMU/HVSMU/MCSMU Support] 0A force offset correction may not work for negative current sweep.
== REVISION HISTORY ==
- [8851] [N1265A + N1266A Support] Diagnosis using HVSMU is to be performed only in case of direct connection of HVSMU.
- [8869] [N1268A Support] Wrong error 694 is reported for missing N1268A instead of appropriate error 940.
- [8872] [N1266A Support] PV mode set by ERHVPV is to be reset at ERHVCA and ERHVCTST? to avoid unexpected failure.
- [8893] [B1512A/B1514A] Voltage pulse falling may be slow for 20A/1A range in case of different polarities for pulse base and peak.
- [8915] [UHCU Support] Power Compliance does not work for I Pulse Sweep.
- Signal Monitor related command support
- [8785] The measurement is not performed at the specified timing, if over 2.097152s is set for Measurement Delay Time.
- [8820] [MM10] Negative base voltage is not output for base hold time in case of 0V source setting.
- [8847] Firmware may stall for racing condition of SRQ and byte-in/out.
- B1511A support for B1505A
- B1513B support for B1505A
- B1514A support for B1505A
- N1265A support for B1505A
- N1266A support for B1505A
- N1268A support for B1505A
- [8242]: [Trigger IO] Support of MM27 and MM28
- [8734]: [Self calibration] Self calibration should abort on abnormal measurement value.
- [8439]: Step voltage of linear search is different from setup value on HVSMU.
- [8454]: [MM16] Unexpected error 224 may occur for issuing XE due to uncleared WNX setting.
- [8468]: Dual HCSMU force 2.3 A constant current causes overflow status.
- [8490]: [MM16] The first step after a polarity change may be output in wrong polarity in case of reaching current compliance.
- [8493]: [*LRN?] Sub-channel number of wrong offset is returned for B1525A.
- [8528]: Output from HPSMU kept at pulse peak value in multi channel pulsed spot measurement.
- [8595]: [MM10] HVSMU may not output negative current for voltage compliance setting above 1500V.
- [8602] FLEX GUI may return a wrong response for ERR? after receiving multiple DCL.
- [8403] [Enh. Req.] weighted factor mode for HRSMU High-resolution ADC integration time
|